Comparison of novel anterior fixation method, W-plate, versus superior parasymphyseal plating in anterior posterior compression type III pelvic ring injuries: a biomechanical study.
Brandi Ann Krieg, Todd Baldini, Christopher Orear, Keenan Masami Onodera, Nicholas Jerome Tucker, Adam Schlauch, Justin E Hellwinkel, Nathaniel Elliot Schaffer, Cyril Mauffrey
{"title":"Comparison of novel anterior fixation method, W-plate, versus superior parasymphyseal plating in anterior posterior compression type III pelvic ring injuries: a biomechanical study.","authors":"Brandi Ann Krieg, Todd Baldini, Christopher Orear, Keenan Masami Onodera, Nicholas Jerome Tucker, Adam Schlauch, Justin E Hellwinkel, Nathaniel Elliot Schaffer, Cyril Mauffrey","doi":"10.1097/BOT.0000000000003304","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To compare stiffness, displacement, and load to failure (LTF) patterns of a novel anterior pelvic plating technique, W-plating, to superior parasymphyseal plating (SPSP) in anterior posterior compression type III (APC III) pelvic ring injuries in a biomechanical model.</p><p><strong>Methods: </strong>Twelve fourth generation composite pelves with induced APC III injuries stabilized with posterior screw fixation and either W-plating or SPSP were stressed via the Instron machine under single leg stance (SLS) axial load and external rotation (ER) load. Motion was analyzed at the pubic symphysis (PS) and sacroiliac joints (SIJ). PS and SIJ displacement, rotation, and stiffness were compared between the W-plate and SPSP groups under simulated stressed conditions. T-test was used to compare means and significance was set at p<0.05. Unadjusted p-values were reported alongside effect sizes (Cohen's d) and 95% confidence intervals (CI) to provide additional context regarding the magnitude and potential biomechanical relevance of the observed differences.</p><p><strong>Results: </strong>The W-plate demonstrated less inferior PS displacement than the SPSP under SLS with a large observed effect size (0.5 ± 0.4 mm vs 1.2 ± 0.5 mm, respectively, Cohen's d = 1.47). Average PS displacement also demonstrated a large effect size favoring the W-plate construct (Cohen's d = 1.07), though statistical significance was not reached (0.3 ± 0.3 mm vs 0.7 ± 0.3 mm respectively, p = 0.07). There were no significant differences in stiffness in the SLS for the W-plate versus SPSP (98.9 ± 11.2 vs 105 ± 6.6 N/mm, respectively, p = 0.6) or in ER (4.4 ± 2.2 v 3.1 ± 1.6, respectively, p = 0.3). When loaded to failure, the W-plate and the SPSP both failed at the posterior fixation with no significant difference between groups (1411± 221 v 1545 ± 225 N, respectively p = 0.4).</p><p><strong>Conclusions: </strong>In a biomechanical analysis of induced APC-III pelvic ring injury, W-plating performed similarly to SPSP in stiffness, displacement, and LTF. W-plating for anterior pelvic ring fixation was found to be biomechanically similar to SPSP with potentially increased resistance to anterior ring displacement.</p><p><strong>Level of evidence: </strong>V.</p>","PeriodicalId":16644,"journal":{"name":"Journal of Orthopaedic Trauma","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-10-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Trauma","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/BOT.0000000000003304","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To compare stiffness, displacement, and load to failure (LTF) patterns of a novel anterior pelvic plating technique, W-plating, to superior parasymphyseal plating (SPSP) in anterior posterior compression type III (APC III) pelvic ring injuries in a biomechanical model.
Methods: Twelve fourth generation composite pelves with induced APC III injuries stabilized with posterior screw fixation and either W-plating or SPSP were stressed via the Instron machine under single leg stance (SLS) axial load and external rotation (ER) load. Motion was analyzed at the pubic symphysis (PS) and sacroiliac joints (SIJ). PS and SIJ displacement, rotation, and stiffness were compared between the W-plate and SPSP groups under simulated stressed conditions. T-test was used to compare means and significance was set at p<0.05. Unadjusted p-values were reported alongside effect sizes (Cohen's d) and 95% confidence intervals (CI) to provide additional context regarding the magnitude and potential biomechanical relevance of the observed differences.
Results: The W-plate demonstrated less inferior PS displacement than the SPSP under SLS with a large observed effect size (0.5 ± 0.4 mm vs 1.2 ± 0.5 mm, respectively, Cohen's d = 1.47). Average PS displacement also demonstrated a large effect size favoring the W-plate construct (Cohen's d = 1.07), though statistical significance was not reached (0.3 ± 0.3 mm vs 0.7 ± 0.3 mm respectively, p = 0.07). There were no significant differences in stiffness in the SLS for the W-plate versus SPSP (98.9 ± 11.2 vs 105 ± 6.6 N/mm, respectively, p = 0.6) or in ER (4.4 ± 2.2 v 3.1 ± 1.6, respectively, p = 0.3). When loaded to failure, the W-plate and the SPSP both failed at the posterior fixation with no significant difference between groups (1411± 221 v 1545 ± 225 N, respectively p = 0.4).
Conclusions: In a biomechanical analysis of induced APC-III pelvic ring injury, W-plating performed similarly to SPSP in stiffness, displacement, and LTF. W-plating for anterior pelvic ring fixation was found to be biomechanically similar to SPSP with potentially increased resistance to anterior ring displacement.
目的:在生物力学模型中比较一种新型骨盆前路镀技术w镀和上副椎管镀(SPSP)在前后路压迫型III (APC III)骨盆环损伤中的刚度、位移和载荷失效(LTF)模式。方法:对12例经后路螺钉固定、w -镀或SPSP稳定的第四代复合椎体APC III型损伤患者,在单腿站立(SLS)轴向载荷和外旋(ER)载荷作用下,在Instron机器上受力。分析耻骨联合(PS)和骶髂关节(SIJ)的运动。比较了w板和SPSP组在模拟应力条件下的PS和SIJ位移、旋转和刚度。结果:在SLS下,w板的PS位移小于SPSP,观察到的效应量较大(分别为0.5±0.4 mm vs 1.2±0.5 mm, Cohen’s d = 1.47)。平均PS位移也显示出较大的效应量(Cohen’s d = 1.07),尽管没有达到统计学意义(分别为0.3±0.3 mm vs 0.7±0.3 mm, p = 0.07)。w型钢板SLS与SPSP的刚度无显著差异(分别为98.9±11.2 vs 105±6.6 N/mm, p = 0.6), ER的刚度分别为4.4±2.2 v 3.1±1.6,p = 0.3)。加载失效时,w型钢板和SPSP在后路固定时均失效,两组间无显著差异(分别为1411±221 N和1545±225 N, p = 0.4)。结论:在诱导APC-III型骨盆环损伤的生物力学分析中,w -镀与SPSP在刚度、位移和LTF方面的表现相似。w电镀骨盆前环固定在生物力学上与SPSP相似,对前环移位的抵抗力可能增加。证据等级:V。
期刊介绍:
Journal of Orthopaedic Trauma is devoted exclusively to the diagnosis and management of hard and soft tissue trauma, including injuries to bone, muscle, ligament, and tendons, as well as spinal cord injuries. Under the guidance of a distinguished international board of editors, the journal provides the most current information on diagnostic techniques, new and improved surgical instruments and procedures, surgical implants and prosthetic devices, bioplastics and biometals; and physical therapy and rehabilitation.