A Hand-Palm Technique for Localization of the Greater Trochanter Tip During Proximal Femoral Nailing: A Prospective Pilot Validation Study.

Sharan Mallya, Ashok Puranik
{"title":"A Hand-Palm Technique for Localization of the Greater Trochanter Tip During Proximal Femoral Nailing: A Prospective Pilot Validation Study.","authors":"Sharan Mallya, Ashok Puranik","doi":"10.13107/jocr.2026.v16.i10.8354","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>An accurate localization of the tip of the greater trochanter (GT) is required for proper skin incision placement and correct entry point creation during proximal femoral nailing (PFN) for intertrochanteric femur fractures. Direct palpation of the GT tip may be unreliable in obese or edematous patients, potentially resulting in extended incisions and improper entry points. This study describes the \"hand-palm technique,\" a surface landmarking method for localization of the GT tip.</p><p><strong>Materials and methods: </strong>The technique was applied in two variations based on the surgeon's glove size. For surgeons with glove size ≥7, the width of four medial fingers placed along the iliac crest from the anterior superior iliac spine was used to identify the GT tip. For surgeons with glove size ≤6.5, the entire palm, including the thumb, was used similarly. Fluoroscopic confirmation was performed in all cases. In addition, an observational pilot study involving 41 patients undergoing PFN was conducted to assess localization accuracy, fluoroscopy usage, operative parameters, and reproducibility of the technique. Descriptive statistical analysis and subgroup comparison between obese and non-obese patients were performed. A body mass index-based subgroup analysis was additionally performed to assess localization accuracy, fluoroscopy usage, reproducibility, and operative parameters across different body habitus categories.</p><p><strong>Results: </strong>A total of 41 patients were included in the pilot study. The mean localization error from the fluoroscopically confirmed GT tip was 2.19 ± 1.74 mm, and the mean number of fluoroscopic images required was 1.66 ± 0.62. The mean operative time was 39.90 ± 7.01 min, the mean blood loss was 119.51 ± 49.85 mL, and the mean incision length was 5.61 ± 0.40 cm. Obese patients demonstrated a trend toward higher localization error and fluoroscopy requirement compared with non-obese patients.</p><p><strong>Conclusion: </strong>The hand-palm technique is a simple and practical surface landmarking method with satisfactory localization accuracy in this pilot study. It may reduce repeated fluoroscopic localization attempts and unnecessary incision extension, particularly in obese or edematous patients. However, fluoroscopic confirmation remains essential, and further prospective studies are needed to validate it.</p>","PeriodicalId":16647,"journal":{"name":"Journal of Orthopaedic Case Reports","volume":"16 10","pages":"619-624"},"PeriodicalIF":0.0000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13646816/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Case Reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.13107/jocr.2026.v16.i10.8354","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: An accurate localization of the tip of the greater trochanter (GT) is required for proper skin incision placement and correct entry point creation during proximal femoral nailing (PFN) for intertrochanteric femur fractures. Direct palpation of the GT tip may be unreliable in obese or edematous patients, potentially resulting in extended incisions and improper entry points. This study describes the "hand-palm technique," a surface landmarking method for localization of the GT tip.

Materials and methods: The technique was applied in two variations based on the surgeon's glove size. For surgeons with glove size ≥7, the width of four medial fingers placed along the iliac crest from the anterior superior iliac spine was used to identify the GT tip. For surgeons with glove size ≤6.5, the entire palm, including the thumb, was used similarly. Fluoroscopic confirmation was performed in all cases. In addition, an observational pilot study involving 41 patients undergoing PFN was conducted to assess localization accuracy, fluoroscopy usage, operative parameters, and reproducibility of the technique. Descriptive statistical analysis and subgroup comparison between obese and non-obese patients were performed. A body mass index-based subgroup analysis was additionally performed to assess localization accuracy, fluoroscopy usage, reproducibility, and operative parameters across different body habitus categories.

Results: A total of 41 patients were included in the pilot study. The mean localization error from the fluoroscopically confirmed GT tip was 2.19 ± 1.74 mm, and the mean number of fluoroscopic images required was 1.66 ± 0.62. The mean operative time was 39.90 ± 7.01 min, the mean blood loss was 119.51 ± 49.85 mL, and the mean incision length was 5.61 ± 0.40 cm. Obese patients demonstrated a trend toward higher localization error and fluoroscopy requirement compared with non-obese patients.

Conclusion: The hand-palm technique is a simple and practical surface landmarking method with satisfactory localization accuracy in this pilot study. It may reduce repeated fluoroscopic localization attempts and unnecessary incision extension, particularly in obese or edematous patients. However, fluoroscopic confirmation remains essential, and further prospective studies are needed to validate it.

在股骨近端钉钉过程中,手掌技术定位大转子尖端:一项前瞻性先导验证研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
128
审稿时长
30 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书