Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.

IF 2.6 3区 医学 Q2 ANESTHESIOLOGY
Shotaro Sakimura, Kaiki Oya, Yuka Sakaguchi, Ken Yamaura
{"title":"Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.","authors":"Shotaro Sakimura, Kaiki Oya, Yuka Sakaguchi, Ken Yamaura","doi":"10.1007/s10877-026-01510-3","DOIUrl":null,"url":null,"abstract":"<p><p>To evaluate the perioperative performance of two minimally/non-invasive hemoglobin monitoring modalities-continuous pulse hemoglobin (SpHb) (Revision O: rev. O) and intermittent point-of-care testing (POCT) using HemoCue® Hb801-against an automated hematology analyzer as the reference standard. Fifty patients undergoing general anesthesia with invasive arterial monitoring for surgery with anticipated blood loss were enrolled prospectively. A total of 281 arterial blood samples were analyzed by an automated hematology analyzer (reference standard), a blood gas analyzer, and HemoCue® Hb801; all comparisons, including HemoCue®, used arterial rather than capillary blood. SpHb (rev. O) was monitored continuously. Correlation and Bland-Altman analyses were performed, and polar plot analysis of changes between successive sampling points assessed SpHb trend-tracking. HemoCue® Hb801 agreed closely with the reference (r = 0.979; bias 0.119 g/dL; limits of agreement [LOA] -0.687 to +0.925 g/dL). SpHb correlated with the reference (r = 0.777) but showed wide LOA (- 2.114 to +2.811 g/dL), proportional bias, and greater random error at perfusion index <1.4 (87/281 measurements, 31.0%). In polar plot analysis of 23 successive-change pairs from 9 patients, angular bias was - 5.8° (95% CI -12.2° to +0.5°) and four-quadrant concordance 100%, but radial limits of agreement (- 36.3° to +24.6°) exceeded the ±30° criterion. HemoCue® Hb801 provided accurate Hb measurement from a minimal arterial sample. SpHb (rev. O) had limited absolute accuracy and could not quantify the magnitude of change within accepted limits, but reliably indicated its direction. A hybrid approach-continuous SpHb monitoring triggering confirmatory HemoCue® measurement-may support safer perioperative hemoglobin management.</p>","PeriodicalId":15513,"journal":{"name":"Journal of Clinical Monitoring and Computing","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2026-10-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Monitoring and Computing","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10877-026-01510-3","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

To evaluate the perioperative performance of two minimally/non-invasive hemoglobin monitoring modalities-continuous pulse hemoglobin (SpHb) (Revision O: rev. O) and intermittent point-of-care testing (POCT) using HemoCue® Hb801-against an automated hematology analyzer as the reference standard. Fifty patients undergoing general anesthesia with invasive arterial monitoring for surgery with anticipated blood loss were enrolled prospectively. A total of 281 arterial blood samples were analyzed by an automated hematology analyzer (reference standard), a blood gas analyzer, and HemoCue® Hb801; all comparisons, including HemoCue®, used arterial rather than capillary blood. SpHb (rev. O) was monitored continuously. Correlation and Bland-Altman analyses were performed, and polar plot analysis of changes between successive sampling points assessed SpHb trend-tracking. HemoCue® Hb801 agreed closely with the reference (r = 0.979; bias 0.119 g/dL; limits of agreement [LOA] -0.687 to +0.925 g/dL). SpHb correlated with the reference (r = 0.777) but showed wide LOA (- 2.114 to +2.811 g/dL), proportional bias, and greater random error at perfusion index <1.4 (87/281 measurements, 31.0%). In polar plot analysis of 23 successive-change pairs from 9 patients, angular bias was - 5.8° (95% CI -12.2° to +0.5°) and four-quadrant concordance 100%, but radial limits of agreement (- 36.3° to +24.6°) exceeded the ±30° criterion. HemoCue® Hb801 provided accurate Hb measurement from a minimal arterial sample. SpHb (rev. O) had limited absolute accuracy and could not quantify the magnitude of change within accepted limits, but reliably indicated its direction. A hybrid approach-continuous SpHb monitoring triggering confirmatory HemoCue® measurement-may support safer perioperative hemoglobin management.

围手术期微创血红蛋白监测的准确性:经皮血红蛋白监测和即时检测设备:前瞻性观察研究。
以自动血液分析仪作为参考标准,评估两种微创/非侵入性血红蛋白监测模式——连续脉冲血红蛋白(SpHb) (Revision O: rev. O)和使用HemoCue®hb801的间歇点护理检测(POCT)的围手术期性能。本研究前瞻性地纳入了50例接受全身麻醉并进行有创动脉监测的手术患者。采用自动血液学分析仪(参考标准)、血气分析仪和HemoCue®Hb801对281份动脉血液样本进行分析;包括HemoCue®在内的所有比较均使用动脉血而非毛细血管血。连续监测SpHb (rev. O)。进行相关分析和Bland-Altman分析,并对连续采样点之间的变化进行极坐标图分析,评估SpHb趋势跟踪。HemoCue®Hb801与参考文献非常吻合(r = 0.979,偏倚0.119 g/dL,一致性限[LOA] -0.687 ~ +0.925 g/dL)。SpHb与参考值相关(r = 0.777),但LOA较宽(- 2.114 ~ +2.811 g/dL),存在比例偏倚,且灌注指数随机误差较大
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.30
自引率
13.60%
发文量
144
审稿时长
6-12 weeks
期刊介绍: The Journal of Clinical Monitoring and Computing is a clinical journal publishing papers related to technology in the fields of anaesthesia, intensive care medicine, emergency medicine, and peri-operative medicine. The journal has links with numerous specialist societies, including editorial board representatives from the European Society for Computing and Technology in Anaesthesia and Intensive Care (ESCTAIC), the Society for Technology in Anesthesia (STA), the Society for Complex Acute Illness (SCAI) and the NAVAt (NAVigating towards your Anaestheisa Targets) group. The journal publishes original papers, narrative and systematic reviews, technological notes, letters to the editor, editorial or commentary papers, and policy statements or guidelines from national or international societies. The journal encourages debate on published papers and technology, including letters commenting on previous publications or technological concerns. The journal occasionally publishes special issues with technological or clinical themes, or reports and abstracts from scientificmeetings. Special issues proposals should be sent to the Editor-in-Chief. Specific details of types of papers, and the clinical and technological content of papers considered within scope can be found in instructions for authors.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书