Dynamic Interfragmentary Compression Through a Tibial Intramedullary Nail Compared to Backslapping Alone: A Biomechanical Analysis.

IF 2 3区 医学 Q3 ORTHOPEDICS
Brittany E Haws, J Daniel Thompson, Jonathan P Yawman, Jana M Davis, Nina Suh, James A Blair
{"title":"Dynamic Interfragmentary Compression Through a Tibial Intramedullary Nail Compared to Backslapping Alone: A Biomechanical Analysis.","authors":"Brittany E Haws, J Daniel Thompson, Jonathan P Yawman, Jana M Davis, Nina Suh, James A Blair","doi":"10.1097/BOT.0000000000003293","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To investigate the amount of interfragmentary compression obtained in a transverse tibia fracture model with backslapping alone (BS) versus the intramedullary nail's dynamic internal compression (IC) mechanism and if that compression is maintained after loading.</p><p><strong>Methods: </strong>A transverse diaphyseal tibia synthetic fracture model was used. Specimens were allocated to four groups (n=5/group) that received one of two intramedullary nails (A or B) and obtained compression either with the IC device or with BS alone. Specimens underwent sinusoidal loading for 135,000 cycles of 700N @3Hz to simulate three months of weightbearing. Interfragmentary pressure measurements were measured at initial compression, procedure completion (pre-cycling) and after loading (post-cycling). Compression at each stage was compared between BS and IC groups as well as between stages. Outcomes were also compared between Nails A and B.</p><p><strong>Results: </strong>There was significantly greater initial (1097N vs 167N, p<0.001) and pre-cycling (391N vs 103N, p<0.001) compression generated with IC compared to BS. The IC group also maintained greater compression than the BS group post-cycling (377N vs 75N, p<0.001). Both groups experienced significant loss of compression between initial and pre-cycling timepoints and during the cycling process (p<0.05 for all). On subanalysis, Nail B was able to generate significantly more initial compression than Nail A with the IC device (1523N vs 382N, p=0.001) and maintain it post-cycling (412N vs 205N, p=0.032). Conversely, Nail A was able to generate and maintain significantly more compression than Nail B at all three time-points with the BS technique (p<0.05 for all).</p><p><strong>Conclusions: </strong>Intramedullary IC devices generated and maintained greater interfragmentary fracture compression than traditional BS techniques alone. However, some compressive force was lost after loading regardless of compression technique utilized. Differences in compressive force were observed between nail manufacturers with both techniques, which may suggest that nail geometry and instrumentation may contribute to fracture compression.</p><p><strong>Level of evidence: </strong>V.</p>","PeriodicalId":16644,"journal":{"name":"Journal of Orthopaedic Trauma","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-10-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Trauma","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/BOT.0000000000003293","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To investigate the amount of interfragmentary compression obtained in a transverse tibia fracture model with backslapping alone (BS) versus the intramedullary nail's dynamic internal compression (IC) mechanism and if that compression is maintained after loading.

Methods: A transverse diaphyseal tibia synthetic fracture model was used. Specimens were allocated to four groups (n=5/group) that received one of two intramedullary nails (A or B) and obtained compression either with the IC device or with BS alone. Specimens underwent sinusoidal loading for 135,000 cycles of 700N @3Hz to simulate three months of weightbearing. Interfragmentary pressure measurements were measured at initial compression, procedure completion (pre-cycling) and after loading (post-cycling). Compression at each stage was compared between BS and IC groups as well as between stages. Outcomes were also compared between Nails A and B.

Results: There was significantly greater initial (1097N vs 167N, p<0.001) and pre-cycling (391N vs 103N, p<0.001) compression generated with IC compared to BS. The IC group also maintained greater compression than the BS group post-cycling (377N vs 75N, p<0.001). Both groups experienced significant loss of compression between initial and pre-cycling timepoints and during the cycling process (p<0.05 for all). On subanalysis, Nail B was able to generate significantly more initial compression than Nail A with the IC device (1523N vs 382N, p=0.001) and maintain it post-cycling (412N vs 205N, p=0.032). Conversely, Nail A was able to generate and maintain significantly more compression than Nail B at all three time-points with the BS technique (p<0.05 for all).

Conclusions: Intramedullary IC devices generated and maintained greater interfragmentary fracture compression than traditional BS techniques alone. However, some compressive force was lost after loading regardless of compression technique utilized. Differences in compressive force were observed between nail manufacturers with both techniques, which may suggest that nail geometry and instrumentation may contribute to fracture compression.

Level of evidence: V.

通过胫骨髓内钉进行动态碎片间压缩与单纯拍打:生物力学分析。
目的:研究单靠拍打(BS)与髓内钉动态内压缩(IC)机制在胫骨横骨折模型中获得的碎片间压缩量,以及加载后是否保持这种压缩。方法:采用胫骨骨干横向合成骨折模型。标本被分为四组(n=5/组),分别接受两种髓内钉(A或B)中的一种,并使用IC装置或单独使用BS进行压迫。试样进行了135,000次700N @3Hz的正弦加载,以模拟三个月的负重。在初始压缩、程序完成(预循环)和加载(后循环)时测量碎片间压力。比较BS组和IC组各阶段以及各阶段的压缩情况。结果:初始1097N比167N明显更大。结论:髓内IC装置产生并维持的骨折碎片间压迫比传统BS技术更大。然而,无论采用何种压缩技术,加载后都会损失一些压缩力。采用这两种技术的钉制造商观察到不同的压缩力,这可能表明钉的几何形状和内固定可能导致骨折压缩。证据等级:V。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Orthopaedic Trauma
Journal of Orthopaedic Trauma 医学-运动科学
CiteScore
3.90
自引率
8.70%
发文量
396
审稿时长
3-8 weeks
期刊介绍: Journal of Orthopaedic Trauma is devoted exclusively to the diagnosis and management of hard and soft tissue trauma, including injuries to bone, muscle, ligament, and tendons, as well as spinal cord injuries. Under the guidance of a distinguished international board of editors, the journal provides the most current information on diagnostic techniques, new and improved surgical instruments and procedures, surgical implants and prosthetic devices, bioplastics and biometals; and physical therapy and rehabilitation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书