{"title":"Improving deep-learning IMRT plan quality for lymphoma using deliverability-aware training","authors":"Denis Kutnár, Ivan Richter Vogelius, Hristo Atanasov Georgiev, Lena Specht, Jens Petersen","doi":"10.1002/acm2.70815","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Deep learning (DL) methods have shown strong performance in predicting dose distributions from patient anatomy and target structures; however, accurate dose prediction does not guarantee a deliverable plan after conversion into machine parameters and dose recalculation in the treatment planning system (TPS), motivating training objectives that explicitly couple prediction to deliverability. Lymphoma presents substantial clinical heterogeneity in target location, extent, and prescription/fractionation, so fast automated methods that produce deliverable plans are particularly valuable.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To investigate whether a deliverability-aware training framework improves the end-to-end deliverability of DL-generated intensity-modulated radiotherapy (IMRT) plans for lymphoma.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A retrospective single-institution dataset of 619 plans (571 patients; 2009–2021) was used. Clinical volumetric modulated arc therapy (VMAT) plans were converted to standardized 15-field coplanar sliding-window IMRT reference plans in Eclipse using automated scripting. Two DL approaches were compared: a baseline dose-only objective model (DL-DO) and a multi-objective model (DL-MO) trained with dose supervision, fluence supervision, and a dose–fluence–dose cycle-consistency constraint implemented via fixed inverse and forward networks. Predicted fluence maps were imported into the TPS, leaf-sequenced, and recalculated with AcurosXB without further optimization; plan quality was evaluated on a held-out test set (<i>n</i> = 61) using planning target volume (PTV) <i>D</i><sub>98</sub>, <i>D</i><sub>2</sub>, and <i>D</i><sub>max</sub> with paired Wilcoxon testing and Holm correction.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>DL-MO produced significantly better machine-deliverable plans than DL-DO, improving coverage and reducing hotspots (median paired differences: +12.31 percentage points (pp) in <i>D</i><sub>98</sub>, −3.43 pp in <i>D</i><sub>2</sub>, and −3.18 pp in <i>D</i><sub>max</sub>; Holm-corrected <i>p</i> < 0.0001 for all). Mean ± SD PTV metrics were <i>D</i><sub>98</sub> = 90.10 ± 3.74% vs. 78.59 ± 5.92%, <i>D</i><sub>2</sub> = 106.28 ± 1.13% vs. 109.65 ± 2.40%, and <i>D</i><sub>max</sub> = 110.32 ± 2.20% vs. 113.27 ± 3.03% (DL-MO vs. DL-DO). DL-MO approached the TPS reference IMRT plans (<i>D</i><sub>98</sub> = 91.62 ± 3.45%) with residual hotspot elevation.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Cycle-consistency training improves the end-to-end deliverability of DL-generated IMRT plans for heterogeneous lymphoma, yielding substantially better TPS-recalculated, machine-deliverable plans than conventional dose-only supervision while maintaining sub-second fluence-map generation. However, statistically significant differences from the TPS reference remained for <i>D</i><sub>98</sub>, <i>D</i><sub>2</sub>, and <i>D</i><sub>max</sub>.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13639240/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70815","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Deep learning (DL) methods have shown strong performance in predicting dose distributions from patient anatomy and target structures; however, accurate dose prediction does not guarantee a deliverable plan after conversion into machine parameters and dose recalculation in the treatment planning system (TPS), motivating training objectives that explicitly couple prediction to deliverability. Lymphoma presents substantial clinical heterogeneity in target location, extent, and prescription/fractionation, so fast automated methods that produce deliverable plans are particularly valuable.
Purpose
To investigate whether a deliverability-aware training framework improves the end-to-end deliverability of DL-generated intensity-modulated radiotherapy (IMRT) plans for lymphoma.
Methods
A retrospective single-institution dataset of 619 plans (571 patients; 2009–2021) was used. Clinical volumetric modulated arc therapy (VMAT) plans were converted to standardized 15-field coplanar sliding-window IMRT reference plans in Eclipse using automated scripting. Two DL approaches were compared: a baseline dose-only objective model (DL-DO) and a multi-objective model (DL-MO) trained with dose supervision, fluence supervision, and a dose–fluence–dose cycle-consistency constraint implemented via fixed inverse and forward networks. Predicted fluence maps were imported into the TPS, leaf-sequenced, and recalculated with AcurosXB without further optimization; plan quality was evaluated on a held-out test set (n = 61) using planning target volume (PTV) D98, D2, and Dmax with paired Wilcoxon testing and Holm correction.
Results
DL-MO produced significantly better machine-deliverable plans than DL-DO, improving coverage and reducing hotspots (median paired differences: +12.31 percentage points (pp) in D98, −3.43 pp in D2, and −3.18 pp in Dmax; Holm-corrected p < 0.0001 for all). Mean ± SD PTV metrics were D98 = 90.10 ± 3.74% vs. 78.59 ± 5.92%, D2 = 106.28 ± 1.13% vs. 109.65 ± 2.40%, and Dmax = 110.32 ± 2.20% vs. 113.27 ± 3.03% (DL-MO vs. DL-DO). DL-MO approached the TPS reference IMRT plans (D98 = 91.62 ± 3.45%) with residual hotspot elevation.
Conclusions
Cycle-consistency training improves the end-to-end deliverability of DL-generated IMRT plans for heterogeneous lymphoma, yielding substantially better TPS-recalculated, machine-deliverable plans than conventional dose-only supervision while maintaining sub-second fluence-map generation. However, statistically significant differences from the TPS reference remained for D98, D2, and Dmax.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic