Characterization of animal tissues for Monte Carlo simulations in ion beam therapy

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Liu Xiaoli, Michael F. Moyers
{"title":"Characterization of animal tissues for Monte Carlo simulations in ion beam therapy","authors":"Liu Xiaoli,&nbsp;Michael F. Moyers","doi":"10.1002/acm2.70836","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To obtain better elemental compositions (ECs), physical densities (PDs), and relative linear stopping powers (RLSPs) of tissues scanned by single or dual energy computed tomography for use by Monte Carlo (MC) simulations of light ion beam therapy.</p>\n </section>\n \n <section>\n \n <h3> Methods and materials</h3>\n \n <p>The RLSPs of 12 animal tissue samples were measured using several different energy proton and carbon ion beams. The EC of each sample was analyzed using the following modern laboratory techniques: Inductively coupled plasma optical emission spectrometry (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS), organic elemental analysis (OEA), and ion chromatography (IC). RLSPs were calculated from both measured ECs and literature ECs and then compared with measured RLSPs. All samples underwent single-energy CT (SECT) and dual-energy CT (DECT) scanning after which a dual-energy index (DEI) was derived. A mapping from computed tomography number (CTN) and DEI to PD was performed.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>For each tissue sample, the measured RLSPs showed no significant dependence on beam energy or ion species, with all values agreeing to within ± 1% except for lung (within ± 2%). RLSPs directly measured and those calculated using measured PDs and ECs agreed within ± 2% except for lung. Measured RLSPs were slightly closer to values from laboratory-determined ECs than to literature ECs.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>RLSP measurements were independent of beam energy and ion species. The main determinants of accurate CTN to RLSP conversion are the PDs and fractional hydrogen content of the tissues. The accuracy of PD and EC for input into Monte Carlo transport calculations can be improved by combining DEI and scaled CTN.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13639258/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70836","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

To obtain better elemental compositions (ECs), physical densities (PDs), and relative linear stopping powers (RLSPs) of tissues scanned by single or dual energy computed tomography for use by Monte Carlo (MC) simulations of light ion beam therapy.

Methods and materials

The RLSPs of 12 animal tissue samples were measured using several different energy proton and carbon ion beams. The EC of each sample was analyzed using the following modern laboratory techniques: Inductively coupled plasma optical emission spectrometry (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS), organic elemental analysis (OEA), and ion chromatography (IC). RLSPs were calculated from both measured ECs and literature ECs and then compared with measured RLSPs. All samples underwent single-energy CT (SECT) and dual-energy CT (DECT) scanning after which a dual-energy index (DEI) was derived. A mapping from computed tomography number (CTN) and DEI to PD was performed.

Results

For each tissue sample, the measured RLSPs showed no significant dependence on beam energy or ion species, with all values agreeing to within ± 1% except for lung (within ± 2%). RLSPs directly measured and those calculated using measured PDs and ECs agreed within ± 2% except for lung. Measured RLSPs were slightly closer to values from laboratory-determined ECs than to literature ECs.

Conclusions

RLSP measurements were independent of beam energy and ion species. The main determinants of accurate CTN to RLSP conversion are the PDs and fractional hydrogen content of the tissues. The accuracy of PD and EC for input into Monte Carlo transport calculations can be improved by combining DEI and scaled CTN.

离子束治疗中用于蒙特卡罗模拟的动物组织表征。
目的:通过单能量或双能量计算机断层扫描获得更好的组织元素组成(ECs)、物理密度(pd)和相对线性停止功率(RLSPs),用于光离子束治疗的蒙特卡罗(MC)模拟。方法与材料:采用不同能量的质子和碳离子束对12种动物组织样品进行rlsp测定。采用电感耦合等离子体发射光谱法(ICP-OES)、电感耦合等离子体质谱法(ICP-MS)、有机元素分析(OEA)和离子色谱法(IC)等现代实验室技术对样品的EC进行分析。根据测量的ECs和文献ECs计算RLSPs,然后与测量的RLSPs进行比较。所有样本分别进行单能CT (SECT)和双能CT (DECT)扫描,并得出双能指数(DEI)。从计算机断层扫描数(CTN)和DEI到PD进行映射。结果:对于每个组织样本,测量的RLSPs对光束能量或离子种类没有显着依赖,除肺外,所有值都在±1%以内(在±2%以内)。除肺外,直接测量的RLSPs和使用测量的pd和ECs计算的RLSPs在±2%范围内一致。测量的RLSPs与实验室测定的ECs值比文献中的ECs值更接近。结论:RLSP测量结果与束流能量和离子种类无关。准确的CTN到RLSP转换的主要决定因素是组织的pd和分数氢含量。结合DEI和缩放CTN,可以提高输入到蒙特卡罗输运计算中的PD和EC的精度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书