{"title":"Radiographic accuracy of implant positioning in unicompartmental knee arthroplasty, does an imageless robotic system make a difference?","authors":"Alice Montagna, Luca Andriollo, Virginia Cinelli, Marina Marescalchi, Luca Matascioli, Francesco Benazzo, Stefano Marco Paolo Rossi","doi":"10.1007/s00402-026-06499-7","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Robotic-assisted technologies and evolving implant designs have been introduced in unicompartmental knee arthroplasty (UKA) to improve the accuracy and reproducibility of component positioning. However, the relative contribution of robotic assistance compared with conventional instrumentation to radiographic outcomes in experienced, high-volume centers remains unclear.</p><h3>Methods</h3><p>This retrospective comparative study included 96 consecutive medial UKAs performed at a single high-volume institution. Patients were stratified into three groups: Journey I UKA implanted with NAVIO imageless robotic assistance (<i>n</i> = 38), Journey II Uni UKA implanted using conventional instrumentation (<i>n</i> = 27), and Journey II UK UKA implanted with CORI imageless robotic assistance (<i>n</i> = 31). Standardized standing long-leg weight-bearing radiographs were obtained preoperatively and postoperatively. Radiographic evaluation included hip–knee–ankle (HKA) angle, femoral component coronal angle (FCCA), tibial component coronal angle (TCCA), and sagittal alignment parameters. Tibial coronal alignment was interpreted according to the Cartier metaphyseal alignment philosophy. Intergroup comparisons were performed using appropriate statistical tests.</p><h3>Results</h3><p>Baseline demographic characteristics and preoperative radiographic parameters were comparable among the three groups (all <i>p</i> > 0.05). Postoperative mean HKA was 177.1° ± 2.1° in the Journey I NAVIO group, 177.3° ± 2.5° in the Journey II Uni conventional group, and 177.7° ± 2.2° in the Journey II UK CORI group, with no statistically significant differences (<i>p</i> = 0.56). Femoral and tibial component coronal alignment, including TCCA, showed no significant intergroup differences. All cases across the three cohorts were within predefined safe zones for HKA, FCCA, and TCCA. Sagittal alignment parameters were also comparable between groups.</p><h3>Conclusion</h3><p>In a high-volume center adopting a consistent alignment philosophy, medial UKA achieves equivalent and highly accurate radiographic outcomes regardless of robotic assistance or conventional instrumentation.</p></div>","PeriodicalId":8326,"journal":{"name":"Archives of Orthopaedic and Trauma Surgery","volume":"146 1","pages":""},"PeriodicalIF":2.2000,"publicationDate":"2026-10-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13634016/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of Orthopaedic and Trauma Surgery","FirstCategoryId":"3","ListUrlMain":"https://link.springer.com/article/10.1007/s00402-026-06499-7","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Robotic-assisted technologies and evolving implant designs have been introduced in unicompartmental knee arthroplasty (UKA) to improve the accuracy and reproducibility of component positioning. However, the relative contribution of robotic assistance compared with conventional instrumentation to radiographic outcomes in experienced, high-volume centers remains unclear.
Methods
This retrospective comparative study included 96 consecutive medial UKAs performed at a single high-volume institution. Patients were stratified into three groups: Journey I UKA implanted with NAVIO imageless robotic assistance (n = 38), Journey II Uni UKA implanted using conventional instrumentation (n = 27), and Journey II UK UKA implanted with CORI imageless robotic assistance (n = 31). Standardized standing long-leg weight-bearing radiographs were obtained preoperatively and postoperatively. Radiographic evaluation included hip–knee–ankle (HKA) angle, femoral component coronal angle (FCCA), tibial component coronal angle (TCCA), and sagittal alignment parameters. Tibial coronal alignment was interpreted according to the Cartier metaphyseal alignment philosophy. Intergroup comparisons were performed using appropriate statistical tests.
Results
Baseline demographic characteristics and preoperative radiographic parameters were comparable among the three groups (all p > 0.05). Postoperative mean HKA was 177.1° ± 2.1° in the Journey I NAVIO group, 177.3° ± 2.5° in the Journey II Uni conventional group, and 177.7° ± 2.2° in the Journey II UK CORI group, with no statistically significant differences (p = 0.56). Femoral and tibial component coronal alignment, including TCCA, showed no significant intergroup differences. All cases across the three cohorts were within predefined safe zones for HKA, FCCA, and TCCA. Sagittal alignment parameters were also comparable between groups.
Conclusion
In a high-volume center adopting a consistent alignment philosophy, medial UKA achieves equivalent and highly accurate radiographic outcomes regardless of robotic assistance or conventional instrumentation.
背景:机器人辅助技术和不断发展的植入物设计已经被引入到单室膝关节置换术(UKA)中,以提高部件定位的准确性和可重复性。然而,在经验丰富的大容量中心,与传统仪器相比,机器人辅助对放射成像结果的相对贡献尚不清楚。方法:这项回顾性比较研究包括在一家大容量机构连续进行的96次医学uka。将患者分为三组:Journey I UKA植入NAVIO无图像机器人辅助(n = 38), Journey II Uni UKA植入常规仪器(n = 27), Journey II UK UKA植入CORI无图像机器人辅助(n = 31)。术前和术后均获得标准化站立长腿负重x线片。影像学评价包括髋关节-膝关节-踝关节(HKA)角、股骨成分冠状角(FCCA)、胫骨成分冠状角(TCCA)和矢状面对齐参数。胫骨冠状位对准按照卡地亚干骺端对准原理进行解释。采用适当的统计检验进行组间比较。结果:三组患者的基线人口学特征和术前影像学参数具有可比性(p < 0.05)。Journey I NAVIO组术后平均HKA为177.1°±2.1°,Journey II Uni常规组为177.3°±2.5°,Journey II UK CORI组为177.7°±2.2°,差异无统计学意义(p = 0.56)。包括TCCA在内的股骨和胫骨成分冠状位对齐在组间无显著差异。三个队列的所有病例都在HKA、FCCA和TCCA预先确定的安全区域内。矢状面对齐参数在两组间也具有可比性。结论:在采用一致对齐理念的大容量中心,无论机器人辅助还是传统仪器,内侧UKA都能获得同等且高度准确的放射成像结果。
期刊介绍:
"Archives of Orthopaedic and Trauma Surgery" is a rich source of instruction and information for physicians in clinical practice and research in the extensive field of orthopaedics and traumatology. The journal publishes papers that deal with diseases and injuries of the musculoskeletal system from all fields and aspects of medicine. The journal is particularly interested in papers that satisfy the information needs of orthopaedic clinicians and practitioners. The journal places special emphasis on clinical relevance.