Kota Hachiya, Hiroki Abiko, Koichi Chida, Yoshikazu Matsuda
{"title":"Transition from a detector-dose-based to a CNR-based cerebral angiography system: evaluation of image quality and radiation dose.","authors":"Kota Hachiya, Hiroki Abiko, Koichi Chida, Yoshikazu Matsuda","doi":"10.1007/s12194-026-01141-y","DOIUrl":null,"url":null,"abstract":"<p><p>To evaluate the image quality and clinical radiation dose of a cerebral angiography system featuring contrast-to-noise ratio (CNR)-based exposure control for neurovascular endovascular treatment (EVT). Image quality was evaluated using a quality control phantom under fluoroscopy and digital subtraction angiography (DSA). Clinical radiation dose was retrospectively compared between a conventional angiography system using detector dose-based exposure control (61 cerebral aneurysm EVT procedures) and a new angiography system using CNR-based exposure control (99 procedures). Patient entrance reference air kerma (K<sub>a,r</sub>) and air kerma-area product (P<sub>KA</sub>) were analyzed. The new system demonstrated significantly higher observer scores than the conventional system for both fluoroscopy and DSA, including spatial resolution, low-contrast resolution, and dynamic range. The median total K<sub>a,r</sub> was 1818 [1301-2744] mGy with the new system and 2283 [1580-3184] mGy with the conventional system, representing a 20% reduction (p < 0.05). Conversely, the median total P<sub>KA</sub> increased by 26% (173 [120-237] vs. 137 [93-207] Gy·cm<sup>2</sup>, p < 0.05), but remained below the current Japanese diagnostic reference level for cerebral aneurysm EVT. Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total K<sub>a,r</sub> during cerebral aneurysm EVT. Although the total P<sub>KA</sub> increased by 26%, this increase may have reflected differences in the magnification strategy between the two systems, suggesting a need for further optimization.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2026-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01141-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
To evaluate the image quality and clinical radiation dose of a cerebral angiography system featuring contrast-to-noise ratio (CNR)-based exposure control for neurovascular endovascular treatment (EVT). Image quality was evaluated using a quality control phantom under fluoroscopy and digital subtraction angiography (DSA). Clinical radiation dose was retrospectively compared between a conventional angiography system using detector dose-based exposure control (61 cerebral aneurysm EVT procedures) and a new angiography system using CNR-based exposure control (99 procedures). Patient entrance reference air kerma (Ka,r) and air kerma-area product (PKA) were analyzed. The new system demonstrated significantly higher observer scores than the conventional system for both fluoroscopy and DSA, including spatial resolution, low-contrast resolution, and dynamic range. The median total Ka,r was 1818 [1301-2744] mGy with the new system and 2283 [1580-3184] mGy with the conventional system, representing a 20% reduction (p < 0.05). Conversely, the median total PKA increased by 26% (173 [120-237] vs. 137 [93-207] Gy·cm2, p < 0.05), but remained below the current Japanese diagnostic reference level for cerebral aneurysm EVT. Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total Ka,r during cerebral aneurysm EVT. Although the total PKA increased by 26%, this increase may have reflected differences in the magnification strategy between the two systems, suggesting a need for further optimization.
期刊介绍:
The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.