Detection and correction of rigid patient motion in SPECT with exponential data consistency conditions.

IF 3.1 3区 医学 Q2 ENGINEERING, BIOMEDICAL
My Hoang Hoa Bui, Antoine Robert, Jean-Noël Badel, Ane Etxebeste, Simon Rit
{"title":"Detection and correction of rigid patient motion in SPECT with exponential data consistency conditions.","authors":"My Hoang Hoa Bui, Antoine Robert, Jean-Noël Badel, Ane Etxebeste, Simon Rit","doi":"10.1088/1361-6560/aeadeb","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective</i>. Rigid patient motion during single photon emission computed tomography (SPECT) can introduce artifacts in the reconstructed images. Patient translation can be estimated using various data-driven methods, e.g. those based on exponential data consistency conditions (eDCCs). eDCCs are mathematical equations that characterize the small redundancy between pairs of exponential projections derived from the application of the attenuated Radon transform. However, rigid motion typically involves both translational and rotational components. This study aimed to develop and evaluate an eDCCs-based approach for detecting and estimating any rigid patient motion during a SPECT acquisition.<i>Approach</i>. We calculated exponential projections from SPECT measurements, assuming that the photon emission was confined to a known convex region<i>K</i>with constant attenuation, and evaluated their pairwise consistency with eDCCs. The rigid motion was characterized by six parameters, three translations and three rotations, and the index indicating when the motion occurred. The six parameters and the motion index were estimated by minimizing the inconsistency between pairs of exponential projections after compensating for the estimated rigid motion in the projection domain. The method was evaluated on several datasets of a simulated liver patient and on experimentally acquired data of a NEMA IEC phantom.<i>Main results</i>. The proposed method accurately estimated the ground truth motion parameters when the SPECT projections were ideally modeled by the attenuated Radon transform, but was affected by degrading physical effects, mainly by the system spatial resolution (intrinsic and collimator spatial resolution) and the scatter. Despite these physical effects, eDCCs recovered the source center position within a two-pixel offset and substantially improved reconstructed image quality. Specifically, the method achieved a relative activity recovery coefficient (rARC) of around 101% relative to images reconstructed from motion-free projections, compared with an rARC of 87% without correction for the studied liver patient's rigid motion occurring at the beginning of the acquisition.<i>Significance</i>. A fully data-driven method with eDCCs was able to estimate and correct for a rigid patient motion during a SPECT acquisition.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":" ","pages":""},"PeriodicalIF":3.1000,"publicationDate":"2026-10-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/aeadeb","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective. Rigid patient motion during single photon emission computed tomography (SPECT) can introduce artifacts in the reconstructed images. Patient translation can be estimated using various data-driven methods, e.g. those based on exponential data consistency conditions (eDCCs). eDCCs are mathematical equations that characterize the small redundancy between pairs of exponential projections derived from the application of the attenuated Radon transform. However, rigid motion typically involves both translational and rotational components. This study aimed to develop and evaluate an eDCCs-based approach for detecting and estimating any rigid patient motion during a SPECT acquisition.Approach. We calculated exponential projections from SPECT measurements, assuming that the photon emission was confined to a known convex regionKwith constant attenuation, and evaluated their pairwise consistency with eDCCs. The rigid motion was characterized by six parameters, three translations and three rotations, and the index indicating when the motion occurred. The six parameters and the motion index were estimated by minimizing the inconsistency between pairs of exponential projections after compensating for the estimated rigid motion in the projection domain. The method was evaluated on several datasets of a simulated liver patient and on experimentally acquired data of a NEMA IEC phantom.Main results. The proposed method accurately estimated the ground truth motion parameters when the SPECT projections were ideally modeled by the attenuated Radon transform, but was affected by degrading physical effects, mainly by the system spatial resolution (intrinsic and collimator spatial resolution) and the scatter. Despite these physical effects, eDCCs recovered the source center position within a two-pixel offset and substantially improved reconstructed image quality. Specifically, the method achieved a relative activity recovery coefficient (rARC) of around 101% relative to images reconstructed from motion-free projections, compared with an rARC of 87% without correction for the studied liver patient's rigid motion occurring at the beginning of the acquisition.Significance. A fully data-driven method with eDCCs was able to estimate and correct for a rigid patient motion during a SPECT acquisition.

指数数据一致性条件下SPECT患者刚性运动的检测与校正。
目的单光子发射计算机断层扫描(SPECT)过程中患者的刚性运动会在重建图像中引入伪影。患者翻译可以使用各种数据驱动的方法进行估计,例如,基于指数数据一致性条件(edcc)的方法。edcc是一种数学方程,其特征是应用衰减Radon变换得到的指数投影对之间的小冗余。然而,刚性运动通常包括平移和旋转分量。本研究旨在开发和评估一种基于edcc的方法,用于在SPECT采集期间检测和估计任何僵硬的患者运动。我们计算了SPECT测量的指数投影,假设光子发射被限制在一个已知的具有恒定衰减的凸区域K中,并评估了它们与edcc的成对一致性。该刚体运动具有6个参数、3个平移和3个旋转以及指示运动发生时间的指标。在投影域对估计的刚性运动进行补偿后,通过最小化指数投影对之间的不一致性来估计六个参数和运动指标。主要结果当衰减的Radon变换对SPECT投影进行理想建模时,所提出的方法能准确地估计出地真运动参数,但受物理效应的影响,主要受系统空间分辨率(固有空间分辨率和准直空间分辨率)和散射的影响。尽管存在这些物理影响,edcc在两个像素的偏移范围内恢复了源中心位置,并大大提高了重建图像的质量。具体来说,该方法实现了相对于无运动投影重建图像的相对活动恢复系数(rARC)约为101%,而对于所研究的肝脏患者在采集开始时发生的刚性运动,未经校正的rARC为87%。具有edcc的完全数据驱动的方法能够在SPECT采集期间估计和纠正患者的刚性运动。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics in medicine and biology
Physics in medicine and biology 医学-工程:生物医学
CiteScore
6.50
自引率
14.30%
发文量
409
审稿时长
2 months
期刊介绍: The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书