Evaluation of a lead-loaded plastic scintillator for next-generation positron emission tomography detectors using Geant4 simulation.

IF 3.5 2区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Makoto Yamazaki, Masahiro Sato
{"title":"Evaluation of a lead-loaded plastic scintillator for next-generation positron emission tomography detectors using Geant4 simulation.","authors":"Makoto Yamazaki, Masahiro Sato","doi":"10.1186/s40658-026-00941-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Lead-loaded plastic scintillators are promising candidates for next-generation positron emission tomography (PET) detectors owing to their short fluorescence decay times, high machinability and low production cost. However, their γ-ray detection performance has not been fully quantified. This study evaluated the detection efficiency and energy-resolution characteristics of a 5 wt% lead-loaded plastic scintillator (EJ-256) and compared its performance with that of a conventional polyvinyltoluene (PVT) scintillator.</p><p><strong>Methods: </strong>Monte Carlo simulations were performed using Geant4. Monoenergetic 511 keV gamma rays (100 000 events) were irradiated onto scintillator blocks with thicknesses varied from 20 mm to 200 mm in 5 mm increments. The resulting absorbed-energy spectra were analysed using ROOT to determine detection efficiency and the simulated full width at half maximum (FWHM) of the 511-keV photopeak. To identify the thickness region at which detection efficiency saturated, a non-parametric Kruskal-Wallis test was conducted to consecutive thickness groups to determine the point beyond which differences were no longer statistically significant (p > 0.05).</p><p><strong>Results: </strong>EJ-256 exhibited a distinct 511 keV photopeak, whereas PVT showed only a broad Compton continuum. The detection efficiency increased from 0.9% at 20 mm to 2.9% at 200 mm and saturated above approximately 135 mm, consistent with the Kruskal-Wallis analysis. The simulated 511-keV photopeak width remained stable at approximately 10% FWHM across all thicknesses.</p><p><strong>Conclusion: </strong>These results indicate that EJ-256 demonstrates photon-absorption performance comparable to that of conventional inorganic PET scintillators, while offering substantially faster temporal response and reduced fabrication cost. This combination suggests strong potential for application in future time-of-flight PET detector modules.</p>","PeriodicalId":11559,"journal":{"name":"EJNMMI Physics","volume":"13 1","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13620012/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40658-026-00941-w","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Lead-loaded plastic scintillators are promising candidates for next-generation positron emission tomography (PET) detectors owing to their short fluorescence decay times, high machinability and low production cost. However, their γ-ray detection performance has not been fully quantified. This study evaluated the detection efficiency and energy-resolution characteristics of a 5 wt% lead-loaded plastic scintillator (EJ-256) and compared its performance with that of a conventional polyvinyltoluene (PVT) scintillator.

Methods: Monte Carlo simulations were performed using Geant4. Monoenergetic 511 keV gamma rays (100 000 events) were irradiated onto scintillator blocks with thicknesses varied from 20 mm to 200 mm in 5 mm increments. The resulting absorbed-energy spectra were analysed using ROOT to determine detection efficiency and the simulated full width at half maximum (FWHM) of the 511-keV photopeak. To identify the thickness region at which detection efficiency saturated, a non-parametric Kruskal-Wallis test was conducted to consecutive thickness groups to determine the point beyond which differences were no longer statistically significant (p > 0.05).

Results: EJ-256 exhibited a distinct 511 keV photopeak, whereas PVT showed only a broad Compton continuum. The detection efficiency increased from 0.9% at 20 mm to 2.9% at 200 mm and saturated above approximately 135 mm, consistent with the Kruskal-Wallis analysis. The simulated 511-keV photopeak width remained stable at approximately 10% FWHM across all thicknesses.

Conclusion: These results indicate that EJ-256 demonstrates photon-absorption performance comparable to that of conventional inorganic PET scintillators, while offering substantially faster temporal response and reduced fabrication cost. This combination suggests strong potential for application in future time-of-flight PET detector modules.

使用Geant4模拟评估下一代正电子发射断层扫描探测器的载铅塑料闪烁体。
目的:载铅塑料闪烁体具有荧光衰减时间短、可加工性高、生产成本低等优点,是下一代正电子发射断层扫描(PET)探测器的理想候选材料。然而,它们的γ射线探测性能尚未完全量化。本研究评估了5 wt%含铅塑料闪烁体(ej256)的探测效率和能量分辨特性,并将其与传统聚乙烯醇(PVT)闪烁体的性能进行了比较。方法:采用genant4进行蒙特卡罗模拟。单能511 keV伽马射线(100,000事件)以5mm的增量照射到厚度从20 mm到200 mm不等的闪烁体块上。利用ROOT对所得的吸收能谱进行了分析,以确定511-keV光峰的探测效率和模拟半峰全宽(FWHM)。为了确定检测效率达到饱和的厚度区域,对连续的厚度组进行非参数Kruskal-Wallis检验,以确定超过该点差异不再具有统计学意义(p > 0.05)。结果:ej256表现出明显的511kev光峰,而PVT仅表现出广泛的康普顿连续谱。检测效率从20 mm处的0.9%提高到200 mm处的2.9%,在约135 mm处达到饱和,与Kruskal-Wallis分析一致。模拟的511-keV光峰宽度在所有厚度上保持稳定在约10% FWHM。结论:这些结果表明,EJ-256具有与传统无机PET闪烁体相当的光子吸收性能,同时具有更快的时间响应和更低的制造成本。这种组合表明在未来飞行时间PET探测器模块中的应用潜力巨大。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
EJNMMI Physics
EJNMMI Physics Physics and Astronomy-Radiation
CiteScore
6.70
自引率
10.00%
发文量
78
审稿时长
13 weeks
期刊介绍: EJNMMI Physics is an international platform for scientists, users and adopters of nuclear medicine with a particular interest in physics matters. As a companion journal to the European Journal of Nuclear Medicine and Molecular Imaging, this journal has a multi-disciplinary approach and welcomes original materials and studies with a focus on applied physics and mathematics as well as imaging systems engineering and prototyping in nuclear medicine. This includes physics-driven approaches or algorithms supported by physics that foster early clinical adoption of nuclear medicine imaging and therapy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书