Ex Vivo Evaluation of Endothelial and Adventitial Temperature during Endovenous Laser Ablation Using a 1470-nm Radial 2ring Slim Fiber: Porcine Jugular Vein Model.

IF 0.5 Q4 PERIPHERAL VASCULAR DISEASE
Annals of vascular diseases Pub Date : 2026-01-01 Epub Date: 2026-09-25 DOI:10.3400/avd.oa.26-00017
Osamu Suzuki
{"title":"Ex Vivo Evaluation of Endothelial and Adventitial Temperature during Endovenous Laser Ablation Using a 1470-nm Radial 2ring Slim Fiber: Porcine Jugular Vein Model.","authors":"Osamu Suzuki","doi":"10.3400/avd.oa.26-00017","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Endovenous laser ablation (EVLA) using a 1470-nm radial 2-ring slim fiber is widely performed, but optimal ablation settings remain empirical. This ex vivo study aimed to evaluate the effects of ablation conditions on endothelial and adventitial temperature dynamics, providing insights into both treatment efficacy and safety.</p><p><strong>Methods: </strong>Porcine jugular veins were used in an ex vivo model. EVLA was performed using a 1470-nm diode laser with the 2-ring slim fiber under 2 conditions: 5W with a linear endovenous energy density (LEED) of 25 J/cm and 6.5W with a LEED of 65 J/cm. Endothelial and adventitial temperatures were continuously measured at 0.1-second intervals. Cumulative durations with endothelial temperature ≥70°C and adventitial temperature ≥43°C were assessed.</p><p><strong>Results: </strong>Endothelial temperatures reached ≥70°C under both conditions, with a longer duration observed at 6.5W. Cumulative adventitial exposure ≥43°C was prolonged at 6.5W, whereas it remained minimal at 5W.</p><p><strong>Conclusions: </strong>In this ex vivo model, 6.5W with a LEED of 65 J/cm tended to produce more sustained endothelial and adventitial thermal exposure, whereas 5W with a LEED of 25 J/cm tended to produce lower adventitial exposure. Although further clinical validation is required, segment-specific settings may be reasonable.</p>","PeriodicalId":7995,"journal":{"name":"Annals of vascular diseases","volume":"19 1","pages":""},"PeriodicalIF":0.5000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13616591/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of vascular diseases","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3400/avd.oa.26-00017","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/9/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: Endovenous laser ablation (EVLA) using a 1470-nm radial 2-ring slim fiber is widely performed, but optimal ablation settings remain empirical. This ex vivo study aimed to evaluate the effects of ablation conditions on endothelial and adventitial temperature dynamics, providing insights into both treatment efficacy and safety.

Methods: Porcine jugular veins were used in an ex vivo model. EVLA was performed using a 1470-nm diode laser with the 2-ring slim fiber under 2 conditions: 5W with a linear endovenous energy density (LEED) of 25 J/cm and 6.5W with a LEED of 65 J/cm. Endothelial and adventitial temperatures were continuously measured at 0.1-second intervals. Cumulative durations with endothelial temperature ≥70°C and adventitial temperature ≥43°C were assessed.

Results: Endothelial temperatures reached ≥70°C under both conditions, with a longer duration observed at 6.5W. Cumulative adventitial exposure ≥43°C was prolonged at 6.5W, whereas it remained minimal at 5W.

Conclusions: In this ex vivo model, 6.5W with a LEED of 65 J/cm tended to produce more sustained endothelial and adventitial thermal exposure, whereas 5W with a LEED of 25 J/cm tended to produce lower adventitial exposure. Although further clinical validation is required, segment-specific settings may be reasonable.

体外评价1470 nm径向2环细纤维猪颈静脉激光消融过程中内皮和外膜温度。
目的:使用1470 nm径向2环细纤维的静脉内激光消融(EVLA)被广泛应用,但最佳消融设置仍是经验经验。这项体外研究旨在评估消融条件对内皮和体外温度动力学的影响,为治疗的有效性和安全性提供见解。方法:采用猪颈静脉离体模型。采用1470 nm二极管激光器和2环细长光纤,在两种条件下进行EVLA: 5W,线性静脉内能量密度(LEED)为25 J/cm, 6.5W, LEED为65 J/cm。每隔0.1秒连续测量内皮温度和表皮温度。评估内皮温度≥70°C和内皮温度≥43°C的累积持续时间。结果:两种条件下内皮温度均达到≥70°C,且在6.5W时观察到的持续时间更长。≥43°C的累积外暴露在6.5W时延长,而在5W时保持最小。结论:在这个离体模型中,LEED为65 J/cm的6.5W倾向于产生更持久的内皮和外热暴露,而LEED为25 J/cm的5W倾向于产生更低的外热暴露。虽然需要进一步的临床验证,但针对特定节段的设置可能是合理的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of vascular diseases
Annals of vascular diseases PERIPHERAL VASCULAR DISEASE-
自引率
0.00%
发文量
41
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书