Ravindra Yaparpalvi, Dinesh K Mynampati, Madhur K Garg, Wolfgang A Tomé
{"title":"Clinical commissioning of a commercial brass GRID collimator for spatially fractionated radiotherapy.","authors":"Ravindra Yaparpalvi, Dinesh K Mynampati, Madhur K Garg, Wolfgang A Tomé","doi":"10.1016/j.meddos.2026.08.006","DOIUrl":null,"url":null,"abstract":"<p><p>Spatially Fractionated Radiation Therapy (SFRT) delivers high-dose radiation in a GRID pattern to treat large tumors. This can be achieved using either a physical GRID block or virtual MLC-based patterns. This report presents our experience commissioning a commercial brass GRID collimator for SFRT. Commissioning was performed using a brass GRID collimator with 6 MV and 10 MV photon beams under both flattening filter (FF) and flattening filter-free (FFF) configurations. Dosimetric measurements were collected in a water phantom and compared with treatment planning system calculations. SFRT test plans were created and validated using a commercial detector array. Valley-to-peak dose ratio (VPDR) ranged from 0.20 to 0.30 (6-MV) and 0.30 to 0.37 (10-MV). The Percent depth doses were lower than those for open fields at all beam energies. Measured vs. calculated profiles agreed within ±5% in dose and 1 mm in distance for all beam energies. TPS-generated plans showed strong agreement with measured distributions, with an average gamma passing rate of 96.5% (3%/3 mm criteria). We successfully commissioned and validated a commercial GRID collimator for SFRT using 6-MV and 10-MV photon beams in both FF and FFF modes. Despite its substantial weight and limited maximum field size of 25 × 25 cm², GRID-based SFRT offers straightforward treatment planning, efficient dose delivery, and the potential to enhance tumor control through its characteristic peak-to-valley dose distribution.</p>","PeriodicalId":49837,"journal":{"name":"Medical Dosimetry","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2026-09-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Dosimetry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.meddos.2026.08.006","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Spatially Fractionated Radiation Therapy (SFRT) delivers high-dose radiation in a GRID pattern to treat large tumors. This can be achieved using either a physical GRID block or virtual MLC-based patterns. This report presents our experience commissioning a commercial brass GRID collimator for SFRT. Commissioning was performed using a brass GRID collimator with 6 MV and 10 MV photon beams under both flattening filter (FF) and flattening filter-free (FFF) configurations. Dosimetric measurements were collected in a water phantom and compared with treatment planning system calculations. SFRT test plans were created and validated using a commercial detector array. Valley-to-peak dose ratio (VPDR) ranged from 0.20 to 0.30 (6-MV) and 0.30 to 0.37 (10-MV). The Percent depth doses were lower than those for open fields at all beam energies. Measured vs. calculated profiles agreed within ±5% in dose and 1 mm in distance for all beam energies. TPS-generated plans showed strong agreement with measured distributions, with an average gamma passing rate of 96.5% (3%/3 mm criteria). We successfully commissioned and validated a commercial GRID collimator for SFRT using 6-MV and 10-MV photon beams in both FF and FFF modes. Despite its substantial weight and limited maximum field size of 25 × 25 cm², GRID-based SFRT offers straightforward treatment planning, efficient dose delivery, and the potential to enhance tumor control through its characteristic peak-to-valley dose distribution.
期刊介绍:
Medical Dosimetry, the official journal of the American Association of Medical Dosimetrists, is the key source of information on new developments for the medical dosimetrist. Practical and comprehensive in coverage, the journal features original contributions and review articles by medical dosimetrists, oncologists, physicists, and radiation therapy technologists on clinical applications and techniques of external beam, interstitial, intracavitary and intraluminal irradiation in cancer management. Articles dealing primarily with physics will be reviewed by a specially appointed team of experts in the field.