Maskless head and neck radiotherapy using surface-guided radiation therapy: initial clinical feasibility of real-time setup, intrafraction monitoring, and automatic beam-hold

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Adi Robinson, Shravan Kandula
{"title":"Maskless head and neck radiotherapy using surface-guided radiation therapy: initial clinical feasibility of real-time setup, intrafraction monitoring, and automatic beam-hold","authors":"Adi Robinson,&nbsp;Shravan Kandula","doi":"10.1002/acm2.70809","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Thermoplastic immobilization remains standard for head and neck radiation therapy, but rigid face masks may produce clinically meaningful anxiety, claustrophobia, and treatment intolerance in some patients. Surface-guided radiation therapy (SGRT) provides non-ionizing surface setup guidance and continuous intrafraction motion monitoring, potentially enabling reduced-immobilization treatment approaches.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To describe and initially evaluate the clinical feasibility of a thermoplastic-maskless head and neck radiation therapy workflow using a posterior dorsal shell, SGRT-guided setup, cone-beam computed tomography (CBCT) verification, and real-time intrafraction monitoring with automatic beam-hold.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Ten patients with head and neck cancer were treated without a thermoplastic face mask. Patients were positioned in a posterior dorsal shell. Daily setup began with SGRT followed by CBCT verification and couch correction. During delivery, SGRT (AlignRT v6.3, Vision RT, London, UK) monitored head position in six degrees of freedom (6DoF); automatic beam-hold was triggered when translational displacement exceeded 2 mm or rotational displacement exceeded 2°. Post-treatment CBCT and SGRT log files were reviewed for every fraction to characterize residual displacement, threshold events, and repositioning. Intrafraction motion and residual displacement were summarized as medians with interquartile range (IQR).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Across 330 delivered fractions, sgrt-recorded intrafraction motion was small:: per-axis signed real-time deltas had a median [IQR] of 0.10 mm [−0.15, 0.30] translationally and 0.10° [−0.05, 0.20] rotationally. No fraction had a mean real-time delta exceeding the ± 2 mm/ ± 2° tolerance, and automatic beam-hold interruptions occurred in approximately 2% of fractions (7 of 330); no fraction was abandoned. Per-axis post-treatment CBCT residuals were −0.01 mm [−0.04, 0.05] translationally and 0.15° [−0.40, 0.55] rotationally. All patients completed the maskless course; 3 of 10 had previously declined or been unable to tolerate mask-based treatment, and all reported the maskless workflow to be well tolerated on an end-of-course questionnaire.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>A thermoplastic-maskless head and neck workflow using SGRT, CBCT verification, predefined beam-hold thresholds, and dorsal shell support was feasible in this initial 10-patient experience, maintaining small observed intrafraction motion while directly addressing mask intolerance. Larger prospective evaluation with validated patient-reported outcomes, dosimetric analysis, and multi-institutional QA is warranted.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70809","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70809","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Thermoplastic immobilization remains standard for head and neck radiation therapy, but rigid face masks may produce clinically meaningful anxiety, claustrophobia, and treatment intolerance in some patients. Surface-guided radiation therapy (SGRT) provides non-ionizing surface setup guidance and continuous intrafraction motion monitoring, potentially enabling reduced-immobilization treatment approaches.

Purpose

To describe and initially evaluate the clinical feasibility of a thermoplastic-maskless head and neck radiation therapy workflow using a posterior dorsal shell, SGRT-guided setup, cone-beam computed tomography (CBCT) verification, and real-time intrafraction monitoring with automatic beam-hold.

Methods

Ten patients with head and neck cancer were treated without a thermoplastic face mask. Patients were positioned in a posterior dorsal shell. Daily setup began with SGRT followed by CBCT verification and couch correction. During delivery, SGRT (AlignRT v6.3, Vision RT, London, UK) monitored head position in six degrees of freedom (6DoF); automatic beam-hold was triggered when translational displacement exceeded 2 mm or rotational displacement exceeded 2°. Post-treatment CBCT and SGRT log files were reviewed for every fraction to characterize residual displacement, threshold events, and repositioning. Intrafraction motion and residual displacement were summarized as medians with interquartile range (IQR).

Results

Across 330 delivered fractions, sgrt-recorded intrafraction motion was small:: per-axis signed real-time deltas had a median [IQR] of 0.10 mm [−0.15, 0.30] translationally and 0.10° [−0.05, 0.20] rotationally. No fraction had a mean real-time delta exceeding the ± 2 mm/ ± 2° tolerance, and automatic beam-hold interruptions occurred in approximately 2% of fractions (7 of 330); no fraction was abandoned. Per-axis post-treatment CBCT residuals were −0.01 mm [−0.04, 0.05] translationally and 0.15° [−0.40, 0.55] rotationally. All patients completed the maskless course; 3 of 10 had previously declined or been unable to tolerate mask-based treatment, and all reported the maskless workflow to be well tolerated on an end-of-course questionnaire.

Conclusions

A thermoplastic-maskless head and neck workflow using SGRT, CBCT verification, predefined beam-hold thresholds, and dorsal shell support was feasible in this initial 10-patient experience, maintaining small observed intrafraction motion while directly addressing mask intolerance. Larger prospective evaluation with validated patient-reported outcomes, dosimetric analysis, and multi-institutional QA is warranted.

使用表面引导放射治疗的无面罩头颈部放疗:实时设置、屈光内监测和自动光束保持的初步临床可行性
背景:热塑性固定仍然是头颈部放射治疗的标准,但刚性口罩可能会在一些患者中产生临床意义上的焦虑、幽闭恐惧症和治疗不耐受。表面引导放射治疗(SGRT)提供非电离的表面设置指导和持续的屈光内运动监测,有可能实现减少固定治疗方法。目的描述并初步评估使用后背壳、sgrt引导装置、锥束计算机断层扫描(CBCT)验证和自动光束保持实时牵伸监测的热塑性无面罩头颈部放射治疗工作流程的临床可行性。方法对10例头颈癌患者进行不戴热塑性口罩的治疗。患者被定位于后背壳。日常设置从SGRT开始,然后是CBCT验证和沙发矫正。在交付过程中,SGRT (AlignRT v6.3, Vision RT,伦敦,英国)在六自由度(6DoF)监测头部位置;当平移位移超过2毫米或旋转位移超过2°时,自动保持光束。处理后的CBCT和SGRT日志文件检查了每个分数,以表征残余位移、阈值事件和重新定位。视场内运动和残余位移被总结为具有四分位间距(IQR)的中位数。在330个传递的分数中,sgrt记录的分数内运动很小:每轴签名实时delta的中位数[IQR]为平移0.10 mm[- 0.15, 0.30],旋转0.10°[- 0.05,0.20]。没有分数的平均实时增量超过±2mm /±2°公差,大约2%的分数(330个中的7个)发生了自动波束保持中断;没有丢弃任何分数。每轴处理后CBCT残差为平移- 0.01 mm[- 0.04, 0.05],旋转0.15°[- 0.40,0.55]。所有患者均完成了无口罩疗程;10名患者中有3名先前拒绝或无法忍受面罩治疗,并且在疗程结束时的问卷调查中,所有患者都报告了对无面罩工作流程的良好耐受。结论:在最初的10例患者中,使用SGRT、CBCT验证、预定义波束保持阈值和背壳支持的热塑性无面罩头颈部工作流程是可行的,在直接解决面罩不耐受的同时,保持了观察到的小的屈光内运动。更大的前瞻性评估,验证患者报告的结果,剂量分析和多机构QA是必要的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书