Hybrid geometric and evolutionary optimization framework for radiosurgery planning: isocenter selection and angle optimization for multiple brain metastases.
IF 1.6 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
{"title":"Hybrid geometric and evolutionary optimization framework for radiosurgery planning: isocenter selection and angle optimization for multiple brain metastases.","authors":"Quang Vinh Pham, Quang Trung Pham","doi":"10.1007/s12194-026-01139-6","DOIUrl":null,"url":null,"abstract":"<p><p>This study aims to develop and evaluate the performance of Collimator Angles Optimization Software version 2 (CAOS-v2) for stereotactic radiosurgery (SRS) planning in patients with multiple brain metastases. Treatment planning was performed in Eclipse v13.6 with RapidArc on the TrueBeam STx. The primary objective is to reduce the volume of normal brain tissue receiving 12 Gy (V12Gy) while maintaining overall plan quality. CAOS-v2 determines the isocenter location using the Smoothed Weiszfeld Algorithm to compute the geometric median of the target centers, in contrast to traditional centroid-based placement. The software subsequently employs a genetic algorithm to optimize couch-collimator angle pairs, reducing the impact of the island blocking problem. Twenty-six clinical cases were retrospectively replanned using CAOS-v2 with a single-fraction prescription dose of 18 Gy applied to all treatment plans. Plans were evaluated using Conformity Index (CI), Gradient Index (GI), Gradient Measure (GM), organs at risk (OARs), and the V12Gy of healthy brain tissue. Application of CAOS-v2 resulted in a reduction of V12Gy by 17.6-38.1% compared to the conventional planning approach. The CI remained stable, indicating consistent dose uniformity across treatment plans. No increase in dose to OARs was observed. Additionally, both the GI and GM significantly decreased, indicating improved dose fall-off characteristics attributable to CAOS-v2. Integration of CAOS-v2 into the planning workflow significantly reduces normal brain dose in multi-metastatic SRS without compromising conformity or gradient performance, supporting its clinical applicability for automated treatment parameter optimization.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2026-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01139-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
This study aims to develop and evaluate the performance of Collimator Angles Optimization Software version 2 (CAOS-v2) for stereotactic radiosurgery (SRS) planning in patients with multiple brain metastases. Treatment planning was performed in Eclipse v13.6 with RapidArc on the TrueBeam STx. The primary objective is to reduce the volume of normal brain tissue receiving 12 Gy (V12Gy) while maintaining overall plan quality. CAOS-v2 determines the isocenter location using the Smoothed Weiszfeld Algorithm to compute the geometric median of the target centers, in contrast to traditional centroid-based placement. The software subsequently employs a genetic algorithm to optimize couch-collimator angle pairs, reducing the impact of the island blocking problem. Twenty-six clinical cases were retrospectively replanned using CAOS-v2 with a single-fraction prescription dose of 18 Gy applied to all treatment plans. Plans were evaluated using Conformity Index (CI), Gradient Index (GI), Gradient Measure (GM), organs at risk (OARs), and the V12Gy of healthy brain tissue. Application of CAOS-v2 resulted in a reduction of V12Gy by 17.6-38.1% compared to the conventional planning approach. The CI remained stable, indicating consistent dose uniformity across treatment plans. No increase in dose to OARs was observed. Additionally, both the GI and GM significantly decreased, indicating improved dose fall-off characteristics attributable to CAOS-v2. Integration of CAOS-v2 into the planning workflow significantly reduces normal brain dose in multi-metastatic SRS without compromising conformity or gradient performance, supporting its clinical applicability for automated treatment parameter optimization.
期刊介绍:
The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.