Bridging TG-116 and TG-232: A retrospective multi-vendor evaluation of deviation index re-centering and process control in digital radiography

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Ezzat AbuAzzah, Faisal Alrehily, Moawia Gameraddin, Sultan Alshoabi, Walaa Alsharif, Kamal Dahan Alsultan, Abdullah Alshamrani, Khaled Al-Radadi, Ramy Badawy, Samer Al-Jabri, Turki Al-Harbi, Nawaf Al-Hazme
{"title":"Bridging TG-116 and TG-232: A retrospective multi-vendor evaluation of deviation index re-centering and process control in digital radiography","authors":"Ezzat AbuAzzah,&nbsp;Faisal Alrehily,&nbsp;Moawia Gameraddin,&nbsp;Sultan Alshoabi,&nbsp;Walaa Alsharif,&nbsp;Kamal Dahan Alsultan,&nbsp;Abdullah Alshamrani,&nbsp;Khaled Al-Radadi,&nbsp;Ramy Badawy,&nbsp;Samer Al-Jabri,&nbsp;Turki Al-Harbi,&nbsp;Nawaf Al-Hazme","doi":"10.1002/acm2.70820","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>The deviation index (DI) is a vendor-neutral indicator of detector exposure relative to the configured target exposure index (TEI). Because TEI is a protocol value rather than a calibrated physical quantity, poorly aligned TEI settings can create systematic DI offsets, obscure protocol problems, and prevent meaningful statistical process control.</p>\n </section>\n \n <section>\n \n <h3> Objective</h3>\n \n <p>This retrospective multi-vendor study evaluated whether routine DICOM metadata can identify biased DI distributions, separate fixed TEI-setting offsets from protocol-adherence variation, estimate candidate TEI updates, and support TG-232-aligned standard deviation-based monitoring after DI re-centering.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>We retrospectively analyzed 38 916 digital radiography examinations performed from January 1, 2024 through August 27, 2024 using three vendor platforms at a tertiary hospital. DICOM metadata were benchmarked against published technical standards using a composite adherence index (CAI) for SID, kVp, mAs, grid use, and AEC use. A secondary gradient-boosting/SHAP analysis was retained as a non-causal variance discovery to rank the metadata patterns used to predict DI. Candidate TEI updates were estimated within the vendor-region-projection groups by algebraically re-centering the observed mean DI (<span></span><math>\n <semantics>\n <mrow>\n <mover>\n <mrow>\n <mi>D</mi>\n <mi>I</mi>\n </mrow>\n <mo>¯</mo>\n </mover>\n <mrow>\n <mo>)</mo>\n </mrow>\n </mrow>\n <annotation>$\\overline {DI} )$</annotation>\n </semantics></math> to 0.0. Robustness was assessed using sensitivity analyses restricted by automatic exposure control, beam energy, and patient habitus proxy measures.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The baseline <span></span><math>\n <semantics>\n <mover>\n <mrow>\n <mi>D</mi>\n <mi>I</mi>\n </mrow>\n <mo>¯</mo>\n </mover>\n <annotation>$\\overline {DI} $</annotation>\n </semantics></math> = −3.1 (SD 3.7), and only 11.3% of examinations were within |DI| ≤ 1. Forty of 45 vendor–region–projection groups (88.9%) had an absolute group <span></span><math>\n <semantics>\n <mover>\n <mrow>\n <mi>D</mi>\n <mi>I</mi>\n </mrow>\n <mo>¯</mo>\n </mover>\n <annotation>$\\overline {DI} $</annotation>\n </semantics></math> greater than 1.0. Separately, the mean CAI was 0.176. The secondary SHAP analysis ranked the configured TEI, manufacturer, SID, exposure, and kVp as high-contribution predictors of DI; these findings were interpreted as predictive attribution, not causal evidence. DI re-centering reduced the overall variance by 42.9% (13.5 to 7.7), increased the proportion of examinations within |DI| ≤ 1 to 34.3%, and reduced severe deviations (|DI| &gt; 3) from 60.3% to 23.1%.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>In this retrospective, multi-vendor study, the TEI setting review was operationalized as a structured, data-driven responsibility of a clinical site. By verifying protocol adherence, equipment performance, image quality, and dose indicators before centering the <span></span><math>\n <semantics>\n <mover>\n <mrow>\n <mi>D</mi>\n <mi>I</mi>\n </mrow>\n <mo>¯</mo>\n </mover>\n <annotation>$\\overline {DI} $</annotation>\n </semantics></math>, departments can support the conversion of the DI from a biased compliance signal into a practical process-control metric for exposure governance.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70820","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70820","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

The deviation index (DI) is a vendor-neutral indicator of detector exposure relative to the configured target exposure index (TEI). Because TEI is a protocol value rather than a calibrated physical quantity, poorly aligned TEI settings can create systematic DI offsets, obscure protocol problems, and prevent meaningful statistical process control.

Objective

This retrospective multi-vendor study evaluated whether routine DICOM metadata can identify biased DI distributions, separate fixed TEI-setting offsets from protocol-adherence variation, estimate candidate TEI updates, and support TG-232-aligned standard deviation-based monitoring after DI re-centering.

Methods

We retrospectively analyzed 38 916 digital radiography examinations performed from January 1, 2024 through August 27, 2024 using three vendor platforms at a tertiary hospital. DICOM metadata were benchmarked against published technical standards using a composite adherence index (CAI) for SID, kVp, mAs, grid use, and AEC use. A secondary gradient-boosting/SHAP analysis was retained as a non-causal variance discovery to rank the metadata patterns used to predict DI. Candidate TEI updates were estimated within the vendor-region-projection groups by algebraically re-centering the observed mean DI ( D I ¯ ) $\overline {DI} )$ to 0.0. Robustness was assessed using sensitivity analyses restricted by automatic exposure control, beam energy, and patient habitus proxy measures.

Results

The baseline D I ¯ $\overline {DI} $  = −3.1 (SD 3.7), and only 11.3% of examinations were within |DI| ≤ 1. Forty of 45 vendor–region–projection groups (88.9%) had an absolute group D I ¯ $\overline {DI} $ greater than 1.0. Separately, the mean CAI was 0.176. The secondary SHAP analysis ranked the configured TEI, manufacturer, SID, exposure, and kVp as high-contribution predictors of DI; these findings were interpreted as predictive attribution, not causal evidence. DI re-centering reduced the overall variance by 42.9% (13.5 to 7.7), increased the proportion of examinations within |DI| ≤ 1 to 34.3%, and reduced severe deviations (|DI| > 3) from 60.3% to 23.1%.

Conclusion

In this retrospective, multi-vendor study, the TEI setting review was operationalized as a structured, data-driven responsibility of a clinical site. By verifying protocol adherence, equipment performance, image quality, and dose indicators before centering the D I ¯ $\overline {DI} $ , departments can support the conversion of the DI from a biased compliance signal into a practical process-control metric for exposure governance.

桥接TG-116和TG-232:数字摄影中偏差指标重新定心和过程控制的回顾性多厂商评估。
背景:偏差指数(DI)是相对于配置的目标暴露指数(TEI)的检测器暴露的供应商中立指标。因为TEI是一个协议值,而不是一个校准过的物理量,所以不正确的TEI设置会造成系统的DI偏移,模糊协议问题,并妨碍有意义的统计过程控制。目的:这项多厂商的回顾性研究评估了常规DICOM元数据是否能够识别有偏差的DI分布,将固定的TEI设置偏移量与协议依从性变化分开,估计候选TEI更新,以及在DI重新定位后支持基于tg -232的标准偏差监测。方法:回顾性分析某三级医院从2024年1月1日至2024年8月27日使用三个供应商平台进行的38916次数字x线检查。DICOM元数据使用SID、kVp、mAs、网格使用和AEC使用的复合粘附指数(CAI)对发布的技术标准进行基准测试。二级梯度增强/SHAP分析被保留为非因果方差发现,用于对用于预测DI的元数据模式进行排序。通过将观测到的平均DI (DI¯)重新集中到0.0,在供应商-区域-投影组中估计候选TEI更新。鲁棒性评估采用受自动暴露控制、光束能量和患者习惯代理测量限制的敏感性分析。结果:基线DI¯= -3.1 (SD 3.7),只有11.3%的检查在|≤1范围内。45个供应商-地区-投影组中有40个(88.9%)的绝对组D¯大于1.0。另外,平均CAI为0.176。二级SHAP分析将配置的TEI、制造商、SID、暴露和kVp列为DI的高贡献预测因子;这些发现被解释为预测性归因,而不是因果证据。DI重新定心使总体方差降低了42.9%(13.5 - 7.7),将|DI|≤1的检查比例提高到34.3%,将严重偏差(|DI| > 3)从60.3%降低到23.1%。结论:在这项多供应商的回顾性研究中,TEI设置审查作为一种结构化的、数据驱动的临床站点责任进行了操作。通过在定心DI¯之前验证方案依从性、设备性能、图像质量和剂量指标,部门可以支持将DI从有偏差的依从性信号转换为暴露治理的实际过程控制度量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书