{"title":"Epicardial Use of Medtronic 3830 Lead in Pediatric Cardiac Resynchronization Therapy with Elevated Thresholds.","authors":"Husnu Firat Altin, Qusi Shaban, Ahmed Reyaz Lone, Shadab Wani, Olivier Ghez, Volkan Tuzcu","doi":"10.1111/pace.70349","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Epicardial pacing is standard in children, but conventional leads often fail due to high thresholds, fractures, or poor sensing after multiple surgeries.</p><p><strong>Case: </strong>A 10-year-old with restrictive cardiomyopathy and prior mitral valve replacement presented with loss of capture from fractured epicardial leads. Redo sternotomy attempts with conventional leads produced thresholds up to 8.0 V. A Medtronic SelectSecure 3830 lead was implanted epicardially, achieving improved thresholds (2.7 V → 1.1 V) and stable cardiac resynchronization therapy (CRT) pacemaker function. At 2 months, thresholds and impedances remained low and stable.</p><p><strong>Conclusion: </strong>Epicardial Medtronic SelectSecure 3830 lead implantation may be a feasible alternative in complex pediatric redo cases.</p>","PeriodicalId":520740,"journal":{"name":"Pacing and clinical electrophysiology : PACE","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pacing and clinical electrophysiology : PACE","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1111/pace.70349","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Epicardial pacing is standard in children, but conventional leads often fail due to high thresholds, fractures, or poor sensing after multiple surgeries.
Case: A 10-year-old with restrictive cardiomyopathy and prior mitral valve replacement presented with loss of capture from fractured epicardial leads. Redo sternotomy attempts with conventional leads produced thresholds up to 8.0 V. A Medtronic SelectSecure 3830 lead was implanted epicardially, achieving improved thresholds (2.7 V → 1.1 V) and stable cardiac resynchronization therapy (CRT) pacemaker function. At 2 months, thresholds and impedances remained low and stable.
Conclusion: Epicardial Medtronic SelectSecure 3830 lead implantation may be a feasible alternative in complex pediatric redo cases.