Quantifying three-dimensional displacement of cardiac implantable electronic devices and implications for safety margins in radiotherapy.

IF 1.6 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Takahiro Kubota, Ryota Onizuka, Kazuki Kajiwara, Tatsunori Saho, Koichiro Muraki
{"title":"Quantifying three-dimensional displacement of cardiac implantable electronic devices and implications for safety margins in radiotherapy.","authors":"Takahiro Kubota, Ryota Onizuka, Kazuki Kajiwara, Tatsunori Saho, Koichiro Muraki","doi":"10.1007/s12194-026-01142-x","DOIUrl":null,"url":null,"abstract":"<p><p>This study aimed to quantitatively evaluate the three-dimensional (3D) displacement of cardiac implantable electronic devices (CIEDs) during radiotherapy and derive safety margins. Sixteen treatment courses (13 patients, 262 fractions) were retrospectively analyzed. Daily device displacement was quantified using orthogonal 2D kV images relative to clinical bone-matching. To evaluate manual registration uncertainty, the intraclass correlation coefficient and minimal detectable change (MDC) were calculated. To assess the impact of patient positioning, displacements were compared between arm-up and arm-down groups. Planning organ at risk volume (PRV) margins were derived from systematic and random errors. A maximum 3D displacement of 24 mm was observed. Manual registration demonstrated good to excellent reproducibility for the primary translational axes. The maximum MDC was 3.6 mm for the 3D displacement vector. While overall displacement distributions showed no significant differences between arm positions, extreme medial displacement significantly increased in the arm-up group (p < 0.05). The calculated PRV margins reached a maximum of 7.3 mm in the longitudinal direction for the arm-up group. Throughout the radiotherapy course, CIEDs can undergo displacement independently of the skeletal structure. The PRV margins derived here provide a quantitative basis for pre-treatment dose evaluation. The observed range of displacement can inform treatment planning and supports verification of the device position at each fraction via image-guided radiotherapy.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01142-x","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

This study aimed to quantitatively evaluate the three-dimensional (3D) displacement of cardiac implantable electronic devices (CIEDs) during radiotherapy and derive safety margins. Sixteen treatment courses (13 patients, 262 fractions) were retrospectively analyzed. Daily device displacement was quantified using orthogonal 2D kV images relative to clinical bone-matching. To evaluate manual registration uncertainty, the intraclass correlation coefficient and minimal detectable change (MDC) were calculated. To assess the impact of patient positioning, displacements were compared between arm-up and arm-down groups. Planning organ at risk volume (PRV) margins were derived from systematic and random errors. A maximum 3D displacement of 24 mm was observed. Manual registration demonstrated good to excellent reproducibility for the primary translational axes. The maximum MDC was 3.6 mm for the 3D displacement vector. While overall displacement distributions showed no significant differences between arm positions, extreme medial displacement significantly increased in the arm-up group (p < 0.05). The calculated PRV margins reached a maximum of 7.3 mm in the longitudinal direction for the arm-up group. Throughout the radiotherapy course, CIEDs can undergo displacement independently of the skeletal structure. The PRV margins derived here provide a quantitative basis for pre-treatment dose evaluation. The observed range of displacement can inform treatment planning and supports verification of the device position at each fraction via image-guided radiotherapy.

心脏植入式电子装置三维位移的量化及其放射治疗安全裕度的意义。
本研究旨在定量评估放射治疗期间心脏植入式电子装置(CIEDs)的三维(3D)位移并得出安全裕度。回顾性分析16个疗程(13例,262份)。使用相对于临床骨匹配的正交2D kV图像来量化每日装置位移。为了评估人工配准的不确定性,计算了类内相关系数和最小可检测变化(最小可检测变化)。为了评估患者体位的影响,比较了上臂组和下臂组之间的位移。计划器官风险体积(PRV)边界是由系统误差和随机误差得出的。观察到最大三维位移为24 mm。手工配准证明了主平移轴具有良好的再现性。三维位移矢量的最大MDC为3.6 mm。虽然整体位移分布在手臂位置之间没有显着差异,但手臂向上组的极端内侧位移显着增加(p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书