Enhancing Far-Forward Resuscitation: A Quantitative Laboratory Analysis of Blood Warmer Devices Used in Austere Environments.

IF 2 3区 医学 Q2 EMERGENCY MEDICINE
Wayne Papalski, Matthew A Tovar, Eric Sulava, Eric J Koch, Mariah A Jensen-Wachspress, Matthew Monaco, Conner Murphy, Emily Friedrich, Fernando U Gonzalez, Lorie Gower, Kay-Cee Patterson, Christopher Treager, Brittany Dellinger
{"title":"Enhancing Far-Forward Resuscitation: A Quantitative Laboratory Analysis of Blood Warmer Devices Used in Austere Environments.","authors":"Wayne Papalski, Matthew A Tovar, Eric Sulava, Eric J Koch, Mariah A Jensen-Wachspress, Matthew Monaco, Conner Murphy, Emily Friedrich, Fernando U Gonzalez, Lorie Gower, Kay-Cee Patterson, Christopher Treager, Brittany Dellinger","doi":"10.1080/10903127.2026.2736179","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>It is well known that hypothermia increases organ dysfunction and mortality in traumatically injured patients. While several small, portable blood warmers are commercially available, comparative data on their warming efficiency and flow dynamics remain limited. Thus, the objective of this study was to identify optimal blood warming devices for far-forward austere trauma resuscitation.</p><p><strong>Methods: </strong>This prospective experimental study assessed six blood warming devices: QinFlow Warrior Lite<sup>™</sup>, MWarmer<sup>™</sup>, Buddy Lite<sup>™</sup>, Thermal Angel<sup>®</sup>, the LifeWarmer Quantum<sup>™</sup>, and the Ready Heat Blanket<sup>™</sup>, using whole blood stored at 4 °C. Devices were tested in triplicate under two pressure conditions (0 and 300 mmHg). Flow was measured volumetrically and converted into average (Qavg), ideal (Qi), and maximum (Qmax) flow rates. Temperature was measured at the distal end of intravenous tubing and analyzed using average (Tavg), ideal (Ti), and maximum (Tmax) definitions. Time-to-one-unit (TT1U) was calculated as a clinically meaningful metric. Two-way ANOVA with interaction terms and post hoc Tukey testing were used for statistical comparison, with significance defined as multiplicity-adjusted <i>p</i> < 0.05.</p><p><strong>Results: </strong>The Quantum exhibited the highest Qi at both pressure levels, achieving a TT1U of 1.55 ± 0.10 min at 300 mmHg, significantly faster than all other devices. However, its warming performance declined significantly with increased pressure, with Ti decreasing from 34.7 °C to 25.6 °C (<i>p</i> < 0.0001). In contrast, the MWarmer demonstrated the highest and most consistent warming performance across all pressure conditions, maintaining Ti of 40-41 °C (<i>p</i> < 0.0001) at both pressure conditions. The Buddy Lite showed poor warming performance with pressure-related degradation, while Ready Heat provided negligible warming capacity with flow comparable to control.</p><p><strong>Conclusions: </strong>This is an independent quantitative comparison of portable blood warming devices using cold-stored whole blood under clinically relevant pressure conditions. Results suggest that the MWarmer offers the most favorable balance of thermal performance and flow for far-forward resuscitation.</p>","PeriodicalId":20336,"journal":{"name":"Prehospital Emergency Care","volume":" ","pages":"1-8"},"PeriodicalIF":2.0000,"publicationDate":"2026-10-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prehospital Emergency Care","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/10903127.2026.2736179","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"EMERGENCY MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: It is well known that hypothermia increases organ dysfunction and mortality in traumatically injured patients. While several small, portable blood warmers are commercially available, comparative data on their warming efficiency and flow dynamics remain limited. Thus, the objective of this study was to identify optimal blood warming devices for far-forward austere trauma resuscitation.

Methods: This prospective experimental study assessed six blood warming devices: QinFlow Warrior Lite™, MWarmer™, Buddy Lite™, Thermal Angel®, the LifeWarmer Quantum™, and the Ready Heat Blanket™, using whole blood stored at 4 °C. Devices were tested in triplicate under two pressure conditions (0 and 300 mmHg). Flow was measured volumetrically and converted into average (Qavg), ideal (Qi), and maximum (Qmax) flow rates. Temperature was measured at the distal end of intravenous tubing and analyzed using average (Tavg), ideal (Ti), and maximum (Tmax) definitions. Time-to-one-unit (TT1U) was calculated as a clinically meaningful metric. Two-way ANOVA with interaction terms and post hoc Tukey testing were used for statistical comparison, with significance defined as multiplicity-adjusted p < 0.05.

Results: The Quantum exhibited the highest Qi at both pressure levels, achieving a TT1U of 1.55 ± 0.10 min at 300 mmHg, significantly faster than all other devices. However, its warming performance declined significantly with increased pressure, with Ti decreasing from 34.7 °C to 25.6 °C (p < 0.0001). In contrast, the MWarmer demonstrated the highest and most consistent warming performance across all pressure conditions, maintaining Ti of 40-41 °C (p < 0.0001) at both pressure conditions. The Buddy Lite showed poor warming performance with pressure-related degradation, while Ready Heat provided negligible warming capacity with flow comparable to control.

Conclusions: This is an independent quantitative comparison of portable blood warming devices using cold-stored whole blood under clinically relevant pressure conditions. Results suggest that the MWarmer offers the most favorable balance of thermal performance and flow for far-forward resuscitation.

加强远前复苏:在严酷环境中使用的血液加热装置的定量实验室分析。
目的:众所周知,低温会增加创伤性损伤患者的器官功能障碍和死亡率。虽然市面上有几种小型便携式血液加热器,但其加热效率和血流动力学的比较数据仍然有限。因此,本研究的目的是确定远前严重创伤复苏的最佳血液升温装置。方法:本前瞻性实验研究评估了六种血液加热装置:QinFlow Warrior Lite™、mwarm™、Buddy Lite™、Thermal Angel®、lifewarm Quantum™和Ready Heat Blanket™,使用4°C保存的全血。设备在两种压力条件下(0和300 mmHg)进行了三次测试。以容积法测量流量,并将其转换为平均流量(Qavg)、理想流量(Qi)和最大流量(Qmax)。在静脉插管远端测量温度,并使用平均(Tavg)、理想(Ti)和最大值(Tmax)定义进行分析。计算每单位时间(TT1U)作为临床有意义的指标。采用带交互项的双因素方差分析和事后Tukey检验进行统计学比较,显著性定义为多重校正p < 0.05。结果:Quantum在两种压力水平下都表现出最高的Qi,在300 mmHg下达到1.55±0.10分钟的TT1U,明显快于所有其他器件。然而,随着压力的增加,其增温性能显著下降,Ti从34.7°C下降到25.6°C (p < 0.0001)。相比之下,mwarm在所有压力条件下都表现出最高和最一致的升温性能,在两种压力条件下都保持40-41°C的Ti (p < 0.0001)。Buddy Lite的升温性能较差,与压力相关的性能下降,而Ready Heat的升温能力可以忽略不计,其流量与对照组相当。结论:这是在临床相关压力条件下使用冷藏全血的便携式血液温血装置的独立定量比较。结果表明,mwarm为远前复苏提供了最有利的热性能和流量平衡。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Prehospital Emergency Care
Prehospital Emergency Care 医学-公共卫生、环境卫生与职业卫生
CiteScore
4.30
自引率
12.50%
发文量
137
审稿时长
1 months
期刊介绍: Prehospital Emergency Care publishes peer-reviewed information relevant to the practice, educational advancement, and investigation of prehospital emergency care, including the following types of articles: Special Contributions - Original Articles - Education and Practice - Preliminary Reports - Case Conferences - Position Papers - Collective Reviews - Editorials - Letters to the Editor - Media Reviews.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书