{"title":"Variability and bias in the calculation of iodine uptake with an adaptive background subtraction method in contrast-enhanced mammography.","authors":"Jorge Patricio Castillo-Lopez, Liliana Moreno Astudillo, Yolanda Villaseñor-Navarro, María-Ester Brandan","doi":"10.1088/1361-6560/aeaac1","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective.</i>This work evaluates variability and bias in the calculation of iodine uptake from contrast-enhanced dual-energy mammography (CEM) images using an adaptive background subtraction method.<i>Approach.</i>The iodine uptake images were custom-generated using weighted subtraction, followed by subtraction of a smooth background estimated by interpolation from neighboring regions outside the lesion (adaptive background subtraction). Three main sources of variability and bias were assessed: differences in breast tissue linear attenuation coefficients, stability of the mammography system, and intra- and inter-observer variability in lesion segmentation. Tissue-related effects were evaluated theoretically. System stability was assessed using weekly phantom measurements acquired over a 4-year period. For observer variability, four radiologists performed two segmentation rounds of iodine-enhancing regions in 40 CEM studies. Typical uptake values were below 5 mg cm<sup>-2</sup>. Whenever possible, variability was expressed as the 95% coverage interval (2σ), where σ represents the standard deviation<i>. Main results.</i>Theoretical evaluation showed that tissue composition differences may introduce biases in iodine uptake exceeding 0.5 mg cm<sup>-2</sup>, although typical lesion compositions produce biases below the visual detection threshold. Long-term phantom measurements demonstrated excellent system stability over four years (2σ ≈0.2 mg cm<sup>-2</sup>, and limit of detection = 0.24 mg cm<sup>-2</sup>), with a 7% change in calibration following a workstation upgrade. In the clinical dataset, the four radiologists achieved high detection agreement when segmenting the primary lesions and greater variability for secondary lesions; the inter-observer variability in mean uptake was approximately 2σ ≈0.6 mg cm<sup>-2</sup>.<i>Significance.</i>These findings indicate that iodine uptake quantification in CEM using an adaptive background subtraction method is technically robust, which is a key element for exploring its potential usefulness as an imaging biomarker.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":" ","pages":""},"PeriodicalIF":3.1000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/aeaac1","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective.This work evaluates variability and bias in the calculation of iodine uptake from contrast-enhanced dual-energy mammography (CEM) images using an adaptive background subtraction method.Approach.The iodine uptake images were custom-generated using weighted subtraction, followed by subtraction of a smooth background estimated by interpolation from neighboring regions outside the lesion (adaptive background subtraction). Three main sources of variability and bias were assessed: differences in breast tissue linear attenuation coefficients, stability of the mammography system, and intra- and inter-observer variability in lesion segmentation. Tissue-related effects were evaluated theoretically. System stability was assessed using weekly phantom measurements acquired over a 4-year period. For observer variability, four radiologists performed two segmentation rounds of iodine-enhancing regions in 40 CEM studies. Typical uptake values were below 5 mg cm-2. Whenever possible, variability was expressed as the 95% coverage interval (2σ), where σ represents the standard deviation. Main results.Theoretical evaluation showed that tissue composition differences may introduce biases in iodine uptake exceeding 0.5 mg cm-2, although typical lesion compositions produce biases below the visual detection threshold. Long-term phantom measurements demonstrated excellent system stability over four years (2σ ≈0.2 mg cm-2, and limit of detection = 0.24 mg cm-2), with a 7% change in calibration following a workstation upgrade. In the clinical dataset, the four radiologists achieved high detection agreement when segmenting the primary lesions and greater variability for secondary lesions; the inter-observer variability in mean uptake was approximately 2σ ≈0.6 mg cm-2.Significance.These findings indicate that iodine uptake quantification in CEM using an adaptive background subtraction method is technically robust, which is a key element for exploring its potential usefulness as an imaging biomarker.
期刊介绍:
The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry