Marie Seret, Claire Dupuis, Vanessa Alves de Araújo, Claudio Annicchiarico, Gabriela Macedo, Caio Pinto Teixeira Araujo, Fabio Taccone, Miriam Cnop, Thomas Desaive, Vincent Uyttendaele, Jean-Charles Preiser
{"title":"Accuracy of an Interstitial Continuous Glucose Management System in Critically Ill Patients: A Prospective Observational Study.","authors":"Marie Seret, Claire Dupuis, Vanessa Alves de Araújo, Claudio Annicchiarico, Gabriela Macedo, Caio Pinto Teixeira Araujo, Fabio Taccone, Miriam Cnop, Thomas Desaive, Vincent Uyttendaele, Jean-Charles Preiser","doi":"10.1177/19322968261486451","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Glycemic management in the intensive care unit (ICU) remains challenging because glucose levels fluctuate rapidly and monitoring is intermittent. Interstitial continuous glucose monitoring (CGM) could improve surveillance, but its performance during rapid glucose excursions remains uncertain. We assessed the accuracy of the Dexcom G7 CGM in critically ill patients under stable conditions and during intravenous glucose tolerance tests (IVGTT).</p><p><strong>Methods: </strong>We conducted a prospective ancillary study of the Determination of Insulin Resistance in ICU (DIRICU) study. Twenty critically ill adults with stress hyperglycemia, no prior diabetes, and expected ICU stay ≥5 days were included. A Dexcom G7 sensor was compared with arterial blood glucose measurements. Accuracy was assessed over the first 10 ICU days using paired measurements, mean bias, mean absolute relative difference (MARD), linear regression (<i>R</i><sup>2</sup>), and Clarke error grid analysis, both outside and during IVGTT. A post hoc 20-minute temporal shift analysis was performed to account for delayed CGM peaks.</p><p><strong>Results: </strong>Outside IVGTT, 716 paired values showed moderate agreement with reference glucose (<i>R</i><sup>2</sup> = .58, mean bias +11 mg/dL, MARD 17%), with 98.9% of values in Clarke zones A/B. During IVGTT, accuracy deteriorated markedly (1018 paired values): <i>R</i><sup>2</sup> = .028, mean bias -34 mg/dL, MARD 24%, and only 80.3% of values in zones A/B, with 18.6% in zone D. After temporal shifting, performance improved (<i>R</i><sup>2</sup> = .34, MARD 19%, 94.7% in zones A/B). No sensor-related adverse events occurred.</p><p><strong>Conclusions: </strong>Dexcom G7 showed acceptable accuracy during stable ICU conditions, but performance deteriorated during rapid glucose excursions. Hybrid monitoring strategies combining CGM with intermittent blood glucose confirmation may therefore be preferable in critically ill patients.</p>","PeriodicalId":15475,"journal":{"name":"Journal of Diabetes Science and Technology","volume":" ","pages":"19322968261486451"},"PeriodicalIF":4.9000,"publicationDate":"2026-09-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13593934/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19322968261486451","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Glycemic management in the intensive care unit (ICU) remains challenging because glucose levels fluctuate rapidly and monitoring is intermittent. Interstitial continuous glucose monitoring (CGM) could improve surveillance, but its performance during rapid glucose excursions remains uncertain. We assessed the accuracy of the Dexcom G7 CGM in critically ill patients under stable conditions and during intravenous glucose tolerance tests (IVGTT).
Methods: We conducted a prospective ancillary study of the Determination of Insulin Resistance in ICU (DIRICU) study. Twenty critically ill adults with stress hyperglycemia, no prior diabetes, and expected ICU stay ≥5 days were included. A Dexcom G7 sensor was compared with arterial blood glucose measurements. Accuracy was assessed over the first 10 ICU days using paired measurements, mean bias, mean absolute relative difference (MARD), linear regression (R2), and Clarke error grid analysis, both outside and during IVGTT. A post hoc 20-minute temporal shift analysis was performed to account for delayed CGM peaks.
Results: Outside IVGTT, 716 paired values showed moderate agreement with reference glucose (R2 = .58, mean bias +11 mg/dL, MARD 17%), with 98.9% of values in Clarke zones A/B. During IVGTT, accuracy deteriorated markedly (1018 paired values): R2 = .028, mean bias -34 mg/dL, MARD 24%, and only 80.3% of values in zones A/B, with 18.6% in zone D. After temporal shifting, performance improved (R2 = .34, MARD 19%, 94.7% in zones A/B). No sensor-related adverse events occurred.
Conclusions: Dexcom G7 showed acceptable accuracy during stable ICU conditions, but performance deteriorated during rapid glucose excursions. Hybrid monitoring strategies combining CGM with intermittent blood glucose confirmation may therefore be preferable in critically ill patients.
期刊介绍:
The Journal of Diabetes Science and Technology (JDST) is a bi-monthly, peer-reviewed scientific journal published by the Diabetes Technology Society. JDST covers scientific and clinical aspects of diabetes technology including glucose monitoring, insulin and metabolic peptide delivery, the artificial pancreas, digital health, precision medicine, social media, cybersecurity, software for modeling, physiologic monitoring, technology for managing obesity, and diagnostic tests of glycation. The journal also covers the development and use of mobile applications and wireless communication, as well as bioengineered tools such as MEMS, new biomaterials, and nanotechnology to develop new sensors. Articles in JDST cover both basic research and clinical applications of technologies being developed to help people with diabetes.