{"title":"Calibration and count rate performance of NaI(Tl) gamma counters for quantifying the activity 89Zr radiopharmaceuticals","authors":"Michael Silosky","doi":"10.1002/acm2.70788","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Quantification of low activity samples using gamma counters is a common technique when investigating the biodistribution of experimental radiopharmaceuticals. Several human and pre-clinical trials have investigated <sup>89</sup>Zr labeled compounds, utilizing gamma counter to quantify activity in blood or animal samples. While the performance of gamma counters has been investigated for other positron emitting radionuclides, the more complex spectrum, including a 909 keV gamma emission, and low positron fraction of <sup>89</sup>Zr may reduce both counting efficiency while increasing the severity of count rate losses.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>The purpose of this study is to evaluate both the counting efficiency and the effect of count rate losses on the quantitation of activity for <sup>89</sup>Zr using NaI(Tl) gamma counters.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Sources of known activity (initially 78.3 kBq) were measured in a gamma counter as they decayed over approximately 15 days. Observed count rates were fit with a paralyzable detector model for two separate energy windows, 511 keV ± 10% and 909 keV ± 10%, as well as their sum. Counting efficiency in the absence of losses as well as the effective deadtime for each dataset were calculated.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Count rates for each energy window were well fit with a paralyzable model with <i>R</i><sup>2</sup> > 0.999. Counting efficiency (mean ± 95% confidence interval) was 10.07% ± 0.07%, 17.38% ± 0.13%, and 27.46% ± 0.16% for the 511 keV, 909 keV, and summed energy windows respectively. Effective dead time (mean, 95% confidence interval) was 14.31 µs, (13.30 µs–15.30 µs), 7.11 µs (6.68 µs–7.55 µs), and 4.77 µs (4.48 µs–5.06 µs) for the 511 keV, 909 keV, and summed energy windows respectively.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Counting efficiency and count rate performance for <sup>89</sup>Zr are poorer than for other positron emitting radionuclides. Count rate losses follow the paralyzable model and are correctible.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70788","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70788","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Quantification of low activity samples using gamma counters is a common technique when investigating the biodistribution of experimental radiopharmaceuticals. Several human and pre-clinical trials have investigated 89Zr labeled compounds, utilizing gamma counter to quantify activity in blood or animal samples. While the performance of gamma counters has been investigated for other positron emitting radionuclides, the more complex spectrum, including a 909 keV gamma emission, and low positron fraction of 89Zr may reduce both counting efficiency while increasing the severity of count rate losses.
Purpose
The purpose of this study is to evaluate both the counting efficiency and the effect of count rate losses on the quantitation of activity for 89Zr using NaI(Tl) gamma counters.
Methods
Sources of known activity (initially 78.3 kBq) were measured in a gamma counter as they decayed over approximately 15 days. Observed count rates were fit with a paralyzable detector model for two separate energy windows, 511 keV ± 10% and 909 keV ± 10%, as well as their sum. Counting efficiency in the absence of losses as well as the effective deadtime for each dataset were calculated.
Results
Count rates for each energy window were well fit with a paralyzable model with R2 > 0.999. Counting efficiency (mean ± 95% confidence interval) was 10.07% ± 0.07%, 17.38% ± 0.13%, and 27.46% ± 0.16% for the 511 keV, 909 keV, and summed energy windows respectively. Effective dead time (mean, 95% confidence interval) was 14.31 µs, (13.30 µs–15.30 µs), 7.11 µs (6.68 µs–7.55 µs), and 4.77 µs (4.48 µs–5.06 µs) for the 511 keV, 909 keV, and summed energy windows respectively.
Conclusion
Counting efficiency and count rate performance for 89Zr are poorer than for other positron emitting radionuclides. Count rate losses follow the paralyzable model and are correctible.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic