Comprehensive acceptance testing and quality control of Single Photon Emission Computed Tomography/Computed Tomography systems.

IF 1.4 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Noor Ul Huda Abbasi, Hannan Younis, Saeed Ur Rehman, Burhan Zia Abbasi, Roya Feroze, Khurram Mehboob, Muhammad Ajaz
{"title":"Comprehensive acceptance testing and quality control of Single Photon Emission Computed Tomography/Computed Tomography systems.","authors":"Noor Ul Huda Abbasi, Hannan Younis, Saeed Ur Rehman, Burhan Zia Abbasi, Roya Feroze, Khurram Mehboob, Muhammad Ajaz","doi":"10.1097/MNM.0000000000002240","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Single photon emission computed tomography/computed tomography (SPECT/CT) systems play a crucial role in nuclear medicine for accurate diagnosis and treatment planning. Ensuring high image quality through rigorous acceptance testing and quality control is essential to maintain diagnostic reliability and patient safety.</p><p><strong>Objective: </strong>This study evaluated the performance of SPECT/CT systems at NORI Hospital Islamabad against National Electrical Manufacturers Association (NEMA) and International Atomic Energy Agency standards and developed a structured quality control program based on acceptance testing results.</p><p><strong>Methods: </strong>Acceptance tests were carried out using Tc-99m, I-131, and Co-57 radioactive sources, in conjunction with quadrant bar and Jaszczak phantoms. Parameters assessed included spatial resolution, intrinsic and extrinsic uniformity, energy resolution, sensitivity, center of rotation (COR), detector shielding, background counts, leakage, CT number accuracy, and nuclear medicine/CT registration. Data were compared with manufacturer specifications and NEMA standards to evaluate system performance and inform quality control strategies.</p><p><strong>Results: </strong>Both detectors exhibited intrinsic and extrinsic uniformity within NEMA limits, ensuring consistent image quality. Spatial resolution testing confirmed sharp and precise imaging, while energy resolution was within manufacturer-specified values. Sensitivity measurements showed minor differences between detectors but remained within acceptable ranges. COR values were precise (<±0.55 mm), confirming proper mechanical and software alignment. CT number accuracy and nuclear medicine/CT registration tests confirmed precise anatomical-functional image overlay. Detector shielding and leakage tests showed minimal interference, supporting reliable imaging.</p><p><strong>Conclusion: </strong>The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality. Acceptance testing validated detector performance and highlighted minor areas for improvement, such as periodic sensitivity calibration. Implementation of a structured quality control program ensures consistent equipment performance, optimal image quality, improved diagnostic accuracy, and superior patient care. Routine adherence to international standards is essential for maintaining high-quality nuclear medicine imaging practices.</p>","PeriodicalId":19708,"journal":{"name":"Nuclear Medicine Communications","volume":" ","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2026-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Medicine Communications","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MNM.0000000000002240","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Single photon emission computed tomography/computed tomography (SPECT/CT) systems play a crucial role in nuclear medicine for accurate diagnosis and treatment planning. Ensuring high image quality through rigorous acceptance testing and quality control is essential to maintain diagnostic reliability and patient safety.

Objective: This study evaluated the performance of SPECT/CT systems at NORI Hospital Islamabad against National Electrical Manufacturers Association (NEMA) and International Atomic Energy Agency standards and developed a structured quality control program based on acceptance testing results.

Methods: Acceptance tests were carried out using Tc-99m, I-131, and Co-57 radioactive sources, in conjunction with quadrant bar and Jaszczak phantoms. Parameters assessed included spatial resolution, intrinsic and extrinsic uniformity, energy resolution, sensitivity, center of rotation (COR), detector shielding, background counts, leakage, CT number accuracy, and nuclear medicine/CT registration. Data were compared with manufacturer specifications and NEMA standards to evaluate system performance and inform quality control strategies.

Results: Both detectors exhibited intrinsic and extrinsic uniformity within NEMA limits, ensuring consistent image quality. Spatial resolution testing confirmed sharp and precise imaging, while energy resolution was within manufacturer-specified values. Sensitivity measurements showed minor differences between detectors but remained within acceptable ranges. COR values were precise (<±0.55 mm), confirming proper mechanical and software alignment. CT number accuracy and nuclear medicine/CT registration tests confirmed precise anatomical-functional image overlay. Detector shielding and leakage tests showed minimal interference, supporting reliable imaging.

Conclusion: The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality. Acceptance testing validated detector performance and highlighted minor areas for improvement, such as periodic sensitivity calibration. Implementation of a structured quality control program ensures consistent equipment performance, optimal image quality, improved diagnostic accuracy, and superior patient care. Routine adherence to international standards is essential for maintaining high-quality nuclear medicine imaging practices.

单光子发射计算机断层扫描/计算机断层扫描系统的全面验收测试和质量控制。
背景:单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)系统在核医学中对准确诊断和治疗计划起着至关重要的作用。通过严格的验收测试和质量控制确保高图像质量对于保持诊断可靠性和患者安全至关重要。目的:本研究根据国家电气制造商协会(NEMA)和国际原子能机构的标准评估了伊斯兰堡NORI医院SPECT/CT系统的性能,并根据验收测试结果制定了结构化的质量控制计划。方法:采用Tc-99m、I-131和Co-57放射源,结合象限棒和Jaszczak幻影进行验收试验。评估的参数包括空间分辨率、内在和外在均匀性、能量分辨率、灵敏度、旋转中心(COR)、探测器屏蔽、背景计数、泄漏、CT数准确性和核医学/CT配准。将数据与制造商规范和NEMA标准进行比较,以评估系统性能并告知质量控制策略。结果:两种检测器在NEMA范围内表现出内在和外在的均匀性,确保了一致的图像质量。空间分辨率测试证实了清晰和精确的成像,而能量分辨率在制造商规定的值之内。灵敏度测量显示探测器之间的差异很小,但仍在可接受的范围内。结论:伊斯兰堡NORI医院的SPECT/CT系统符合既定的NEMA图像质量标准。验收测试验证了探测器的性能,并强调了需要改进的小领域,如周期性灵敏度校准。结构化质量控制程序的实施确保了一致的设备性能、最佳的图像质量、改进的诊断准确性和卓越的患者护理。常规遵守国际标准对于保持高质量的核医学成像实践至关重要。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.20
自引率
6.70%
发文量
212
审稿时长
3-8 weeks
期刊介绍: Nuclear Medicine Communications, the official journal of the British Nuclear Medicine Society, is a rapid communications journal covering nuclear medicine and molecular imaging with radionuclides, and the basic supporting sciences. As well as clinical research and commentary, manuscripts describing research on preclinical and basic sciences (radiochemistry, radiopharmacy, radiobiology, radiopharmacology, medical physics, computing and engineering, and technical and nursing professions involved in delivering nuclear medicine services) are welcomed, as the journal is intended to be of interest internationally to all members of the many medical and non-medical disciplines involved in nuclear medicine. In addition to papers reporting original studies, frankly written editorials and topical reviews are a regular feature of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书