Mark-Alexander Solf, R Tomasi, C-E Kamla, J Kreuzer, M Feuerecker, P Scheiermann, D-J Höchter, T Schumann, C Müller
{"title":"[Left ventricular assist devices in emergency medicine : Case-based analysis for emergency medical services and emergency departments].","authors":"Mark-Alexander Solf, R Tomasi, C-E Kamla, J Kreuzer, M Feuerecker, P Scheiermann, D-J Höchter, T Schumann, C Müller","doi":"10.1007/s00063-026-01490-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Left ventricular assist devices (LVAD) are an established treatment for advanced heart failure. Due to the continuous flow, pulse palpation, noninvasive blood pressure measurement and pulse oximetry are often unreliable in acute care.</p><p><strong>Objective: </strong>Case-based analysis of the acute care of patients with left ventricular assist devices in emergency medical services and emergency departments, exemplified by a patient supported with HeartMate 3® and derivation of practical recommendations for action.</p><p><strong>Material and methods: </strong>Anonymized prehospital and clinical documentation of a 67-year-old female patient was evaluated. Monitoring, clinical assessment, device and power supply checks, pump parameters, diagnostics and triage were analyzed using consensus documents, reviews, registry data and manufacturer's data.</p><p><strong>Results: </strong>In this study three key problem areas emerged: unreliable standard monitoring, initial misclassification as technical device failure and necessary contextual interpretation of pump parameters. Prehospital findings included dyspnea, cyanosis, impaired consciousness and inability to obtain blood pressure or oxygen saturation values. The later device readout showed stable parameters without acute pump malfunction. The underlying cause was severe Staphylococcus aureus bacteremia with renal replacement therapy-dependent renal failure.</p><p><strong>Conclusion: </strong>Severe clinical deterioration in LVAD patients is not necessarily pump-related. Key elements are perfusion assessment, device and power supply checks, interpretation of pump parameters, extended diagnostics and early consultation with an LVAD center.</p>","PeriodicalId":49019,"journal":{"name":"Medizinische Klinik-Intensivmedizin Und Notfallmedizin","volume":" ","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2026-09-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medizinische Klinik-Intensivmedizin Und Notfallmedizin","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00063-026-01490-x","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Left ventricular assist devices (LVAD) are an established treatment for advanced heart failure. Due to the continuous flow, pulse palpation, noninvasive blood pressure measurement and pulse oximetry are often unreliable in acute care.
Objective: Case-based analysis of the acute care of patients with left ventricular assist devices in emergency medical services and emergency departments, exemplified by a patient supported with HeartMate 3® and derivation of practical recommendations for action.
Material and methods: Anonymized prehospital and clinical documentation of a 67-year-old female patient was evaluated. Monitoring, clinical assessment, device and power supply checks, pump parameters, diagnostics and triage were analyzed using consensus documents, reviews, registry data and manufacturer's data.
Results: In this study three key problem areas emerged: unreliable standard monitoring, initial misclassification as technical device failure and necessary contextual interpretation of pump parameters. Prehospital findings included dyspnea, cyanosis, impaired consciousness and inability to obtain blood pressure or oxygen saturation values. The later device readout showed stable parameters without acute pump malfunction. The underlying cause was severe Staphylococcus aureus bacteremia with renal replacement therapy-dependent renal failure.
Conclusion: Severe clinical deterioration in LVAD patients is not necessarily pump-related. Key elements are perfusion assessment, device and power supply checks, interpretation of pump parameters, extended diagnostics and early consultation with an LVAD center.
期刊介绍:
Medizinische Klinik – Intensivmedizin und Notfallmedizin is an internationally respected interdisciplinary journal. It is intended for physicians, nurses, respiratory and physical therapists active in intensive care and accident/emergency units, but also for internists, anesthesiologists, surgeons, neurologists, and pediatricians with special interest in intensive care medicine.
Comprehensive reviews describe the most recent advances in the field of internal medicine with special focus on intensive care problems. Freely submitted original articles present important studies in this discipline and promote scientific exchange, while articles in the category Photo essay feature interesting cases and aim at optimizing diagnostic and therapeutic strategies. In the rubric journal club well-respected experts comment on outstanding international publications. Review articles under the rubric "Continuing Medical Education" present verified results of scientific research and their integration into daily practice. The rubrics "Nursing practice" and "Physical therapy" round out the information.