Efectos de la termohumidificación sobre la interacción mecánica interfase–superficie facial durante ventilación mecánica no invasiva: un estudio experimental en modelo bench

IF 0.2 Q4 MEDICINE, GENERAL & INTERNAL
Revista Medica Clinica Las Condes Pub Date : 2026-07-01 Epub Date: 2026-09-15 DOI:10.1016/j.rmclc.2026.08.003
Hugo Riquelme, Tomás Salinas, David Wood, Matías Soublett, Bárbara Becerra, Javiera Navarrete
{"title":"Efectos de la termohumidificación sobre la interacción mecánica interfase–superficie facial durante ventilación mecánica no invasiva: un estudio experimental en modelo bench","authors":"Hugo Riquelme,&nbsp;Tomás Salinas,&nbsp;David Wood,&nbsp;Matías Soublett,&nbsp;Bárbara Becerra,&nbsp;Javiera Navarrete","doi":"10.1016/j.rmclc.2026.08.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Noninvasive ventilation (NIV) performance depends on effective mechanical coupling between the interface and the facial surface. However, the effects of thermohumidification on this interaction and its relationship with facial pressure redistribution, system leak, and ventilatory variables remain poorly characterized.</div></div><div><h3>Objective</h3><div>To evaluate the effect of thermohumidification on interface–facial surface mechanical interaction and its relationship with facial pressure redistribution, system leak, and ventilatory variables.</div></div><div><h3>Methods</h3><div>This controlled experimental bench study included 14 independent runs (7 per condition) comparing NIV with and without thermohumidification (35<!--> <!-->°C). Experiments were performed under controlled environmental conditions (20<!--> <!-->°C, 50% relative humidity, phantom surface temperature 36<!--> <!-->°C). Interface–facial surface temperature (IFST), the Pressure Redistribution Index (PRI), system leak, and ventilatory variables were assessed. Group, time, and interaction effects were analyzed using linear mixed-effects models.</div></div><div><h3>Results</h3><div>Thermohumidification increased IFST by 4–5<!--> <!-->°C (p<!--> <!-->&lt;<!--> <!-->0.001), progressively increased PRI (group<!--> <!-->×<!--> <!-->time interaction, p<!--> <!-->&lt;<!--> <!-->0.01), and reduced system leak by 11.5–12.4 l/min. During the final phase of the protocol (45–60<!--> <!-->min), tidal volume increased by 58–61<!--> <!-->ml and peak inspiratory flow by 5.9–6.3 l/min. Post-baseline changes in IFST correlated positively with changes in PRI (ρ=0.81, p<!--> <!-->&lt;<!--> <!-->0.001), whereas changes in PRI correlated inversely with system leak (ρ=<!--> <!-->−<!--> <!-->0.74, p<!--> <!-->=<!--> <!-->0.002).</div></div><div><h3>Conclusions</h3><div>Thermohumidification was associated with facial pressure redistribution, reduced system leak, and changes in ventilatory variables. These findings are consistent with the hypothesis that thermohumidification may influence the mechanical interaction between the interface and the facial surface. However, the underlying mechanisms were not directly evaluated and require confirmation in dedicated experimental and clinical studies.</div></div>","PeriodicalId":31544,"journal":{"name":"Revista Medica Clinica Las Condes","volume":"37 4","pages":"Pages 482-493"},"PeriodicalIF":0.2000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Medica Clinica Las Condes","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0716864026000763","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/9/15 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Noninvasive ventilation (NIV) performance depends on effective mechanical coupling between the interface and the facial surface. However, the effects of thermohumidification on this interaction and its relationship with facial pressure redistribution, system leak, and ventilatory variables remain poorly characterized.

Objective

To evaluate the effect of thermohumidification on interface–facial surface mechanical interaction and its relationship with facial pressure redistribution, system leak, and ventilatory variables.

Methods

This controlled experimental bench study included 14 independent runs (7 per condition) comparing NIV with and without thermohumidification (35 °C). Experiments were performed under controlled environmental conditions (20 °C, 50% relative humidity, phantom surface temperature 36 °C). Interface–facial surface temperature (IFST), the Pressure Redistribution Index (PRI), system leak, and ventilatory variables were assessed. Group, time, and interaction effects were analyzed using linear mixed-effects models.

Results

Thermohumidification increased IFST by 4–5 °C (p < 0.001), progressively increased PRI (group × time interaction, p < 0.01), and reduced system leak by 11.5–12.4 l/min. During the final phase of the protocol (45–60 min), tidal volume increased by 58–61 ml and peak inspiratory flow by 5.9–6.3 l/min. Post-baseline changes in IFST correlated positively with changes in PRI (ρ=0.81, p < 0.001), whereas changes in PRI correlated inversely with system leak (ρ= − 0.74, p = 0.002).

Conclusions

Thermohumidification was associated with facial pressure redistribution, reduced system leak, and changes in ventilatory variables. These findings are consistent with the hypothesis that thermohumidification may influence the mechanical interaction between the interface and the facial surface. However, the underlying mechanisms were not directly evaluated and require confirmation in dedicated experimental and clinical studies.
非侵入式机械通风中热加湿对机械界面-面部相互作用的影响:长凳模型的实验研究
无创通气(NIV)的性能取决于界面和面部表面之间的有效机械耦合。然而,热湿化对这种相互作用的影响及其与面部压力再分布、系统泄漏和通风变量的关系仍然缺乏表征。目的评价热湿化对界面-面-面力学相互作用的影响及其与面部压力重分布、系统泄漏和通气变量的关系。方法在对照实验台上进行了14次独立试验(每种条件7次),比较了35°C热加湿和不加湿的NIV。实验在受控环境条件下进行(20°C, 50%相对湿度,幻影表面温度36°C)。评估了界面表面温度(IFST)、压力再分布指数(PRI)、系统泄漏和通风变量。采用线性混合效应模型分析组效应、时间效应和相互作用效应。结果热加湿使IFST增加4 ~ 5°C (p < 0.001), PRI逐渐增加(组×时间交互作用,p < 0.01),系统泄漏减少11.5 ~ 12.4 l/min。在方案的最后阶段(45-60 min),潮气量增加58-61 ml,峰值吸气流量增加5.9-6.3 l/min。基线后IFST变化与PRI变化呈正相关(ρ=0.81, p < 0.001),而PRI变化与系统泄漏呈负相关(ρ= - 0.74, p = 0.002)。结论热湿化与面部压力重新分布、减少系统泄漏和通气变量的变化有关。这些发现与热湿化可能影响界面和表面之间的机械相互作用的假设是一致的。然而,潜在的机制没有直接评估,需要在专门的实验和临床研究中确认。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Revista Medica Clinica Las Condes
Revista Medica Clinica Las Condes MEDICINE, GENERAL & INTERNAL-
CiteScore
0.80
自引率
0.00%
发文量
65
审稿时长
81 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书