Alex Grafton, Lynn Thomson, Kathryn Beardsall, Joan Lasenby
{"title":"Camera-based respiratory rate monitoring in unmodified neonatal intensive care - a cross-sectional study in a UK hospital.","authors":"Alex Grafton, Lynn Thomson, Kathryn Beardsall, Joan Lasenby","doi":"10.1016/j.eclinm.2026.104175","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Camera-based methods have shown potential for respiratory rate measurement in neonatal intensive care. However existing work is tested on short recordings, which do not represent the range of environments present in the neonatal intensive care unit (NICU) or assess challenging conditions such as movement or poor visibility. This study evaluates a fully automatic respiratory rate measurement system over continuous 24-h periods in the unmodified NICU environment.</p><p><strong>Methods: </strong>In this single-centre cross-sectional study, 20 babies in the NICU in Addenbrooke's Hospital, Cambridge, UK were recorded using a colour + depth camera, each for around 24 h. Recordings took place between 11 July 2022 and 11 November 2025. Inclusion criteria were: pre-term and nursed in closed incubators; no specific exclusion criteria applied. Existing methods were used to automatically locate the baby, and we used a novel clustering method to extract a respiratory signal and detect the patient's breaths. The primary outcome was the measurement of respiratory rate using the camera, assessed by comparison with data from the clinical standard patient monitor, recorded concurrently. Further post-hoc analysis was conducted on clinical and demographic subsets. This trial is registered on clinicaltrials.gov, ID NCT04831242.</p><p><strong>Findings: </strong>After excluding intervals where the camera or baby was removed from the incubator, 380 h of valid recording from 20 participants remained. During non-intervention periods (338 h, 77.6% of all, 88.7% of valid), we configured the patient monitor signal quality indices to 57.9% valid measurement, a likely overestimate; the camera achieved 57.8% valid measurement with 1.44 breaths/minute agreement (mean error 0.42) averaged over all periods where both camera and monitor had valid data (136 h, 40.4% of non-intervention, 31.4% of all). Covering affects the valid time (63.6% when uncovered, 51.3% when covered), as do intervals where the baby is more active (61.4% when inactive, 53.5% when active). The error and valid time were worse when the monitor respiratory rate was below 35 breaths/minute (31.0% vs 57.8% valid, 1.98 vs 1.44 error) and more so below 25 breaths/minute (25.6% valid, 7.57 error), though data in this range was limited (0.2% of non-intervention) and the patient monitor may be unreliable.</p><p><strong>Interpretation: </strong>Respiratory rate measurement using cameras has potential for continuous use in the NICU, beyond previously studied short recordings, achieving mean absolute error (MAE) averaged across all data below 2 breaths/minute while providing valid data 57.8% of the time. The relationship between covering/activity and valid measurement time shows the importance of long recordings in the real-world NICU. The MAE and valid time was worse for babies under 1000 g, requiring continuous positive airway pressure, or when the respiratory rate is below 35 breaths/minute. Further work should focus on bradypnoea/apnoea detection, the aforementioned demographic groups, and results should be confirmed in a cohort with greater ethnic diversity.</p><p><strong>Funding: </strong>Rosetrees Trust, Stoneygate Trust, Isaac Newton Trust, EPSRC Impact Acceleration Account.</p>","PeriodicalId":11393,"journal":{"name":"EClinicalMedicine","volume":"99 ","pages":"104175"},"PeriodicalIF":12.8000,"publicationDate":"2026-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13562425/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EClinicalMedicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.eclinm.2026.104175","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/9/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Camera-based methods have shown potential for respiratory rate measurement in neonatal intensive care. However existing work is tested on short recordings, which do not represent the range of environments present in the neonatal intensive care unit (NICU) or assess challenging conditions such as movement or poor visibility. This study evaluates a fully automatic respiratory rate measurement system over continuous 24-h periods in the unmodified NICU environment.
Methods: In this single-centre cross-sectional study, 20 babies in the NICU in Addenbrooke's Hospital, Cambridge, UK were recorded using a colour + depth camera, each for around 24 h. Recordings took place between 11 July 2022 and 11 November 2025. Inclusion criteria were: pre-term and nursed in closed incubators; no specific exclusion criteria applied. Existing methods were used to automatically locate the baby, and we used a novel clustering method to extract a respiratory signal and detect the patient's breaths. The primary outcome was the measurement of respiratory rate using the camera, assessed by comparison with data from the clinical standard patient monitor, recorded concurrently. Further post-hoc analysis was conducted on clinical and demographic subsets. This trial is registered on clinicaltrials.gov, ID NCT04831242.
Findings: After excluding intervals where the camera or baby was removed from the incubator, 380 h of valid recording from 20 participants remained. During non-intervention periods (338 h, 77.6% of all, 88.7% of valid), we configured the patient monitor signal quality indices to 57.9% valid measurement, a likely overestimate; the camera achieved 57.8% valid measurement with 1.44 breaths/minute agreement (mean error 0.42) averaged over all periods where both camera and monitor had valid data (136 h, 40.4% of non-intervention, 31.4% of all). Covering affects the valid time (63.6% when uncovered, 51.3% when covered), as do intervals where the baby is more active (61.4% when inactive, 53.5% when active). The error and valid time were worse when the monitor respiratory rate was below 35 breaths/minute (31.0% vs 57.8% valid, 1.98 vs 1.44 error) and more so below 25 breaths/minute (25.6% valid, 7.57 error), though data in this range was limited (0.2% of non-intervention) and the patient monitor may be unreliable.
Interpretation: Respiratory rate measurement using cameras has potential for continuous use in the NICU, beyond previously studied short recordings, achieving mean absolute error (MAE) averaged across all data below 2 breaths/minute while providing valid data 57.8% of the time. The relationship between covering/activity and valid measurement time shows the importance of long recordings in the real-world NICU. The MAE and valid time was worse for babies under 1000 g, requiring continuous positive airway pressure, or when the respiratory rate is below 35 breaths/minute. Further work should focus on bradypnoea/apnoea detection, the aforementioned demographic groups, and results should be confirmed in a cohort with greater ethnic diversity.
Funding: Rosetrees Trust, Stoneygate Trust, Isaac Newton Trust, EPSRC Impact Acceleration Account.
背景:基于相机的方法在新生儿重症监护中显示出呼吸频率测量的潜力。然而,现有的工作是在短记录上进行测试的,这并不能代表新生儿重症监护病房(NICU)中存在的环境范围,也不能评估诸如运动或能见度低等具有挑战性的条件。本研究在未经修改的NICU环境中对全自动呼吸速率测量系统进行了连续24小时的评估。方法:在这项单中心横断面研究中,使用彩色+深度摄像机对英国剑桥阿登布鲁克医院NICU的20名婴儿进行记录,每次记录约24小时。记录时间为2022年7月11日至2025年11月11日。纳入标准为:早产儿和封闭式保温箱护理;没有具体的排除标准。利用现有的方法对婴儿进行自动定位,并采用一种新颖的聚类方法提取呼吸信号并检测患者的呼吸。主要结果是使用相机测量呼吸频率,通过与同时记录的临床标准患者监护仪的数据进行比较来评估。对临床和人口亚群进行了进一步的事后分析。该试验已在clinicaltrials.gov注册,编号NCT04831242。结果:在排除相机或婴儿从培养箱中取出的间隔后,20名参与者的380小时有效记录仍然存在。在非干预期(338小时,77.6%,88.7%有效),我们将患者监护信号质量指标配置为57.9%有效测量,可能高估;在摄像机和监测器均有有效数据的所有时间段(136小时,40.4%的非干预,31.4%的所有时间段),摄像机的有效测量率为57.8%,一致性为1.44次/分钟(平均误差0.42)。覆盖影响有效时间(未覆盖时为63.6%,覆盖时为51.3%),婴儿更活跃的间隔时间(不活动时为61.4%,活动时为53.5%)也是如此。当监测呼吸频率低于35次/分钟时(31.0% vs 57.8%,有效误差1.98 vs 1.44),当监测呼吸频率低于25次/分钟时(25.6%,有效误差7.57),误差和有效时间更大,尽管该范围内的数据有限(0.2%的非干预),患者监护仪可能不可靠。解释:使用摄像机测量呼吸速率在新生儿重症监护病房中具有连续使用的潜力,超越了先前研究的短记录,在所有低于2次呼吸/分钟的数据中实现平均绝对误差(MAE),同时提供57.8%的有效数据。覆盖/活动与有效测量时间之间的关系显示了在现实NICU中长记录的重要性。对于小于1000 g、需要持续气道正压或呼吸频率低于35次/分钟的婴儿,MAE和有效时间更差。进一步的工作应该集中在呼吸缓慢/呼吸暂停的检测上,上述人口群体,结果应该在一个种族多样性更大的队列中得到证实。资助:Rosetrees信托,Stoneygate信托,Isaac Newton信托,EPSRC影响加速账户。
期刊介绍:
eClinicalMedicine is a gold open-access clinical journal designed to support frontline health professionals in addressing the complex and rapid health transitions affecting societies globally. The journal aims to assist practitioners in overcoming healthcare challenges across diverse communities, spanning diagnosis, treatment, prevention, and health promotion. Integrating disciplines from various specialties and life stages, it seeks to enhance health systems as fundamental institutions within societies. With a forward-thinking approach, eClinicalMedicine aims to redefine the future of healthcare.