Sagarika Jain, Ashlee Ewing, Michael Weldon, Allison Palmiero, Daniel Christ, Emile Gogineni, Simeng Zhu, Sung Jun Ma, John C Grecula, Darrion L Mitchell, Sujith Baliga, Dukagjin M Blakaj, David J Konieczkowski
{"title":"Density override-based VMAT optimization to improve robustness at tissue-air interfaces in head and neck radiotherapy","authors":"Sagarika Jain, Ashlee Ewing, Michael Weldon, Allison Palmiero, Daniel Christ, Emile Gogineni, Simeng Zhu, Sung Jun Ma, John C Grecula, Darrion L Mitchell, Sujith Baliga, Dukagjin M Blakaj, David J Konieczkowski","doi":"10.1002/acm2.70737","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Volumetric Modulated Arc Therapy (VMAT) is integral to head and neck (HN) radiation therapy; however, plan robustness may be compromised at air-tissue interfaces. Highly modulated fluence directed at these areas can produce significant unexpected high-dose regions in the event of minor variations in tissue density.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This study characterizes an instability in VMAT planning for HN radiation therapy (RT) occurring at the interface between target volume (e.g. mucosal primary tumor) and internal air spaces (e.g. pharyngeal lumen, sinuses, nasal cavity, etc.). A density-override based planning technique is presented and validated to mitigate this instability.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Treatment plans for fifteen HN patients receiving VMAT RT at our institution between 2023 and 2024 were retrospectively analyzed. All patients received 60–69.96 Gy in 30–33 daily fractions using 6 MV photons. To model swelling or tumor growth near the air–tumor interface, verification plans were generated by overriding internal air within 3 mm of the planning target volume (PTV) to soft tissue density (HU = 0). A robustness-enhancing technique was evaluated, wherein optimization was performed with luminal air overridden to HU = –300, followed by recalculation and renormalization using the original CT HU values. Robustness was then reassessed after overriding luminal air to HU = 0.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Baseline plans demonstrated controlled hotspot magnitudes within the BODY, with a median D0.1cc of 109.0% [Interquartile range: 107.8–110.4]. Simulated tissue-air interface changes produced substantially elevated hotspots in verification plans, increasing to a median D0.1cc of 115.6% [111.9–126.9], corresponding to a median paired increase of 9.0% (Wilcoxon <i>p</i> = 1.22 × 10<sup>−</sup><sup>4</sup>). Robust plans generated using the proposed density-override optimization maintained comparable baseline dosimetry (median D0.1cc 108.1% [107.4–108.7]) while demonstrating stable hotspot behavior under simulated density perturbations (median D0.1cc 108.1% [107.1–108.6], <i>p</i> = 0.18). Target coverage remained preserved, with robust verification plan D95 values ranging from 99.7% to 100.5% (<i>p</i> = 0.26).</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>A density-override optimization approach mitigates hotspot formation at air–tumor interfaces, producing VMAT plans that remain dosimetrically stable despite variation at internal air-tumor interfaces.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 9","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70737","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70737","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Volumetric Modulated Arc Therapy (VMAT) is integral to head and neck (HN) radiation therapy; however, plan robustness may be compromised at air-tissue interfaces. Highly modulated fluence directed at these areas can produce significant unexpected high-dose regions in the event of minor variations in tissue density.
Purpose
This study characterizes an instability in VMAT planning for HN radiation therapy (RT) occurring at the interface between target volume (e.g. mucosal primary tumor) and internal air spaces (e.g. pharyngeal lumen, sinuses, nasal cavity, etc.). A density-override based planning technique is presented and validated to mitigate this instability.
Methods
Treatment plans for fifteen HN patients receiving VMAT RT at our institution between 2023 and 2024 were retrospectively analyzed. All patients received 60–69.96 Gy in 30–33 daily fractions using 6 MV photons. To model swelling or tumor growth near the air–tumor interface, verification plans were generated by overriding internal air within 3 mm of the planning target volume (PTV) to soft tissue density (HU = 0). A robustness-enhancing technique was evaluated, wherein optimization was performed with luminal air overridden to HU = –300, followed by recalculation and renormalization using the original CT HU values. Robustness was then reassessed after overriding luminal air to HU = 0.
Results
Baseline plans demonstrated controlled hotspot magnitudes within the BODY, with a median D0.1cc of 109.0% [Interquartile range: 107.8–110.4]. Simulated tissue-air interface changes produced substantially elevated hotspots in verification plans, increasing to a median D0.1cc of 115.6% [111.9–126.9], corresponding to a median paired increase of 9.0% (Wilcoxon p = 1.22 × 10−4). Robust plans generated using the proposed density-override optimization maintained comparable baseline dosimetry (median D0.1cc 108.1% [107.4–108.7]) while demonstrating stable hotspot behavior under simulated density perturbations (median D0.1cc 108.1% [107.1–108.6], p = 0.18). Target coverage remained preserved, with robust verification plan D95 values ranging from 99.7% to 100.5% (p = 0.26).
Conclusion
A density-override optimization approach mitigates hotspot formation at air–tumor interfaces, producing VMAT plans that remain dosimetrically stable despite variation at internal air-tumor interfaces.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic