Differential Air Entrainment of Mini-Basket and Circular Mapping Catheters Through the FARADRIVE Sheath: Mitigation With a Liquid-Sealed Insertion Method
{"title":"Differential Air Entrainment of Mini-Basket and Circular Mapping Catheters Through the FARADRIVE Sheath: Mitigation With a Liquid-Sealed Insertion Method","authors":"Thai Duy Vo, Ting-Yung Chang, Chin-Yu Lin, Pei-Yi Wang, Kuo-Hsiang Lu, Hsin-Lei Huang","doi":"10.1111/jce.70492","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Iatrogenic air entrainment is a preventable embolic hazard in pulsed field ablation (PFA). Mapping catheter insertion through large-bore PFA sheaths may be a vulnerable step, but the roles of catheter geometry and liquid-sealed insertion are unclear.</p>\n </section>\n \n <section>\n \n <h3> Methods and Results</h3>\n \n <p>In an ex vivo saline tank model, ORION mini-basket and SPIRAL circular mapping catheters were inserted through the FARADRIVE steerable sheath using saline infusion only, continuous manual suction, or liquid-sealed insertion with a custom Adapter Sink. The primary endpoint was macroscopic air recovered distally after each trial of 10 insertions; the trial was the unit of analysis, and 0.2 mL was the smallest quantitatively reportable volume. In 60 trials (600 insertions), mitigation strategy significantly affected air entrainment (Kruskal–Wallis <i>p</i> < 0.001). Under saline infusion alone, ORION generated more air than SPIRAL (23.86 ± 3.63 vs. 7.20 ± 2.09 mL per 10 insertions; 2.39 vs. 0.72 mL per insertion; adjusted <i>p</i> < 0.001). Manual suction reduced entrainment to 3.96 ± 1.01 and 1.02 ± 0.45 mL per 10 insertions (0.40 and 0.10 mL per insertion), but residual air persisted. With liquid-sealed insertion, no visually detectable or quantitatively measurable macroscopic air was observed in any trial.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Air entrainment varied according to mapping-catheter geometry and mitigation strategy. Timed manual suction substantially reduced, but did not abolish, measurable air. Liquid-sealed insertion was associated with no measurable macroscopic air within this ex vivo model, warranting further evaluation under physiologic pressure conditions and in clinical PFA workflows.</p>\n </section>\n </div>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":"37 10","pages":"2317-2325"},"PeriodicalIF":2.4000,"publicationDate":"2026-10-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jce.70492","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/9/8 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Iatrogenic air entrainment is a preventable embolic hazard in pulsed field ablation (PFA). Mapping catheter insertion through large-bore PFA sheaths may be a vulnerable step, but the roles of catheter geometry and liquid-sealed insertion are unclear.
Methods and Results
In an ex vivo saline tank model, ORION mini-basket and SPIRAL circular mapping catheters were inserted through the FARADRIVE steerable sheath using saline infusion only, continuous manual suction, or liquid-sealed insertion with a custom Adapter Sink. The primary endpoint was macroscopic air recovered distally after each trial of 10 insertions; the trial was the unit of analysis, and 0.2 mL was the smallest quantitatively reportable volume. In 60 trials (600 insertions), mitigation strategy significantly affected air entrainment (Kruskal–Wallis p < 0.001). Under saline infusion alone, ORION generated more air than SPIRAL (23.86 ± 3.63 vs. 7.20 ± 2.09 mL per 10 insertions; 2.39 vs. 0.72 mL per insertion; adjusted p < 0.001). Manual suction reduced entrainment to 3.96 ± 1.01 and 1.02 ± 0.45 mL per 10 insertions (0.40 and 0.10 mL per insertion), but residual air persisted. With liquid-sealed insertion, no visually detectable or quantitatively measurable macroscopic air was observed in any trial.
Conclusions
Air entrainment varied according to mapping-catheter geometry and mitigation strategy. Timed manual suction substantially reduced, but did not abolish, measurable air. Liquid-sealed insertion was associated with no measurable macroscopic air within this ex vivo model, warranting further evaluation under physiologic pressure conditions and in clinical PFA workflows.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.