Effects of spiral deflation and warming on cuff diameter, insertion force, extubation, and sealing performance in polyvinyl chloride-cuffed tracheal tubes.

IF 1.5 4区 医学 Q3 ANESTHESIOLOGY
Hiroki Yonezawa, Michiko Kinoshita, Chiaki Murakami, Takako Ikezaki, Katsuya Tanaka
{"title":"Effects of spiral deflation and warming on cuff diameter, insertion force, extubation, and sealing performance in polyvinyl chloride-cuffed tracheal tubes.","authors":"Hiroki Yonezawa, Michiko Kinoshita, Chiaki Murakami, Takako Ikezaki, Katsuya Tanaka","doi":"10.1177/0310057X261478645","DOIUrl":null,"url":null,"abstract":"<p><p>Cuffed tracheal tubes are widely used in paediatric airway management, but passage through the relatively narrow subglottic airway may increase insertion resistance and risk mucosal trauma. This study evaluated whether spiral deflation (i.e. folding the cuff inward like an umbrella) and warming the tube could reduce cuff diameter and insertion force in a simulated trachea. Four commercially available polyvinyl chloride-cuffed tubes with an internal diameter of 5.0 mm (PORTEX<sup>®</sup> (Smiths Medical International Ltd, Ashford, UK), Sheridan<sup>®</sup> (Teleflex Inc., Morrisville, NC, USA), Shiley™ (Medtronic, Minneapolis, MN, USA), and VentiSeal™ (Flexicare Medical Ltd, Mountain Ash, UK)) were examined. Spiral deflation significantly reduced cuff diameter, with reductions of (mean difference, MD) -3.6 mm (95% confidence intervals, CI -4.5 to -2.7, <i>P</i> <0.001), -2.8 mm (-4.0 to -1.6, <i>P</i> <0.001), -3.5 mm (-5.0 to -2.1, <i>P</i>=0.001), and -4.9 mm (-6.0 to -3.9, <i>P</i> <0.001), respectively. Spiral deflation also decreased insertion force, with reductions of (MD (CI)) -1.01 N (-1.52 to -0.50, <i>P</i>=0.008), -0.34 N (-0.63 to -0.05, <i>P</i>=0.115), -0.49 N (-0.61 to -0.37, <i>P</i> <0.001), and -2.33 N (-3.83 to -0.84, <i>P</i>=0.028), respectively. Additional warming at 40°C further lowered insertion force, with reductions of (MD (CI)) -0.12 N (-0.19 to -0.05, <i>P</i>=0.020), -0.06 N (-0.08 to -0.03, <i>P</i>=0.002), -0.07 N (-0.11 to -0.03, <i>P</i>=0.010), and -0.48 N (-0.86 to -0.11, <i>P</i>=0.068), respectively. Spiral deflation with warming did not affect extubation force or sealing performance. Overall, spiral deflation effectively reduced cuff diameter by approximately 25% and lowered insertion force, while warming further enhanced ease of insertion. These approaches did not compromise sealing or extubation performance.</p>","PeriodicalId":7746,"journal":{"name":"Anaesthesia and Intensive Care","volume":" ","pages":"310057X261478645"},"PeriodicalIF":1.5000,"publicationDate":"2026-09-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Anaesthesia and Intensive Care","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/0310057X261478645","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Cuffed tracheal tubes are widely used in paediatric airway management, but passage through the relatively narrow subglottic airway may increase insertion resistance and risk mucosal trauma. This study evaluated whether spiral deflation (i.e. folding the cuff inward like an umbrella) and warming the tube could reduce cuff diameter and insertion force in a simulated trachea. Four commercially available polyvinyl chloride-cuffed tubes with an internal diameter of 5.0 mm (PORTEX® (Smiths Medical International Ltd, Ashford, UK), Sheridan® (Teleflex Inc., Morrisville, NC, USA), Shiley™ (Medtronic, Minneapolis, MN, USA), and VentiSeal™ (Flexicare Medical Ltd, Mountain Ash, UK)) were examined. Spiral deflation significantly reduced cuff diameter, with reductions of (mean difference, MD) -3.6 mm (95% confidence intervals, CI -4.5 to -2.7, P <0.001), -2.8 mm (-4.0 to -1.6, P <0.001), -3.5 mm (-5.0 to -2.1, P=0.001), and -4.9 mm (-6.0 to -3.9, P <0.001), respectively. Spiral deflation also decreased insertion force, with reductions of (MD (CI)) -1.01 N (-1.52 to -0.50, P=0.008), -0.34 N (-0.63 to -0.05, P=0.115), -0.49 N (-0.61 to -0.37, P <0.001), and -2.33 N (-3.83 to -0.84, P=0.028), respectively. Additional warming at 40°C further lowered insertion force, with reductions of (MD (CI)) -0.12 N (-0.19 to -0.05, P=0.020), -0.06 N (-0.08 to -0.03, P=0.002), -0.07 N (-0.11 to -0.03, P=0.010), and -0.48 N (-0.86 to -0.11, P=0.068), respectively. Spiral deflation with warming did not affect extubation force or sealing performance. Overall, spiral deflation effectively reduced cuff diameter by approximately 25% and lowered insertion force, while warming further enhanced ease of insertion. These approaches did not compromise sealing or extubation performance.

螺旋放气和加热对聚氯乙烯套管袖口直径、插入力、拔管和密封性能的影响。
带口气管管广泛应用于儿科气道管理,但通过相对狭窄的声门下气道可能增加插入阻力和粘膜损伤的风险。本研究评估了螺旋放气(即像雨伞一样向内折叠袖带)和加热管是否可以减少袖带直径和在模拟气管中的插入力。检测了四根内径为5.0 mm的市售聚氯乙烯套管(PORTEX®(Smiths Medical International Ltd, Ashford, UK)、Sheridan®(Teleflex Inc., Morrisville, NC, USA)、Shiley™(Medtronic, Minneapolis, MN, USA)和VentiSeal™(Flexicare Medical Ltd, Mountain Ash, UK))。螺旋充气显著降低袖带直径,(平均差值,MD)分别减少3.6 mm(95%可信区间,CI -4.5至-2.7,P P=0.001), -4.9 mm(-6.0至-3.9,P P=0.008), -0.34 N(-0.63至-0.05,P=0.115), -0.49 N(-0.61至-0.37,P P=0.028)。在40°C的额外升温进一步降低了插入力,分别降低了(MD (CI)) -0.12 N(-0.19至-0.05,P=0.020)、-0.06 N(-0.08至-0.03,P=0.002)、-0.07 N(-0.11至-0.03,P=0.010)和-0.48 N(-0.86至-0.11,P=0.068)。加温螺旋式放气不影响拔管力和密封性能。总的来说,螺旋放气有效地减少了袖带直径约25%,降低了插入力,而加热进一步提高了插入的便利性。这些方法不影响密封或拔管性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.70
自引率
13.30%
发文量
150
审稿时长
3 months
期刊介绍: Anaesthesia and Intensive Care is an international journal publishing timely, peer reviewed articles that have educational value and scientific merit for clinicians and researchers associated with anaesthesia, intensive care medicine, and pain medicine.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书