A real-time patient orientation verification system using pose estimation in CT simulation

IF 1 4区 物理与天体物理 Q3 PHYSICS, MULTIDISCIPLINARY
Hyunji Amy Kim, Elliott Choi, Estella Choi, Youngjin Lee, Daegun Kim, Jae-Hong Jung, Kyung-Bae Lee, James J. Sohn
{"title":"A real-time patient orientation verification system using pose estimation in CT simulation","authors":"Hyunji Amy Kim,&nbsp;Elliott Choi,&nbsp;Estella Choi,&nbsp;Youngjin Lee,&nbsp;Daegun Kim,&nbsp;Jae-Hong Jung,&nbsp;Kyung-Bae Lee,&nbsp;James J. Sohn","doi":"10.1007/s40042-026-01664-y","DOIUrl":null,"url":null,"abstract":"<div><p>Patient orientation mis-entry during computed tomography (CT) simulation represents a critical failure mode that corrupts the digital imaging and communications in medicine (DICOM) coordinate system, necessitating patient recall for re-scanning or complex manual corrections. We implemented a tool called just another secondary evaluation (JASE), an automated real-time verification system to prevent these errors. JASE uses a standard webcam to classify patient orientation into four standard categories: head-first-supine (HFS), head-first-prone (HFP), feet-first-supine (FFS), and feet-first-prone (FFP). We compared two approaches: transfer learning with MobileNetV2 and a multilayer perceptron (MLP) using MediaPipe pose landmarks. The system displays real-time orientation on an LCD screen adjacent to the CT console. The MLP model achieved 98.3% accuracy (1/49 misclassifications) versus 96.2% for MobileNetV2 (3/80 misclassifications) on an independent test set. Real-time implementation demonstrated stable performance, which indicates the potential for future clinical deployment pending validation in actual CT simulation workflows. JASE provides a low-cost, effective solution for preventing orientation errors at the point of entry. The system’s applicability extends beyond radiation oncology to radiology and nuclear medicine departments using similar imaging protocols.</p></div>","PeriodicalId":677,"journal":{"name":"Journal of the Korean Physical Society","volume":"89 6","pages":"677 - 685"},"PeriodicalIF":1.0000,"publicationDate":"2026-06-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Korean Physical Society","FirstCategoryId":"101","ListUrlMain":"https://link.springer.com/article/10.1007/s40042-026-01664-y","RegionNum":4,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PHYSICS, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0

Abstract

Patient orientation mis-entry during computed tomography (CT) simulation represents a critical failure mode that corrupts the digital imaging and communications in medicine (DICOM) coordinate system, necessitating patient recall for re-scanning or complex manual corrections. We implemented a tool called just another secondary evaluation (JASE), an automated real-time verification system to prevent these errors. JASE uses a standard webcam to classify patient orientation into four standard categories: head-first-supine (HFS), head-first-prone (HFP), feet-first-supine (FFS), and feet-first-prone (FFP). We compared two approaches: transfer learning with MobileNetV2 and a multilayer perceptron (MLP) using MediaPipe pose landmarks. The system displays real-time orientation on an LCD screen adjacent to the CT console. The MLP model achieved 98.3% accuracy (1/49 misclassifications) versus 96.2% for MobileNetV2 (3/80 misclassifications) on an independent test set. Real-time implementation demonstrated stable performance, which indicates the potential for future clinical deployment pending validation in actual CT simulation workflows. JASE provides a low-cost, effective solution for preventing orientation errors at the point of entry. The system’s applicability extends beyond radiation oncology to radiology and nuclear medicine departments using similar imaging protocols.

CT仿真中基于位姿估计的实时患者方向验证系统
在计算机断层扫描(CT)模拟过程中,患者方向输入错误是一种严重的故障模式,它破坏了医学数字成像和通信(DICOM)坐标系,需要召回患者进行重新扫描或复杂的手动更正。我们实现了一个工具,称为另一个二次评估(JASE),一个自动的实时验证系统,以防止这些错误。JASE使用一个标准的网络摄像头将患者的取向分为四个标准类别:头先仰卧(HFS)、头先俯卧(HFP)、脚先仰卧(FFS)和脚先俯卧(FFP)。我们比较了两种方法:使用MobileNetV2的迁移学习和使用MediaPipe姿势地标的多层感知器(MLP)。该系统在CT控制台旁边的LCD屏幕上显示实时方位。MLP模型在独立测试集上的准确率为98.3%(1/49错误分类),而MobileNetV2的准确率为96.2%(3/80错误分类)。实时实现显示出稳定的性能,这表明在实际CT模拟工作流程中等待验证的未来临床部署的潜力。JASE提供了一种低成本、有效的解决方案,用于防止进入点的方向错误。该系统的适用性从放射肿瘤学扩展到使用类似成像协议的放射学和核医学部门。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of the Korean Physical Society
Journal of the Korean Physical Society PHYSICS, MULTIDISCIPLINARY-
CiteScore
1.20
自引率
16.70%
发文量
276
审稿时长
5.5 months
期刊介绍: The Journal of the Korean Physical Society (JKPS) covers all fields of physics spanning from statistical physics and condensed matter physics to particle physics. The manuscript to be published in JKPS is required to hold the originality, significance, and recent completeness. The journal is composed of Full paper, Letters, and Brief sections. In addition, featured articles with outstanding results are selected by the Editorial board and introduced in the online version. For emphasis on aspect of international journal, several world-distinguished researchers join the Editorial board. High quality of papers may be express-published when it is recommended or requested.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书