Sake Salverda, Ernst van der Wal, Thomas Toet, Mischa Hoogeman, Kelvin Ng Wei Siang
{"title":"Impact of collimator misalignments on dosimetric robustness in proton minibeam radiotherapy using an automated measurement setup","authors":"Sake Salverda, Ernst van der Wal, Thomas Toet, Mischa Hoogeman, Kelvin Ng Wei Siang","doi":"10.1002/mp.70671","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Proton minibeam radiation therapy is an emerging technique that combines the tissue-sparing benefits of protons with the spatial modulation used for minibeams to reduce toxicity in healthy tissue. Achieving accurate delivery of planned dose distributions depends on precise collimator-to-beam alignment, for which it is essential to understand the behavior of both multi-slit and single-slit configurations. Understanding single-slit behavior is particularly important, as it also linearly affects the heterogeneity (peaks and valleys) of the overall minibeam. While several studies have investigated misalignment effects with multi-slit collimators, there is limited experimental characterisation of single-slit behavior and its impact on multi-slit arrays.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This study aims to quantify the dosimetric effects of translational and rotational collimator-to-beam misalignments for single-slit proton minibeams using our in-house automated measurement setup. Additionally, the robustness of lateral dose profiles for minibeam arrays will be investigated under varying collimator misalignments.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A single slit collimator with <span></span><math>\n <semantics>\n <mrow>\n <mn>150</mn>\n <mspace></mspace>\n <mi>M</mi>\n <mi>e</mi>\n <mi>V</mi>\n </mrow>\n <annotation>$150 \\,\\mathrm{M}\\mathrm{e\\mathrm{V}}$</annotation>\n </semantics></math> protons was used to produce a single minibeam with adjustable collimator slit widths (0.4 and <span></span><math>\n <semantics>\n <mrow>\n <mn>0.7</mn>\n <mspace></mspace>\n <mi>m</mi>\n <mi>m</mi>\n </mrow>\n <annotation>$0.7 \\,\\mathrm{m}\\mathrm{m}$</annotation>\n </semantics></math>). We introduced rotational and translational misalignments using motorized stages. We developed a quick measurement setup to measure lateral profiles using a microDiamond detector. The lateral profiles were measured at depths of 1 and <span></span><math>\n <semantics>\n <mrow>\n <mn>13</mn>\n <mspace></mspace>\n <mi>c</mi>\n <mi>m</mi>\n </mrow>\n <annotation>$13 \\,\\mathrm{c}\\mathrm{m}$</annotation>\n </semantics></math> using a water-equivalent phantom. We validated the setup by comparing the measurements to gafchromic film measurements using local gamma analysis. A custom fitting model combining two Gaussians and a Lorentzian was developed to fit the lateral profiles. From the fitted profiles, we determined the maximum dose, the lateral position of maximum dose, the full width at half maximum, and the symmetry. Minibeam arrays were simulated via superposition of single-slit profiles. Specific misalignment combinations were measured for 140, 160, and <span></span><math>\n <semantics>\n <mrow>\n <mn>170</mn>\n <mspace></mspace>\n <mi>M</mi>\n <mi>e</mi>\n <mi>V</mi>\n </mrow>\n <annotation>$170 \\,\\mathrm{M}\\mathrm{e\\mathrm{V}}$</annotation>\n </semantics></math> protons, energies typically used in the clinic for deep seated tumours, to assess how misalignments affect the minibeam at different energies.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Rotational misalignments caused substantial asymmetry and peak dose reductions up to 68.4% at shallow depths, with asymmetry reducing at greater depths. Translational shifts resulted in a translation of the minibeam profile in the same direction and magnitude without changing the minibeams shape. Despite these distortions, simulated minibeam arrays achieved acceptable dose homogeneity at <span></span><math>\n <semantics>\n <mrow>\n <mn>13</mn>\n <mspace></mspace>\n <mi>c</mi>\n <mi>m</mi>\n </mrow>\n <annotation>$13 \\,\\mathrm{c}\\mathrm{m}$</annotation>\n </semantics></math> depth for central 80% field-widths with flatness below a 3% threshold. The proposed measurement setup enabled profile acquisition of a single minibeam in under one minute with submillimetre resolution and strong agreement with gafchromic film (3%/<span></span><math>\n <semantics>\n <mrow>\n <mn>0.2</mn>\n <mspace></mspace>\n <mi>m</mi>\n <mi>m</mi>\n </mrow>\n <annotation>$0.2 \\,\\mathrm{m}\\mathrm{m}$</annotation>\n </semantics></math> gamma pass rates of 100%).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Collimator misalignments substantially affect the minibeam dose robustness and should be considered in clinical treatment planning and quality assurance (QA). The presented measurement setup enables, rapid, high-resolution profile measurements using commercially available detectors and can facilitate routine clinical QA.</p>\n </section>\n </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"53 9","pages":""},"PeriodicalIF":3.2000,"publicationDate":"2026-09-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13548184/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/mp.70671","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Proton minibeam radiation therapy is an emerging technique that combines the tissue-sparing benefits of protons with the spatial modulation used for minibeams to reduce toxicity in healthy tissue. Achieving accurate delivery of planned dose distributions depends on precise collimator-to-beam alignment, for which it is essential to understand the behavior of both multi-slit and single-slit configurations. Understanding single-slit behavior is particularly important, as it also linearly affects the heterogeneity (peaks and valleys) of the overall minibeam. While several studies have investigated misalignment effects with multi-slit collimators, there is limited experimental characterisation of single-slit behavior and its impact on multi-slit arrays.
Purpose
This study aims to quantify the dosimetric effects of translational and rotational collimator-to-beam misalignments for single-slit proton minibeams using our in-house automated measurement setup. Additionally, the robustness of lateral dose profiles for minibeam arrays will be investigated under varying collimator misalignments.
Methods
A single slit collimator with protons was used to produce a single minibeam with adjustable collimator slit widths (0.4 and ). We introduced rotational and translational misalignments using motorized stages. We developed a quick measurement setup to measure lateral profiles using a microDiamond detector. The lateral profiles were measured at depths of 1 and using a water-equivalent phantom. We validated the setup by comparing the measurements to gafchromic film measurements using local gamma analysis. A custom fitting model combining two Gaussians and a Lorentzian was developed to fit the lateral profiles. From the fitted profiles, we determined the maximum dose, the lateral position of maximum dose, the full width at half maximum, and the symmetry. Minibeam arrays were simulated via superposition of single-slit profiles. Specific misalignment combinations were measured for 140, 160, and protons, energies typically used in the clinic for deep seated tumours, to assess how misalignments affect the minibeam at different energies.
Results
Rotational misalignments caused substantial asymmetry and peak dose reductions up to 68.4% at shallow depths, with asymmetry reducing at greater depths. Translational shifts resulted in a translation of the minibeam profile in the same direction and magnitude without changing the minibeams shape. Despite these distortions, simulated minibeam arrays achieved acceptable dose homogeneity at depth for central 80% field-widths with flatness below a 3% threshold. The proposed measurement setup enabled profile acquisition of a single minibeam in under one minute with submillimetre resolution and strong agreement with gafchromic film (3%/ gamma pass rates of 100%).
Conclusions
Collimator misalignments substantially affect the minibeam dose robustness and should be considered in clinical treatment planning and quality assurance (QA). The presented measurement setup enables, rapid, high-resolution profile measurements using commercially available detectors and can facilitate routine clinical QA.
背景:质子微束放射治疗是一种新兴的技术,结合了质子的组织保护优势和用于微束的空间调制以减少健康组织的毒性。实现计划剂量分布的准确传递取决于精确的准直器-光束对准,因此了解多缝和单缝配置的行为至关重要。理解单缝的行为是特别重要的,因为它也线性地影响整个微光束的非均质性(峰和谷)。虽然有一些研究研究了多缝准直器的错位效应,但对单缝行为及其对多缝阵列影响的实验表征有限。目的:本研究旨在利用我们内部的自动测量装置,量化单狭缝质子微束的平移和旋转准直-光束错位的剂量学效应。此外,在不同准直器失调的情况下,将研究微束阵列侧剂量分布的鲁棒性。方法:利用150 μ Me V质子的单缝准直器产生可调节准直缝宽度(0.4和0.7 μ M)的单束微光束。我们引入了旋转和平移错位使用机动阶段。我们开发了一种快速测量装置,使用微金刚石探测器测量横向剖面。横向剖面在1和13厘米的深度使用水等效模体测量。我们通过将测量值与使用局部伽玛分析的变色薄膜测量值进行比较来验证该设置。结合两个高斯函数和一个洛伦兹函数的定制拟合模型被开发来拟合横向剖面。根据拟合的剖面,我们确定了最大剂量、最大剂量的横向位置、半最大时的全宽度和对称性。通过单狭缝剖面的叠加模拟了微光束阵列。我们测量了140、160和170 M e V质子(临床用于深部肿瘤的典型能量)的特定错位组合,以评估不同能量下的错位对微束的影响。结果:旋转错位引起了大量的不对称,在浅深度处峰值剂量降低高达68.4%,深度越大,不对称程度越低。平移位移导致微光束轮廓在相同方向和量级上的平移,而不改变微光束的形状。尽管存在这些畸变,模拟的微束阵列在中心80%场宽度的13 cm深度处获得了可接受的剂量均匀性,平坦度低于3%阈值。所提出的测量装置能够在一分钟内以亚毫米分辨率获取单个微光束的剖面,并且与变色薄膜的一致性很强(3%/ 0.2 mm伽马通过率为100%)。结论:准直器错位严重影响微束剂量稳健性,在临床治疗计划和质量保证(QA)中应予以考虑。所提出的测量装置可以使用市售检测器进行快速,高分辨率的剖面测量,并可以促进常规临床质量保证。
期刊介绍:
Medical Physics publishes original, high impact physics, imaging science, and engineering research that advances patient diagnosis and therapy through contributions in 1) Basic science developments with high potential for clinical translation 2) Clinical applications of cutting edge engineering and physics innovations 3) Broadly applicable and innovative clinical physics developments
Medical Physics is a journal of global scope and reach. By publishing in Medical Physics your research will reach an international, multidisciplinary audience including practicing medical physicists as well as physics- and engineering based translational scientists. We work closely with authors of promising articles to improve their quality.