Use of the internal astigmatism axis to determine the toric intraocular lens axis: Comparison between C-loop and plate-haptic lenses.

IF 1 4区 医学 Q3 OPHTHALMOLOGY
Arquivos brasileiros de oftalmologia Pub Date : 2026-08-28 eCollection Date: 2026-01-01 DOI:10.5935/0004-2749.2025-0395
Alper Can Yilmaz, Onder Ayyildiz
{"title":"Use of the internal astigmatism axis to determine the toric intraocular lens axis: Comparison between C-loop and plate-haptic lenses.","authors":"Alper Can Yilmaz, Onder Ayyildiz","doi":"10.5935/0004-2749.2025-0395","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate the agreement between the internal astigmatism axis obtained using the optical path difference-Scan III (NIDEK Co., Japan) and conventional slit-lamp biomicroscopic measurements for determining the toric intraocular lens axis in non-dilated eyes.</p><p><strong>Methods: </strong>This retrospective study included 56 eyes that underwent toric intraocular lens implantation and were evaluated at postoperative week 6. The toric intraocular lens axis was measured using the optical path difference-derived internal astigmatism axis under mesopic conditions without pharmacologic pupil dilation and compared with slit-lamp biomicroscopic measurements obtained after pharmacologic dilation. Agreement between the two methods was assessed using Bland-Altman analysis, the intraclass correlation coefficient, angular error analysis, and the distribution of clinically relevant deviations. Subgroup analyses were performed for plate-haptic and C-loop haptic intraocular lens.</p><p><strong>Results: </strong>The mean signed angular difference between the two methods was 0.00° (95% CI:, -0.41° to +0.41°), with 95% limits of agreement ranging from -2.99° to +2.99°. No systematic or proportional bias was observed. The mean absolute angular error was 1.21°±0.91°, and no eye demonstrated an intermethod discrepancy exceeding 5°. The difference between the two methods was ≤1° in 73.2% of eyes, ≤3° in 98.2%, and ≤5° in all eyes. Excellent agreement was observed between the two techniques (intraclass correlation coefficient >0.99, p<0.001). Comparable agreement was also observed in both the plate-haptic and C-loop haptic intraocular lens subgroups.</p><p><strong>Conclusions: </strong>Optical path difference-derived internal astigmatism axis measurements demonstrated excellent agreement with conventional slit-lamp biomicroscopic assessment for postoperative evaluation of toric intraocular lens alignment. The low angular error, narrow limits of agreement, and absence of clinically meaningful discrepancies suggest that this technique may serve as a reliable alternative for determining the toric intraocular lens axis in non-dilated eyes with adequate pupil size.</p>","PeriodicalId":8397,"journal":{"name":"Arquivos brasileiros de oftalmologia","volume":"89 5","pages":"e20250395"},"PeriodicalIF":1.0000,"publicationDate":"2026-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arquivos brasileiros de oftalmologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5935/0004-2749.2025-0395","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aimed to evaluate the agreement between the internal astigmatism axis obtained using the optical path difference-Scan III (NIDEK Co., Japan) and conventional slit-lamp biomicroscopic measurements for determining the toric intraocular lens axis in non-dilated eyes.

Methods: This retrospective study included 56 eyes that underwent toric intraocular lens implantation and were evaluated at postoperative week 6. The toric intraocular lens axis was measured using the optical path difference-derived internal astigmatism axis under mesopic conditions without pharmacologic pupil dilation and compared with slit-lamp biomicroscopic measurements obtained after pharmacologic dilation. Agreement between the two methods was assessed using Bland-Altman analysis, the intraclass correlation coefficient, angular error analysis, and the distribution of clinically relevant deviations. Subgroup analyses were performed for plate-haptic and C-loop haptic intraocular lens.

Results: The mean signed angular difference between the two methods was 0.00° (95% CI:, -0.41° to +0.41°), with 95% limits of agreement ranging from -2.99° to +2.99°. No systematic or proportional bias was observed. The mean absolute angular error was 1.21°±0.91°, and no eye demonstrated an intermethod discrepancy exceeding 5°. The difference between the two methods was ≤1° in 73.2% of eyes, ≤3° in 98.2%, and ≤5° in all eyes. Excellent agreement was observed between the two techniques (intraclass correlation coefficient >0.99, p<0.001). Comparable agreement was also observed in both the plate-haptic and C-loop haptic intraocular lens subgroups.

Conclusions: Optical path difference-derived internal astigmatism axis measurements demonstrated excellent agreement with conventional slit-lamp biomicroscopic assessment for postoperative evaluation of toric intraocular lens alignment. The low angular error, narrow limits of agreement, and absence of clinically meaningful discrepancies suggest that this technique may serve as a reliable alternative for determining the toric intraocular lens axis in non-dilated eyes with adequate pupil size.

利用内部散光轴确定环形人工晶状体轴:c -环与板触觉晶状体的比较。
目的:本研究旨在评估使用光学程差扫描III(日本NIDEK公司)获得的内散光轴与传统的裂隙灯生物显微测量方法确定非扩张眼的环形人工晶状体轴之间的一致性。方法:回顾性研究56只眼接受环形人工晶状体植入术,并于术后第6周进行评估。在无药物瞳孔扩张的中观条件下,采用光程差衍生的内散光轴测量环形人工晶状体轴,并与药物瞳孔扩张后的裂隙灯生物显微镜测量结果进行比较。采用Bland-Altman分析、类内相关系数、角度误差分析和临床相关偏差分布来评估两种方法之间的一致性。对平板触觉和c环触觉人工晶状体进行亚组分析。结果:两种方法的平均有符号角差为0.00°(95% CI: -0.41°至+0.41°),95%的一致性范围为-2.99°至+2.99°。未观察到系统性或比例偏倚。平均绝对角度误差为1.21°±0.91°,没有眼睛显示出超过5°的方法间差异。两种方法的差异为73.2%的眼≤1°,98.2%的眼≤3°,所有眼≤5°。两种技术之间的一致性非常好(类内相关系数>0.99,p)。结论:光程差衍生的内散光轴测量结果与传统的裂隙灯生物显微评估方法在环形人工晶状体排列术后评价中的一致性非常好。低的角度误差,狭窄的一致性范围,以及缺乏临床意义上的差异表明,该技术可以作为一种可靠的替代方法,用于确定瞳孔足够大的非扩张眼睛的环形人工晶状体轴。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.60
自引率
0.00%
发文量
200
审稿时长
6-12 weeks
期刊介绍: The ABO-ARQUIVOS BRASILEIROS DE OFTALMOLOGIA (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - electronic version), the official bimonthly publication of the Brazilian Council of Ophthalmology (CBO), aims to disseminate scientific studies in Ophthalmology, Visual Science and Health public, by promoting research, improvement and updating of professionals related to the field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书