Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory.

IF 1
Jagdish Chandarana, Kinjal Butani
{"title":"Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory.","authors":"Jagdish Chandarana, Kinjal Butani","doi":"10.1093/labmed/lmag052","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Hemoglobin A1c has a stringent total allowable error (TEa), making high Six Sigma values difficult to attain. In addition, data from ambulatory accredited laboratories are sparse.</p><p><strong>Methods: </strong>Internal quality control (QC) data for 2 levels were collected over 8 months on a single-lot Bio-Rad D-10 high-performance liquid chromatography analyzer. Long-term imprecision was the coefficient of variation (s/x¯) × 100 (CV) of all internal QC results combined, capturing between-month variation. Bias was derived from external quality assessment (EQAS) and assigned QC targets, applied as absolute values. The Sigma metric was (TEa ‒ absolute bias) / CV at TEa levels of 6%, 7%, and 8%; the Quality Goal Index distinguished imprecision-driven from bias-driven performance.</p><p><strong>Results: </strong>Long-term CVs were 3.04% (level 1) and 1.34% (level 2); EQASs were satisfactory throughout. At a TEa of 6% with EQAS-derived bias, Sigma metrics were 1.19 and 2.70 and, with a level-specific QC target bias, 1.46 and 4.22. At the Clinical Laboratory Improvement Amendments 2025 criterion of 8%, the Sigma metrics were 2.11 and 5.71. The level 1 Quality Goal Index (0.52) identified imprecision as predominant.</p><p><strong>Discussion: </strong>Despite satisfactory external proficiency, Sigma metrics fell below 3 at both levels against a 6% TEa and below 2 at the clinically pivotal lower concentration, driven by imprecision rather than bias. No control strategy compensates for this; therefore, we proposed a precision-focused corrective plan that targets between-month instability.</p>","PeriodicalId":94124,"journal":{"name":"Laboratory medicine","volume":"57 5","pages":""},"PeriodicalIF":1.0000,"publicationDate":"2026-08-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/labmed/lmag052","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Hemoglobin A1c has a stringent total allowable error (TEa), making high Six Sigma values difficult to attain. In addition, data from ambulatory accredited laboratories are sparse.

Methods: Internal quality control (QC) data for 2 levels were collected over 8 months on a single-lot Bio-Rad D-10 high-performance liquid chromatography analyzer. Long-term imprecision was the coefficient of variation (s/x¯) × 100 (CV) of all internal QC results combined, capturing between-month variation. Bias was derived from external quality assessment (EQAS) and assigned QC targets, applied as absolute values. The Sigma metric was (TEa ‒ absolute bias) / CV at TEa levels of 6%, 7%, and 8%; the Quality Goal Index distinguished imprecision-driven from bias-driven performance.

Results: Long-term CVs were 3.04% (level 1) and 1.34% (level 2); EQASs were satisfactory throughout. At a TEa of 6% with EQAS-derived bias, Sigma metrics were 1.19 and 2.70 and, with a level-specific QC target bias, 1.46 and 4.22. At the Clinical Laboratory Improvement Amendments 2025 criterion of 8%, the Sigma metrics were 2.11 and 5.71. The level 1 Quality Goal Index (0.52) identified imprecision as predominant.

Discussion: Despite satisfactory external proficiency, Sigma metrics fell below 3 at both levels against a 6% TEa and below 2 at the clinically pivotal lower concentration, driven by imprecision rather than bias. No control strategy compensates for this; therefore, we proposed a precision-focused corrective plan that targets between-month instability.

西格玛度量和质量目标指数评估血红蛋白A1c分析性能在一个流动认可的实验室。
简介:血红蛋白A1c有严格的总允许误差(TEa),使得高六西格玛值难以达到。此外,来自流动认可实验室的数据很少。方法:在单批Bio-Rad D-10高效液相色谱分析仪上采集8个月内2个级别的内部质量控制数据。长期不精度为所有内部QC结果的变异系数(s/x¯)× 100 (CV),捕获月间的变化。偏差来源于外部质量评估(EQAS)和指定的QC目标,作为绝对值。在TEa水平为6%、7%和8%时,Sigma指标为(TEa -绝对偏倚)/ CV;质量目标指数区分了不精确驱动和偏见驱动的表现。结果:长期cv分别为3.04%(水平1)和1.34%(水平2);EQASs在整个过程中都令人满意。在TEa为6%时,eqas衍生偏差,Sigma指标为1.19和2.70,在特定水平的QC目标偏差下,Sigma指标为1.46和4.22。在临床实验室改进修订2025标准为8%时,Sigma指标为2.11和5.71。1级质量目标指数(0.52)确定不精确是主要的。讨论:尽管外部熟练程度令人满意,但在6% TEa水平下,两个水平的Sigma指标都低于3,在临床关键低浓度下,Sigma指标低于2,这是由不精确而非偏倚驱动的。没有任何控制策略可以弥补这一点;因此,我们提出了针对月间不稳定性的精确校正计划。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书