Ventricular pulsed field ablation from bench to bedside: preclinical and clinical evidence from a systematic review and meta-analysis.

IF 10.2 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
Europace Pub Date : 2026-09-02 DOI:10.1093/europace/euag221
Paolo Compagnucci, Luca Finori, Michela Casella, Alessio Gasperetti, Henry D Huang, Yari Valeri, Leonardo D'Angelo, Francesca Campanelli, Giorgio Giacomini, Giovanni Volpato, Laura Cipolletta, Quintino Parisi, Luigi Di Biase, Andrea Natale, Pasquale Santangeli, Antonio Dello Russo
{"title":"Ventricular pulsed field ablation from bench to bedside: preclinical and clinical evidence from a systematic review and meta-analysis.","authors":"Paolo Compagnucci, Luca Finori, Michela Casella, Alessio Gasperetti, Henry D Huang, Yari Valeri, Leonardo D'Angelo, Francesca Campanelli, Giorgio Giacomini, Giovanni Volpato, Laura Cipolletta, Quintino Parisi, Luigi Di Biase, Andrea Natale, Pasquale Santangeli, Antonio Dello Russo","doi":"10.1093/europace/euag221","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>Evidence supporting the use of pulsed field ablation (PFA) for ventricular arrhythmia (VA) ablation remains limited and heterogeneous, and the relationship between preclinical lesion characteristics and clinical outcomes is incompletely defined. We sought to systematically summarize preclinical lesion metrics and clinical outcomes of ventricular PFA.</p><p><strong>Methods and results: </strong>We performed a systematic review and meta-analysis of preclinical and clinical studies of ventricular PFA. Preclinical studies reporting lesion depth or transmurality and clinical studies detailing procedural efficacy and safety were included. Random-effects models were used to pool lesion characteristics and clinical outcomes. Seventy-five studies were included (33 preclinical, 43 clinical; 355 patients). In preclinical models, mean lesion depth was 5.96 mm(95%CI, 5.07-6.84) in healthy endocardium and 5.19 mm (95%CI, 2.58-7.79) in healthy epicardium. Notably, lesion depth was quantitatively similar in models of scarred myocardium (5.96 mm; 95%CI, 5.39-6.53). Lesion transmurality was variable in healthy tissues (32%; 95%CI, 15-52%) but higher within scar (53%; 95%CI, 36-70%). Consistently, lesion depth was similar between PFA and RF in healthy myocardium, but greater with PFA in scarred models (within-study mean depth difference, 2.09 mm; 95%CI, 0.84-3.34). Greater contact force, repeated applications, and higher energy/duration settings were associated with increased lesion depth. Clinically, acute success was high for both premature ventricular complexes (PVCs,90%; 95%CI, 83-96%) and ventricular tachycardia (VT;92%; 95%CI, 87-96%) in studies with ≥3patients. Over a mean follow-up of approximately 5months, recurrence rates were 19% (95%CI, 12-28%) for PVCs and 21% (95%CI, 10-35%) for VT. The overall complication rate was 10% (95%CI, 5-17%) in studies with ≥3patients, with coronary artery spasm and conduction disturbances representing the most frequent adverse events.</p><p><strong>Conclusion: </strong>Ventricular PFA demonstrates high acute efficacy and encouraging mid-term outcomes, supported by preclinical evidence of favourable lesion penetration within scarred myocardium. Safety considerations-particularly coronary and conduction system effects-and the heterogeneity of preclinical evidence highlight the need for further investigation.</p>","PeriodicalId":11981,"journal":{"name":"Europace","volume":" ","pages":""},"PeriodicalIF":10.2000,"publicationDate":"2026-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13574506/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Europace","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/europace/euag221","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Aims: Evidence supporting the use of pulsed field ablation (PFA) for ventricular arrhythmia (VA) ablation remains limited and heterogeneous, and the relationship between preclinical lesion characteristics and clinical outcomes is incompletely defined. We sought to systematically summarize preclinical lesion metrics and clinical outcomes of ventricular PFA.

Methods and results: We performed a systematic review and meta-analysis of preclinical and clinical studies of ventricular PFA. Preclinical studies reporting lesion depth or transmurality and clinical studies detailing procedural efficacy and safety were included. Random-effects models were used to pool lesion characteristics and clinical outcomes. Seventy-five studies were included (33 preclinical, 43 clinical; 355 patients). In preclinical models, mean lesion depth was 5.96 mm(95%CI, 5.07-6.84) in healthy endocardium and 5.19 mm (95%CI, 2.58-7.79) in healthy epicardium. Notably, lesion depth was quantitatively similar in models of scarred myocardium (5.96 mm; 95%CI, 5.39-6.53). Lesion transmurality was variable in healthy tissues (32%; 95%CI, 15-52%) but higher within scar (53%; 95%CI, 36-70%). Consistently, lesion depth was similar between PFA and RF in healthy myocardium, but greater with PFA in scarred models (within-study mean depth difference, 2.09 mm; 95%CI, 0.84-3.34). Greater contact force, repeated applications, and higher energy/duration settings were associated with increased lesion depth. Clinically, acute success was high for both premature ventricular complexes (PVCs,90%; 95%CI, 83-96%) and ventricular tachycardia (VT;92%; 95%CI, 87-96%) in studies with ≥3patients. Over a mean follow-up of approximately 5months, recurrence rates were 19% (95%CI, 12-28%) for PVCs and 21% (95%CI, 10-35%) for VT. The overall complication rate was 10% (95%CI, 5-17%) in studies with ≥3patients, with coronary artery spasm and conduction disturbances representing the most frequent adverse events.

Conclusion: Ventricular PFA demonstrates high acute efficacy and encouraging mid-term outcomes, supported by preclinical evidence of favourable lesion penetration within scarred myocardium. Safety considerations-particularly coronary and conduction system effects-and the heterogeneity of preclinical evidence highlight the need for further investigation.

从实验室到床边的心室脉冲场消融:来自系统回顾和荟萃分析的临床前和临床证据。
背景和目的:支持使用脉冲场消融(PFA)治疗室性心律失常(VA)的证据仍然有限且不均匀,临床前病变特征与临床结果之间的关系尚不完全明确。我们试图系统地总结心室PFA的临床前病变指标和临床结果。方法:我们对心室PFA的临床前和临床研究进行了系统回顾和荟萃分析。包括报告病变深度或跨壁性的临床前研究和详细说明手术疗效和安全性的临床研究。随机效应模型用于汇总病变特征和临床结果。结果:纳入75项研究(33项临床前研究,43项临床研究,355例患者)。在临床前模型中,健康心内膜的平均病变深度为5.96mm(95%CI,5.07-6.84),健康心外膜的平均病变深度为5.19mm (95%CI,2.58-7.79)。值得注意的是,瘢痕心肌模型的病变深度在数量上相似(5.96mm; 95%CI,5.39-6.53)。病变跨壁性在健康组织中是不同的(32%;95%CI,15%-52%),但在疤痕组织中较高(53%;95%CI,36%-70%)。同样,PFA和RF在健康心肌中的病变深度相似,但PFA在疤痕模型中的病变深度更大(研究内平均深度差为2.09mm; 95%CI,0.84-3.34)。更大的接触力、重复应用和更高的能量/持续时间设置与损伤深度增加有关。临床上,在≥3例患者的研究中,室性早搏(早搏,90%;95%CI,83%-96%)和室性心动过速(VT, 92%; 95%CI,87%-96%)的急性成功率都很高。在平均约5个月的随访中,室性心动过速的复发率为19% (95%CI,12%-28%),室性心动过速的复发率为21% (95%CI,10%-35%)。在≥3例患者的研究中,总并发症发生率为10% (95%CI,5%-17%),冠状动脉痉挛和传导障碍是最常见的不良事件。结论:心室PFA具有较高的急性疗效和令人鼓舞的中期预后,临床前证据支持病变在瘢痕心肌内的良好渗透。安全性方面的考虑——尤其是冠状动脉和传导系统的影响——以及临床前证据的异质性强调了进一步研究的必要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Europace
Europace 医学-心血管系统
CiteScore
10.30
自引率
8.20%
发文量
851
审稿时长
3-6 weeks
期刊介绍: EP - Europace - European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology. The journal aims to provide an avenue of communication of top quality European and international original scientific work and reviews in the fields of Arrhythmias, Pacing and Cellular Electrophysiology. The Journal offers the reader a collection of contemporary original peer-reviewed papers, invited papers and editorial comments together with book reviews and correspondence.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书