Proton Bragg peak flash stereotactic radiosurgery for recurrent glioblastoma reirradiation

IF 3.2 Q2 ONCOLOGY
Yangguang Ma, Yuyu Wang, Tengda Zhang, Xingyi Zhao, Balaji Selvaraj, Chingyun Cheng, Michael Pennock, Jocelyn Jackson, Wei Guo, Xuanqin Mou, Benjamin Durkee, Jehee Isabelle Choi, Haibo Lin, Charles B. Simone II, Minglei Kang, Hui Wu
{"title":"Proton Bragg peak flash stereotactic radiosurgery for recurrent glioblastoma reirradiation","authors":"Yangguang Ma,&nbsp;Yuyu Wang,&nbsp;Tengda Zhang,&nbsp;Xingyi Zhao,&nbsp;Balaji Selvaraj,&nbsp;Chingyun Cheng,&nbsp;Michael Pennock,&nbsp;Jocelyn Jackson,&nbsp;Wei Guo,&nbsp;Xuanqin Mou,&nbsp;Benjamin Durkee,&nbsp;Jehee Isabelle Choi,&nbsp;Haibo Lin,&nbsp;Charles B. Simone II,&nbsp;Minglei Kang,&nbsp;Hui Wu","doi":"10.1016/j.phro.2026.101044","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and Purpose</h3><div>Stereotactic Radiosurgery with robotic image-guided systems is promising for recurrent glioblastoma (rGBM), but long delivery times may increase morbidity and reduce accuracy. Proton single-energy Bragg-peak FLASH-RT (SEBP-FLASH) may address these limits by delivering each field in &lt;1 s, with potential for normal-tissue sparing through the FLASH effect.</div></div><div><h3>Materials and Methods</h3><div>The SEBP-FLASH technique was used to reoptimize treatment for rGBM patients treated with robotic image-guided systems. Patient-specific aperture devices were incorporated into the SEBP-FLASH technique for sharper penumbra. The gross tumor volume (GTV) coverage, conformity index (CI), dose gradient index (GI<sub>50</sub> and GI<sub>30</sub>), and dose-volume histogram (DVHs), were evaluated. Additionally, ultra-high dose-rate ratios in critical organs-at-risk were assessed.</div></div><div><h3>Results</h3><div>SEBP-FLASH successfully met all clinical objectives. Both SEBP-FLASH and robotic techniques showed comparable performance, with GI<sub>50</sub> and GI<sub>30</sub> values of 2.9 ± 0.5 and 6.2 ± 1.8 for SEBP-FLASH, versus 2.7 ± 0.2 and 5.5 ± 0.5 for robotic (<em>p</em>-values at 0.085 and 0.189). SEBP-FLASH demonstrated superior conformality, achieving a CI of 1.4 ± 0.2 compared to CK's 1.9 ± 0.3 (p-value 0.001). All other metrics were similar across both techniques. For the brain, the FLASH dose rate coverage (V<sub>40 Gy/s</sub>) exceeded 80% under a 5 Gy dose threshold and minimum monitor unit of 400.</div></div><div><h3>Conclusions</h3><div>This study evaluated the dose-volume characteristics of SEBP-FLASH and robotic image-guided systems and found broadly comparable dose distributions, suggesting that SEBP-FLASH may be a feasible option for rGBM reirradiation. SEBP-FLASH also showed potential advantages in conformity, normal tissue sparing, and reduced on-table time.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"40 ","pages":"Article 101044"},"PeriodicalIF":3.2000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631626001442","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/20 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and Purpose

Stereotactic Radiosurgery with robotic image-guided systems is promising for recurrent glioblastoma (rGBM), but long delivery times may increase morbidity and reduce accuracy. Proton single-energy Bragg-peak FLASH-RT (SEBP-FLASH) may address these limits by delivering each field in <1 s, with potential for normal-tissue sparing through the FLASH effect.

Materials and Methods

The SEBP-FLASH technique was used to reoptimize treatment for rGBM patients treated with robotic image-guided systems. Patient-specific aperture devices were incorporated into the SEBP-FLASH technique for sharper penumbra. The gross tumor volume (GTV) coverage, conformity index (CI), dose gradient index (GI50 and GI30), and dose-volume histogram (DVHs), were evaluated. Additionally, ultra-high dose-rate ratios in critical organs-at-risk were assessed.

Results

SEBP-FLASH successfully met all clinical objectives. Both SEBP-FLASH and robotic techniques showed comparable performance, with GI50 and GI30 values of 2.9 ± 0.5 and 6.2 ± 1.8 for SEBP-FLASH, versus 2.7 ± 0.2 and 5.5 ± 0.5 for robotic (p-values at 0.085 and 0.189). SEBP-FLASH demonstrated superior conformality, achieving a CI of 1.4 ± 0.2 compared to CK's 1.9 ± 0.3 (p-value 0.001). All other metrics were similar across both techniques. For the brain, the FLASH dose rate coverage (V40 Gy/s) exceeded 80% under a 5 Gy dose threshold and minimum monitor unit of 400.

Conclusions

This study evaluated the dose-volume characteristics of SEBP-FLASH and robotic image-guided systems and found broadly comparable dose distributions, suggesting that SEBP-FLASH may be a feasible option for rGBM reirradiation. SEBP-FLASH also showed potential advantages in conformity, normal tissue sparing, and reduced on-table time.

Abstract Image

质子Bragg峰值闪光立体定向放射治疗复发性胶质母细胞瘤再照射
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书