{"title":"[33 years of pacemaker longevity-an unusual case].","authors":"Martin Koval, Timur Kyashif, Torsten Doenst","doi":"10.1007/s00399-026-01164-3","DOIUrl":null,"url":null,"abstract":"<p><p>We report on an 86-year-old patient whose unipolar single-chamber pacemaker (Vitatron Ceryx 623), implanted in 1991 due to sinus node arrest, had been continuously active for 33 years without any replacement of the pacemaker unit. No routine outpatient check-up had been carried out for over a decade. The ECG revealed asynchronous ventricular pacing (VOO) at a fixed rate of 50/min, consistent with the manufacturer's defined EOS fallback mode. Despite the failure to switch modes, the patient remained clinically event-free. Upon replacement of the device, complete shell-like calcification of the pocket (\"porcelain pocket\") was observed; the lead measurements (stimulation threshold 0.5 V/0.4 ms, sensing 7.2 mV, impedance 351 Ω) were excellent, thus the lead was left in place. This case illustrates the remarkable long-term performance of older devices and leads, at the same time it underlines the need for a regular follow-up and the risks of asynchronous pacing in EOS mode.</p>","PeriodicalId":52403,"journal":{"name":"Herzschrittmachertherapie und Elektrophysiologie","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2026-08-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Herzschrittmachertherapie und Elektrophysiologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s00399-026-01164-3","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
We report on an 86-year-old patient whose unipolar single-chamber pacemaker (Vitatron Ceryx 623), implanted in 1991 due to sinus node arrest, had been continuously active for 33 years without any replacement of the pacemaker unit. No routine outpatient check-up had been carried out for over a decade. The ECG revealed asynchronous ventricular pacing (VOO) at a fixed rate of 50/min, consistent with the manufacturer's defined EOS fallback mode. Despite the failure to switch modes, the patient remained clinically event-free. Upon replacement of the device, complete shell-like calcification of the pocket ("porcelain pocket") was observed; the lead measurements (stimulation threshold 0.5 V/0.4 ms, sensing 7.2 mV, impedance 351 Ω) were excellent, thus the lead was left in place. This case illustrates the remarkable long-term performance of older devices and leads, at the same time it underlines the need for a regular follow-up and the risks of asynchronous pacing in EOS mode.
期刊介绍:
Mit wissenschaftlichen Original- und Übersichtsarbeiten, Berichten über moderne Operationstechniken und experimentelle Methoden ist die Zeitschrift Herzschrittmachertherapie + Elektrophysiologie ein Diskussionsforum für Themen wie:
- Zelluläre Elektrophysiologie
- Theoretische Elektrophysiologie
- Klinische Elektrophysiologie
- Angewandte Herzschrittmachertherapie
- Bradykarde und tachykarde Herzrhythmusstörungen
- Plötzlicher Herztod und Risikostratifikation
- Elektrokardiographie
- Elektromedizinische Technologie
- Experimentelle und klinische Pharmakologie
- Herzchirurgie bei Herzrhythmusstörungen
Mitteilungen der Arbeitsgruppen Herzschrittmacher und Arrhythmie der Deutschen Gesellschaft für Kardiologie - Herz und Kreislaufforschung e.V. (DGK) sowie Stellungnahmen und praktische Hinweise runden das breite Spektrum dieser Zeitschrift ab.
Interessensgebiete: Kardiologie, Herzschrittmachertherapie, Herzschrittmachertechnologie, klinische Elektrophysiologie