{"title":"A Siemens Single-Shot T1 Mapping Sequence as an Alternative to MOLLI.","authors":"Mariem Dali, Narjes Benameur, Younes Arous, Salam Labidi","doi":"10.62438/tunismed.v104i05.6107","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Native T1 mapping has become an essential tool for quantitative myocardial tissue characterization in cardiac magnetic resonance (CMR). However, widely used techniques such as Modified Look-Locker Inversion Recovery (MOLLI) remain limited by their proprietary nature and limited accessibility, particularly in resource-constrained settings.</p><p><strong>Aim: </strong>This study aimed to develop and validate a more accessible Single-Shot T1 mapping technique as a practical alternative to MOLLI, and to assess its agreement and diagnostic performance against a reference standard (Circle Cardiovascular Imaging, CVI⁴²).</p><p><strong>Methods: </strong>A retrospective study was conducted on 40 subjects (25 healthy controls and 15 patients with cardiac pathologies) who underwent 3T CMR imaging. Native T1 values were acquired using a modified Single-Shot inversion recovery scheme (3(3)3(3)5). Quantitative analysis was performed using custom MATLAB-based post-processing and compared with reference measurements obtained using CVI⁴² software. Agreement was assessed using Bland-Altman analysis, while group differences were evaluated using independent t-tests.</p><p><strong>Results: </strong>Pathological subjects demonstrated significantly elevated native T1 values compared to healthy controls (1566 ± 65 ms vs. 1079 ± 80 ms, p < 0.001). The proposed method showed strong agreement with the reference standard, with a mean difference of 5.7 ms and no statistically significant bias. Additionally, the Single-Shot approach exhibited significantly reduced inter-subject variability compared to CVI⁴²-based measurements.</p><p><strong>Conclusion: </strong>The proposed Single-Shot T1 mapping technique provides a reliable, reproducible, and accessible alternative to MOLLI, with comparable accuracy and improved numerical stability. This approach may expand access to advanced myocardial tissue characterization, particularly in low-resource settings.</p>","PeriodicalId":38818,"journal":{"name":"Tunisie Medicale","volume":"104 4","pages":"651-658"},"PeriodicalIF":0.0000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Tunisie Medicale","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.62438/tunismed.v104i05.6107","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Native T1 mapping has become an essential tool for quantitative myocardial tissue characterization in cardiac magnetic resonance (CMR). However, widely used techniques such as Modified Look-Locker Inversion Recovery (MOLLI) remain limited by their proprietary nature and limited accessibility, particularly in resource-constrained settings.
Aim: This study aimed to develop and validate a more accessible Single-Shot T1 mapping technique as a practical alternative to MOLLI, and to assess its agreement and diagnostic performance against a reference standard (Circle Cardiovascular Imaging, CVI⁴²).
Methods: A retrospective study was conducted on 40 subjects (25 healthy controls and 15 patients with cardiac pathologies) who underwent 3T CMR imaging. Native T1 values were acquired using a modified Single-Shot inversion recovery scheme (3(3)3(3)5). Quantitative analysis was performed using custom MATLAB-based post-processing and compared with reference measurements obtained using CVI⁴² software. Agreement was assessed using Bland-Altman analysis, while group differences were evaluated using independent t-tests.
Results: Pathological subjects demonstrated significantly elevated native T1 values compared to healthy controls (1566 ± 65 ms vs. 1079 ± 80 ms, p < 0.001). The proposed method showed strong agreement with the reference standard, with a mean difference of 5.7 ms and no statistically significant bias. Additionally, the Single-Shot approach exhibited significantly reduced inter-subject variability compared to CVI⁴²-based measurements.
Conclusion: The proposed Single-Shot T1 mapping technique provides a reliable, reproducible, and accessible alternative to MOLLI, with comparable accuracy and improved numerical stability. This approach may expand access to advanced myocardial tissue characterization, particularly in low-resource settings.
背景:原生T1定位已成为心脏磁共振(CMR)定量心肌组织表征的重要工具。然而,广泛使用的技术,如改进的Look-Locker反演恢复(MOLLI),仍然受到其专有性质和有限的可及性的限制,特别是在资源受限的环境中。目的:本研究旨在开发和验证一种更容易获得的单次T1映射技术,作为MOLLI的实用替代方案,并根据参考标准(Circle Cardiovascular Imaging, CVI⁴²)评估其一致性和诊断性能。方法:对40例接受3T CMR成像的患者(健康对照25例,心脏病变患者15例)进行回顾性研究。原生T1值使用改进的Single-Shot反演恢复方案获得(3(3)3(3)5)。使用定制的基于matlab的后处理进行定量分析,并与使用CVI⁴软件获得的参考测量值进行比较。使用Bland-Altman分析评估一致性,使用独立t检验评估组间差异。结果:与健康对照组相比,病理组T1值明显升高(1566±65 ms vs 1079±80 ms, p < 0.001)。该方法与参考标准具有较强的一致性,平均差值为5.7 ms,无统计学显著偏差。此外,与基于CVI⁴²的测量相比,单次射击方法显着降低了受试者间的可变性。结论:提出的单次T1定位技术提供了一种可靠的、可重复的、可获得的替代MOLLI的方法,具有相当的精度和更高的数值稳定性。这种方法可以扩大对高级心肌组织特征的获取,特别是在资源匮乏的环境中。