The impact of data extraction percentage and deep learning-based reconstruction on image quality in gated PET/computed tomography.

IF 1.4 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Hiroki Nosaka, Masaya Suda, Noriaki Miyaji, Hiraku Fuse, Kenji Yasue, Norikazu Koori, Shin Miyakawa, Masato Takahashi, Koichi Hanada, Shogo Imai
{"title":"The impact of data extraction percentage and deep learning-based reconstruction on image quality in gated PET/computed tomography.","authors":"Hiroki Nosaka, Masaya Suda, Noriaki Miyaji, Hiraku Fuse, Kenji Yasue, Norikazu Koori, Shin Miyakawa, Masato Takahashi, Koichi Hanada, Shogo Imai","doi":"10.1097/MNM.0000000000002222","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Respiratory motion artifacts degrade PET/computed tomography (PET/CT) image quality. Data-driven gated (DDG) PET/CT addresses this issue by extracting respiratory signals directly from PET data, eliminating the need for external monitoring devices. This study investigated the effects of data extraction percentage (%count) and deep learning-based reconstruction [Advanced Intelligent Clear-IQ Engine-integrated (AiCE-i)] on image quality in DDG-PET under different respiratory conditions using a phantom model.</p><p><strong>Methods: </strong>A body phantom containing six spheres (10-37 mm) was imaged using a silicon photomultiplier-based PET/CT system. Four respiratory waveforms (no-motion, sinusoidal, representative patient, and baseline shift) and four %count levels (20, 30, 40, and 50%) were evaluated using AiCE-i reconstruction. Image quality was assessed using background variability (N10 mm), percentage contrast (QH,10 mm), contrast-to-noise ratio (QH,10 mm/N10 mm), and recovery coefficient.</p><p><strong>Results: </strong>Increasing %count consistently reduced N10 mm across all respiratory waveforms. QH,10 mm and QH,10 mm/N10 mm generally increased with increasing %count; however, at the 50% threshold, significant reductions were observed in the baseline shift and sinusoidal waveforms compared with the no-motion condition. Recovery coefficient analysis demonstrated the partial volume effect in smaller spheres and showed that quantitative performance was maintained across the evaluated gating conditions.</p><p><strong>Conclusion: </strong>The combination of DDG and AiCE-i maintained stable image quality across the respiratory conditions evaluated. Under the conditions of this phantom study, a %count range of 30-40% provided a favorable balance between image noise and the effects of respiratory motion for standard 120-s acquisitions.</p>","PeriodicalId":19708,"journal":{"name":"Nuclear Medicine Communications","volume":" ","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2026-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Medicine Communications","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MNM.0000000000002222","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Respiratory motion artifacts degrade PET/computed tomography (PET/CT) image quality. Data-driven gated (DDG) PET/CT addresses this issue by extracting respiratory signals directly from PET data, eliminating the need for external monitoring devices. This study investigated the effects of data extraction percentage (%count) and deep learning-based reconstruction [Advanced Intelligent Clear-IQ Engine-integrated (AiCE-i)] on image quality in DDG-PET under different respiratory conditions using a phantom model.

Methods: A body phantom containing six spheres (10-37 mm) was imaged using a silicon photomultiplier-based PET/CT system. Four respiratory waveforms (no-motion, sinusoidal, representative patient, and baseline shift) and four %count levels (20, 30, 40, and 50%) were evaluated using AiCE-i reconstruction. Image quality was assessed using background variability (N10 mm), percentage contrast (QH,10 mm), contrast-to-noise ratio (QH,10 mm/N10 mm), and recovery coefficient.

Results: Increasing %count consistently reduced N10 mm across all respiratory waveforms. QH,10 mm and QH,10 mm/N10 mm generally increased with increasing %count; however, at the 50% threshold, significant reductions were observed in the baseline shift and sinusoidal waveforms compared with the no-motion condition. Recovery coefficient analysis demonstrated the partial volume effect in smaller spheres and showed that quantitative performance was maintained across the evaluated gating conditions.

Conclusion: The combination of DDG and AiCE-i maintained stable image quality across the respiratory conditions evaluated. Under the conditions of this phantom study, a %count range of 30-40% provided a favorable balance between image noise and the effects of respiratory motion for standard 120-s acquisitions.

数据提取百分比和基于深度学习的重建对门控PET/计算机断层扫描图像质量的影响。
背景:呼吸运动伪影降低PET/计算机断层扫描(PET/CT)图像质量。数据驱动门控(DDG) PET/CT通过直接从PET数据中提取呼吸信号来解决这一问题,从而消除了对外部监测设备的需求。本研究使用幻影模型研究了数据提取百分比(%count)和基于深度学习的重建[集成高级智能Clear-IQ引擎(aisi)]对不同呼吸条件下DDG-PET图像质量的影响。方法:采用基于硅光电倍增管的PET/CT系统对含有6个10-37 mm球体的体影进行成像。使用AiCE-i重建评估四种呼吸波形(无运动、正弦、代表性患者和基线移位)和四种%计数水平(20%、30%、40%和50%)。使用背景变异性(N10 mm)、对比度百分比(QH,10 mm)、对比度噪声比(QH,10 mm/N10 mm)和恢复系数评估图像质量。结果:在所有呼吸波形中,增加%计数持续减少N10 mm。QH、10mm和QH、10mm / n10mm一般随%计数的增加而增加;然而,在50%阈值下,与无运动条件相比,基线移位和正弦波形明显减少。回收系数分析表明,在较小的球体中存在部分体积效应,并表明在评价的浇注条件下,定量性能保持不变。结论:DDG联合AiCE-i在评估呼吸状况时保持稳定的图像质量。在这个幻像研究的条件下,30-40%的%计数范围在图像噪声和呼吸运动的影响之间提供了一个有利的平衡,用于标准的120-s采集。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.20
自引率
6.70%
发文量
212
审稿时长
3-8 weeks
期刊介绍: Nuclear Medicine Communications, the official journal of the British Nuclear Medicine Society, is a rapid communications journal covering nuclear medicine and molecular imaging with radionuclides, and the basic supporting sciences. As well as clinical research and commentary, manuscripts describing research on preclinical and basic sciences (radiochemistry, radiopharmacy, radiobiology, radiopharmacology, medical physics, computing and engineering, and technical and nursing professions involved in delivering nuclear medicine services) are welcomed, as the journal is intended to be of interest internationally to all members of the many medical and non-medical disciplines involved in nuclear medicine. In addition to papers reporting original studies, frankly written editorials and topical reviews are a regular feature of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书