{"title":"Effect of 70-kVp chest CT on surface dose at the mammary gland level in a neonatal phantom under matched CTDIvol conditions.","authors":"Nanami Usui, Takanori Masuda, Norimi Nishiyama, Junichi Nakagawa, Mako Yamada, Nagito Kuwada, Hitoshi Takahata, Takashi Amano, Shinichi Arao","doi":"10.1007/s12194-026-01117-y","DOIUrl":null,"url":null,"abstract":"<p><p>To compare surface dose at the mammary gland level between 70- and 120-kVp non-contrast neonatal chest computed tomography (CT) under matched volume CT dose index (CTDIvol) conditions, a neonatal anthropomorphic phantom was scanned using a 256-row CT scanner. Real-time dosimeters were placed bilaterally on the anterior chest wall at the mammary gland level and on the posterior chest wall at the same axial level. For the 70-kVp protocol, the noise index was adjusted to match the console-displayed CTDIvol of the 120-kVp protocol. Each protocol was repeated 15 times, and image noise was measured on axial images. Anterior surface dose was significantly higher with 70 kVp than with 120 kVp (median [IQR], 5.67 [5.17-5.86] vs. 5.29 [4.84-5.43] mGy; p < 0.01). Posterior surface dose and image noise did not differ significantly. Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":"1460-1465"},"PeriodicalIF":1.6000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01117-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/7 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
To compare surface dose at the mammary gland level between 70- and 120-kVp non-contrast neonatal chest computed tomography (CT) under matched volume CT dose index (CTDIvol) conditions, a neonatal anthropomorphic phantom was scanned using a 256-row CT scanner. Real-time dosimeters were placed bilaterally on the anterior chest wall at the mammary gland level and on the posterior chest wall at the same axial level. For the 70-kVp protocol, the noise index was adjusted to match the console-displayed CTDIvol of the 120-kVp protocol. Each protocol was repeated 15 times, and image noise was measured on axial images. Anterior surface dose was significantly higher with 70 kVp than with 120 kVp (median [IQR], 5.67 [5.17-5.86] vs. 5.29 [4.84-5.43] mGy; p < 0.01). Posterior surface dose and image noise did not differ significantly. Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.
期刊介绍:
The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.