Effect of 70-kVp chest CT on surface dose at the mammary gland level in a neonatal phantom under matched CTDIvol conditions.

IF 1.6 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Radiological Physics and Technology Pub Date : 2026-09-01 Epub Date: 2026-08-07 DOI:10.1007/s12194-026-01117-y
Nanami Usui, Takanori Masuda, Norimi Nishiyama, Junichi Nakagawa, Mako Yamada, Nagito Kuwada, Hitoshi Takahata, Takashi Amano, Shinichi Arao
{"title":"Effect of 70-kVp chest CT on surface dose at the mammary gland level in a neonatal phantom under matched CTDIvol conditions.","authors":"Nanami Usui, Takanori Masuda, Norimi Nishiyama, Junichi Nakagawa, Mako Yamada, Nagito Kuwada, Hitoshi Takahata, Takashi Amano, Shinichi Arao","doi":"10.1007/s12194-026-01117-y","DOIUrl":null,"url":null,"abstract":"<p><p>To compare surface dose at the mammary gland level between 70- and 120-kVp non-contrast neonatal chest computed tomography (CT) under matched volume CT dose index (CTDIvol) conditions, a neonatal anthropomorphic phantom was scanned using a 256-row CT scanner. Real-time dosimeters were placed bilaterally on the anterior chest wall at the mammary gland level and on the posterior chest wall at the same axial level. For the 70-kVp protocol, the noise index was adjusted to match the console-displayed CTDIvol of the 120-kVp protocol. Each protocol was repeated 15 times, and image noise was measured on axial images. Anterior surface dose was significantly higher with 70 kVp than with 120 kVp (median [IQR], 5.67 [5.17-5.86] vs. 5.29 [4.84-5.43] mGy; p < 0.01). Posterior surface dose and image noise did not differ significantly. Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":"1460-1465"},"PeriodicalIF":1.6000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01117-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/8/7 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

To compare surface dose at the mammary gland level between 70- and 120-kVp non-contrast neonatal chest computed tomography (CT) under matched volume CT dose index (CTDIvol) conditions, a neonatal anthropomorphic phantom was scanned using a 256-row CT scanner. Real-time dosimeters were placed bilaterally on the anterior chest wall at the mammary gland level and on the posterior chest wall at the same axial level. For the 70-kVp protocol, the noise index was adjusted to match the console-displayed CTDIvol of the 120-kVp protocol. Each protocol was repeated 15 times, and image noise was measured on axial images. Anterior surface dose was significantly higher with 70 kVp than with 120 kVp (median [IQR], 5.67 [5.17-5.86] vs. 5.29 [4.84-5.43] mGy; p < 0.01). Posterior surface dose and image noise did not differ significantly. Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.

在匹配CTDIvol条件下,70 kvp胸部CT对新生儿乳腺水平表面剂量的影响。
为了比较匹配容积CT剂量指数(CTDIvol)条件下70 kvp和120 kvp的非对比新生儿胸部计算机断层扫描(CT)在乳腺水平的表面剂量,使用256排CT扫描仪扫描新生儿拟人化幻影。实时剂量仪放置于双侧乳腺水平的前胸壁和相同轴向水平的后胸壁。对于70-kVp协议,调整噪声指数以匹配控制台显示的120-kVp协议的CTDIvol。每个方案重复15次,并在轴向图像上测量图像噪声。70 kVp组前表面剂量显著高于120 kVp组(中位[IQR], 5.67[5.17-5.86]比5.29 [4.84-5.43]mGy
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书