Agreement and Accuracy of a Commercial Pulse Oximeter Compared with Arterial Oxygen Saturation During Controlled Hypoxemia in Healthy Volunteers.

IF 1.5 Q4 ENGINEERING, BIOMEDICAL
Medical Devices-Evidence and Research Pub Date : 2026-07-14 eCollection Date: 2026-01-01 DOI:10.2147/MDER.S626574
Jinqi Lin, Shiyan Pan, Lan Ma, Jilun Ye, Chunsheng Liu, Fan Wang, Xu Zhang
{"title":"Agreement and Accuracy of a Commercial Pulse Oximeter Compared with Arterial Oxygen Saturation During Controlled Hypoxemia in Healthy Volunteers.","authors":"Jinqi Lin, Shiyan Pan, Lan Ma, Jilun Ye, Chunsheng Liu, Fan Wang, Xu Zhang","doi":"10.2147/MDER.S626574","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Pulse oxygen saturation (SpO<sub>2</sub>) is widely used for continuous non-invasive oxygenation monitoring, but its accuracy can be affected by sensor design, probe placement, peripheral perfusion, skin pigmentation, motion artifacts, and device algorithms. This study evaluated the accuracy and agreement of a commercial pulse oximetry system (model M201 SpO<sub>2</sub> module, Shenzhen Witleaf Medical Electronics Co. Ltd.) with a matched fingertip sensor during non-motion controlled hypoxemia in healthy volunteers, using arterial oxygen saturation (SaO<sub>2</sub>) measured by arterial blood gas/co-oximetry analysis as the reference standard.</p><p><strong>Material and methods: </strong>Eighteen healthy volunteers underwent two phases of controlled hypoxemia. Each phase comprised three target SpO<sub>2</sub> plateau ranges (90-100%, 80-89%, and 70-79%). Phase 1 was the desaturation phase, and Phase 2 was a re-oxygenation and repetition phase. After each target plateau stabilized, SpO<sub>2</sub> values from the investigational system were recorded, and paired arterial blood samples were collected for SaO<sub>2</sub> measurement. Measurement error was defined as SpO<sub>2</sub>-SaO<sub>2</sub>. Bias, mean absolute error (MAE), accuracy root mean square (ARMS), 95% limits of agreement (LOA), and proportions of absolute errors exceeding predefined thresholds were calculated.</p><p><strong>Results: </strong>A total of 432 paired SpO<sub>2</sub>-SaO<sub>2</sub> observations were analyzed. SpO<sub>2</sub> showed strong correlation with SaO<sub>2</sub> (Pearson r = 0.990; Spearman r = 0.986; both p < 0.001). Overall bias was -0.061%, with 95% LOA from -2.65% to 2.53%. MAE and ARMS were 1.05% and 1.32%, respectively. Absolute errors exceeding ±2%, ±3%, and ±4% occurred in 11.57%, 2.55%, and 0% of observations, respectively. ARMS values were 1.03%, 1.19%, and 1.66% across the 90-100%, 80-89%, and 70-79% platform phases, respectively, with increased error dispersion and mild SaO<sub>2</sub> underestimation in the 70-79% phase.</p><p><strong>Conclusion: </strong>The commercial pulse oximetry system showed good overall agreement with reference SaO<sub>2</sub> during controlled hypoxemia, with an ARMS of 1.32% (well below the 4% threshold). Error dispersion increased in the 70-79% saturation range, with mild underestimation of SaO<sub>2</sub> by SpO<sub>2</sub>. These findings support the clinical performance of this device within the tested saturation range.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"19 ","pages":"626574"},"PeriodicalIF":1.5000,"publicationDate":"2026-07-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13381327/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Devices-Evidence and Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/MDER.S626574","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Pulse oxygen saturation (SpO2) is widely used for continuous non-invasive oxygenation monitoring, but its accuracy can be affected by sensor design, probe placement, peripheral perfusion, skin pigmentation, motion artifacts, and device algorithms. This study evaluated the accuracy and agreement of a commercial pulse oximetry system (model M201 SpO2 module, Shenzhen Witleaf Medical Electronics Co. Ltd.) with a matched fingertip sensor during non-motion controlled hypoxemia in healthy volunteers, using arterial oxygen saturation (SaO2) measured by arterial blood gas/co-oximetry analysis as the reference standard.

Material and methods: Eighteen healthy volunteers underwent two phases of controlled hypoxemia. Each phase comprised three target SpO2 plateau ranges (90-100%, 80-89%, and 70-79%). Phase 1 was the desaturation phase, and Phase 2 was a re-oxygenation and repetition phase. After each target plateau stabilized, SpO2 values from the investigational system were recorded, and paired arterial blood samples were collected for SaO2 measurement. Measurement error was defined as SpO2-SaO2. Bias, mean absolute error (MAE), accuracy root mean square (ARMS), 95% limits of agreement (LOA), and proportions of absolute errors exceeding predefined thresholds were calculated.

Results: A total of 432 paired SpO2-SaO2 observations were analyzed. SpO2 showed strong correlation with SaO2 (Pearson r = 0.990; Spearman r = 0.986; both p < 0.001). Overall bias was -0.061%, with 95% LOA from -2.65% to 2.53%. MAE and ARMS were 1.05% and 1.32%, respectively. Absolute errors exceeding ±2%, ±3%, and ±4% occurred in 11.57%, 2.55%, and 0% of observations, respectively. ARMS values were 1.03%, 1.19%, and 1.66% across the 90-100%, 80-89%, and 70-79% platform phases, respectively, with increased error dispersion and mild SaO2 underestimation in the 70-79% phase.

Conclusion: The commercial pulse oximetry system showed good overall agreement with reference SaO2 during controlled hypoxemia, with an ARMS of 1.32% (well below the 4% threshold). Error dispersion increased in the 70-79% saturation range, with mild underestimation of SaO2 by SpO2. These findings support the clinical performance of this device within the tested saturation range.

商业脉搏血氧仪与健康志愿者控制低氧血症期间动脉血氧饱和度的一致性和准确性。
目的:脉冲氧饱和度(SpO2)广泛用于连续无创氧合监测,但其准确性会受到传感器设计、探头放置、外周灌注、皮肤色素沉着、运动伪影和设备算法的影响。本研究以动脉血气/共氧法测定的动脉血氧饱和度(SaO2)为参考标准,评估了带有匹配指尖传感器的商用脉搏血氧仪(M201型SpO2模块,深圳威立夫医疗电子有限公司)在健康志愿者非运动控制低氧血症时的准确性和一致性。材料和方法:18名健康志愿者接受两个阶段的控制低氧血症。每个阶段包括三个目标SpO2平台范围(90-100%,80-89%和70-79%)。第1阶段是去饱和阶段,第2阶段是再氧化和重复阶段。在每个目标平台稳定后,记录研究系统的SpO2值,并收集成对的动脉血样本进行SaO2测量。测量误差定义为SpO2-SaO2。计算偏倚、平均绝对误差(MAE)、精度均方根(ARMS)、95%一致限(LOA)和绝对误差超过预定义阈值的比例。结果:共分析了432对SpO2-SaO2观测值。SpO2与SaO2有较强的相关性(Pearson r = 0.990; Spearman r = 0.986, p均< 0.001)。总体偏倚为-0.061%,95% LOA从-2.65%到2.53%。MAE和ARMS分别为1.05%和1.32%。绝对误差超过±2%、±3%和±4%的观测值分别为11.57%、2.55%和0%。在90-100%、80-89%和70-79%的平台阶段,ARMS值分别为1.03%、1.19%和1.66%,在70-79%的平台阶段,误差分散增加,SaO2轻度低估。结论:在控制低氧血症时,商用脉搏血氧仪与参考SaO2的总体一致性良好,ARMS为1.32%(远低于4%的阈值)。在70-79%饱和度范围内,误差色散增大,SpO2对SaO2的估计轻微低估。这些发现支持该设备在测试饱和范围内的临床性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Medical Devices-Evidence and Research
Medical Devices-Evidence and Research ENGINEERING, BIOMEDICAL-
CiteScore
2.80
自引率
0.00%
发文量
41
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书