Effect of CT Acquisition Parameters on RetinaNet-Based Lung Nodule Detection: A Physics-Guided Validation Study Across Dose, Slice Thickness, and Reconstruction Kernel
IF 4.7 2区 医学Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
{"title":"Effect of CT Acquisition Parameters on RetinaNet-Based Lung Nodule Detection: A Physics-Guided Validation Study Across Dose, Slice Thickness, and Reconstruction Kernel","authors":"Daniel Soliman MS","doi":"10.1016/j.acra.2026.06.056","DOIUrl":null,"url":null,"abstract":"<div><h3>Rationale and Objectives</h3><div>Computed tomography acquisition parameters, including dose level, slice thickness, and reconstruction kernel, vary substantially across clinical sites, yet artificial intelligence (AI)-assisted lung nodule detection systems are rarely evaluated against this variability. We propose and demonstrate a physics-guided framework for evaluating detection model sensitivity to systematic acquisition parameter variation.</div></div><div>Computed tomography acquisition parameters, including dose level, slice thickness, and reconstruction kernel, vary substantially across clinical sites, yet artificial intelligence (AI)-assisted lung nodule detection systems are rarely evaluated against this variability. We propose and demonstrate a physics-guided framework for evaluating detection model sensitivity to systematic acquisition parameter variation.</div><div><h3>Materials and Methods</h3><div>154 cases from the LIDC-IDRI dataset were evaluated using a Medical Open Network for AI (MONAI) RetinaNet model pretrained on LUNA16 (fold 0, no fine-tuning) across six imaging conditions: baseline, 25% dose reduction, 50% dose reduction, 3 mm slice thickness, 5 mm slice thickness, and soft kernel reconstruction. Dose reduction was simulated via image-domain Gaussian noise; slice thickness via z-axis moving average. Detection sensitivity was computed at confidence threshold 0.5 with a 15 mm matching criterion.</div></div><div>154 cases from the LIDC-IDRI dataset were evaluated using a Medical Open Network for AI (MONAI) RetinaNet model pretrained on LUNA16 (fold 0, no fine-tuning) across six imaging conditions: baseline, 25% dose reduction, 50% dose reduction, 3 mm slice thickness, 5 mm slice thickness, and soft kernel reconstruction. Dose reduction was simulated via image-domain Gaussian noise; slice thickness via z-axis moving average. Detection sensitivity was computed at confidence threshold 0.5 with a 15 mm matching criterion.</div><div><h3>Results</h3><div>Baseline sensitivity was 84.8% (95% CI: 80.2–89.8%). Soft kernel and 5 mm slice thickness produced the largest decreases: 74.3% (−10.5 pp, <em>p <</em> 0.05) and 71.6% (−13.2 pp, <em>p <</em> 0.05), respectively. Dose reduction caused moderate asymmetric degradation: 76.0% at 25% dose (−8.8 pp, <em>p <</em> 0.05) and 80.0% at 50% dose (−4.8 pp, <em>p <</em> 0.05). 3 mm slice thickness produced a small but significant reduction (−2.4 pp, <em>p</em> = 0.009). Reductions were most pronounced in the 3–6 mm range and consistent across confidence thresholds.</div></div><div>Baseline sensitivity was 84.8% (95% CI: 80.2–89.8%). Soft kernel and 5 mm slice thickness produced the largest decreases: 74.3% (−10.5 pp, <em>p <</em> 0.05) and 71.6% (−13.2 pp, <em>p <</em> 0.05), respectively. Dose reduction caused moderate asymmetric degradation: 76.0% at 25% dose (−8.8 pp, <em>p <</em> 0.05) and 80.0% at 50% dose (−4.8 pp, <em>p <</em> 0.05). 3 mm slice thickness produced a small but significant reduction (−2.4 pp, <em>p</em> = 0.009). Reductions were most pronounced in the 3–6 mm range and consistent across confidence thresholds.</div><div><h3>Conclusion</h3><div>Slice thickness and soft reconstruction kernel represent stronger constraints on AI detection performance than image noise under these conditions. The proposed framework is reproducible, requires no proprietary scanner data, and provides a practical basis for post-deployment acquisition QA.</div></div><div>Slice thickness and soft reconstruction kernel represent stronger constraints on AI detection performance than image noise under these conditions. The proposed framework is reproducible, requires no proprietary scanner data, and provides a practical basis for post-deployment acquisition QA.</div>","PeriodicalId":50928,"journal":{"name":"Academic Radiology","volume":"33 10","pages":"Pages 4625-4633"},"PeriodicalIF":4.7000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Academic Radiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S107663322600468X","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/23 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Rationale and Objectives
Computed tomography acquisition parameters, including dose level, slice thickness, and reconstruction kernel, vary substantially across clinical sites, yet artificial intelligence (AI)-assisted lung nodule detection systems are rarely evaluated against this variability. We propose and demonstrate a physics-guided framework for evaluating detection model sensitivity to systematic acquisition parameter variation.
Computed tomography acquisition parameters, including dose level, slice thickness, and reconstruction kernel, vary substantially across clinical sites, yet artificial intelligence (AI)-assisted lung nodule detection systems are rarely evaluated against this variability. We propose and demonstrate a physics-guided framework for evaluating detection model sensitivity to systematic acquisition parameter variation.
Materials and Methods
154 cases from the LIDC-IDRI dataset were evaluated using a Medical Open Network for AI (MONAI) RetinaNet model pretrained on LUNA16 (fold 0, no fine-tuning) across six imaging conditions: baseline, 25% dose reduction, 50% dose reduction, 3 mm slice thickness, 5 mm slice thickness, and soft kernel reconstruction. Dose reduction was simulated via image-domain Gaussian noise; slice thickness via z-axis moving average. Detection sensitivity was computed at confidence threshold 0.5 with a 15 mm matching criterion.
154 cases from the LIDC-IDRI dataset were evaluated using a Medical Open Network for AI (MONAI) RetinaNet model pretrained on LUNA16 (fold 0, no fine-tuning) across six imaging conditions: baseline, 25% dose reduction, 50% dose reduction, 3 mm slice thickness, 5 mm slice thickness, and soft kernel reconstruction. Dose reduction was simulated via image-domain Gaussian noise; slice thickness via z-axis moving average. Detection sensitivity was computed at confidence threshold 0.5 with a 15 mm matching criterion.
Results
Baseline sensitivity was 84.8% (95% CI: 80.2–89.8%). Soft kernel and 5 mm slice thickness produced the largest decreases: 74.3% (−10.5 pp, p < 0.05) and 71.6% (−13.2 pp, p < 0.05), respectively. Dose reduction caused moderate asymmetric degradation: 76.0% at 25% dose (−8.8 pp, p < 0.05) and 80.0% at 50% dose (−4.8 pp, p < 0.05). 3 mm slice thickness produced a small but significant reduction (−2.4 pp, p = 0.009). Reductions were most pronounced in the 3–6 mm range and consistent across confidence thresholds.
Baseline sensitivity was 84.8% (95% CI: 80.2–89.8%). Soft kernel and 5 mm slice thickness produced the largest decreases: 74.3% (−10.5 pp, p < 0.05) and 71.6% (−13.2 pp, p < 0.05), respectively. Dose reduction caused moderate asymmetric degradation: 76.0% at 25% dose (−8.8 pp, p < 0.05) and 80.0% at 50% dose (−4.8 pp, p < 0.05). 3 mm slice thickness produced a small but significant reduction (−2.4 pp, p = 0.009). Reductions were most pronounced in the 3–6 mm range and consistent across confidence thresholds.
Conclusion
Slice thickness and soft reconstruction kernel represent stronger constraints on AI detection performance than image noise under these conditions. The proposed framework is reproducible, requires no proprietary scanner data, and provides a practical basis for post-deployment acquisition QA.
Slice thickness and soft reconstruction kernel represent stronger constraints on AI detection performance than image noise under these conditions. The proposed framework is reproducible, requires no proprietary scanner data, and provides a practical basis for post-deployment acquisition QA.
期刊介绍:
Academic Radiology publishes original reports of clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, image-guided interventions and related techniques. It also includes brief technical reports describing original observations, techniques, and instrumental developments; state-of-the-art reports on clinical issues, new technology and other topics of current medical importance; meta-analyses; scientific studies and opinions on radiologic education; and letters to the Editor.