Multicentre audit of the efficiency of hospital-wide central vacuum systems

IF 7.4 1区 医学 Q1 ANESTHESIOLOGY
Anaesthesia Pub Date : 2026-10-13 Epub Date: 2026-07-21 DOI:10.1111/anae.70307
Asmaa Haamid, Nawras Osman, Alexander Lang, Steven Chawk, Kenneth Barker
{"title":"Multicentre audit of the efficiency of hospital-wide central vacuum systems","authors":"Asmaa Haamid,&nbsp;Nawras Osman,&nbsp;Alexander Lang,&nbsp;Steven Chawk,&nbsp;Kenneth Barker","doi":"10.1111/anae.70307","DOIUrl":null,"url":null,"abstract":"<p>Vacuum suction is a critical clinical tool used to maintain airway clearance; manage respiratory conditions; and facilitate drainage therapy and specialised wound care. Most hospitals provide wall-mounted suction ports connected to a central vacuum system. These systems include multiple pumps; backup units; and extensive pipework. While designed for reliability, they are prone to entrainment and leaks due to the high number of end ports, resulting in energy consumption even when suction is not used actively. Portable on-demand suction devices, powered by batteries or mains electricity or driven by compressed air or oxygen through a Venturi effect, offer an alternative that operates only when required and may reduce environmental impact and costs.</p><p>A prospective audit was conducted across three UK hospitals (Raigmore Hospital, Inverness; Aberdeen Royal Infirmary, Aberdeen and Sheffield Children's Hospital, Sheffield). The data collected focused on equipment usage, with no patient interaction and included the number of suction ports for each ward and the frequency and duration of suction use. Electrical consumption was inferred from logbooks and kW ratings of pumps. Carbon emissions were calculated using UK government greenhouse gas conversion factors [<span>1</span>].</p><p>Across all hospitals, high-acuity areas such as operating theatres, endoscopy and ICUs reported daily suction use. In contrast, low-acuity wards showed substantial underutilisation. For example, single episodes of suction use of &lt; 30 s over the course of a month were reported in medical wards with ≤ 26 wall ports. Similar patterns were observed across the audited hospitals, indicating systemic overprovision of central vacuum system infrastructure (Table 1). Most suction procedures last &lt; 3 min, with many lasting &lt; 30 s. Usage for 3 min represents just 0.2% of a 24-h period, meaning the system remains clinically unused for 99.8% of the day while pumps maintain a continuous negative pressure.</p><p>Compressor logbooks at Raigmore Hospital revealed intermittent operation of 3 kW motors on a ‘duty cycle’ – sequentially running whenever the vacuum pressure falls to a lower limit. Even with minimal clinical demand, negative pressure must be maintained across the entire pipe network. Average weekly energy use was logged at 490.5 kWh, giving an annual total of 25,506 kWh which emits 4515 kg of carbon dioxide equivalent (CO<sub>2</sub>e) and equates to 1300 tonnes of CO<sub>2</sub>e emissions across the UK.</p><p>Annual financial costs for running the Raigmore Hospital central vacuum system pumps amount to £7396 ($9941, €8551) for electricity, not including maintenance. Based on 123,000 acute beds across the UK [<span>2</span>], the potential annual green dividend of reviewing central vacuum system provision is &gt; £2.0 million ($2.69 million, €2.31 million). For new build hospitals, a current stage 4 tender for a 450-bed hospital lists the central vacuum installation at £1.47 million ($1.97 million, €1.70 million) [personal communication].</p><p>The findings reveal a fundamental inefficiency in hospital-wide central vacuum system deployment. Whilst deemed essential in high-acuity areas, central vacuum system infrastructure is vastly underutilised in general wards. The design of the system, with the need to maintain vacuum over and above any allowable loss, multiple cycling pumps, extensive pipework and unavoidable leaks, means that energy consumption is driven not by clinical need but by the requirement to maintain a negative vacuum pressure.</p><p>Portable on-demand suction devices offer a practical alternative with the environmental advantage of reduced energy consumption. They operate only when needed in general wards and for emergencies as mains and battery powered units (though require maintenance and a battery replacement programme). Venturi units have the advantage of very low maintenance requirements and at a cost-point similar to the existing vacuum control devices. The units are powered by compressed medical air or oxygen, and some allow simultaneous vacuum and oxygen delivery. However, because they are powered by compressed gas, the embedded carbon and financial cost must be considered when analysing leanest service options. For high-volume surgical use in operating theatres and endoscopy units, on-demand systems such as Stryker Neptune (Stryker UK, Newbury, UK) [<span>3</span>] or Drainstar (M.A.S. Medical s.r.l., Racconigi, Italy) [<span>4</span>] allow direct discharge to grey water systems, which improves staff wellbeing by reducing moving and handling requirements, and decreases clinical waste tonnage and transport of surgical fluid for high temperature incineration [<span>5</span>]. Though the recurring cost of filter procurement for this equipment is high, it should be compared with the capital and maintenance costs of hospital-wide central vacuum systems.</p><p>This work shows that centralised vacuum systems are environmentally and financially unsustainable in low-acuity hospital wards and could be replaced by on-demand units, with efficient alternatives also available for high-acuity areas. Despite suction being used for &lt; 1% of the day in many areas, central vacuum system pumps must always maintain vacuum, generating substantial carbon emissions and financial costs. Transitioning to a hybrid model, or designing future hospitals without central vacuum systems, offers an opportunity to reduce carbon emissions, lower costs and maintain high standards of clinical care.</p>","PeriodicalId":7742,"journal":{"name":"Anaesthesia","volume":"81 11","pages":"1593-1594"},"PeriodicalIF":7.4000,"publicationDate":"2026-10-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/anae.70307","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Anaesthesia","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/anae.70307","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/21 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Vacuum suction is a critical clinical tool used to maintain airway clearance; manage respiratory conditions; and facilitate drainage therapy and specialised wound care. Most hospitals provide wall-mounted suction ports connected to a central vacuum system. These systems include multiple pumps; backup units; and extensive pipework. While designed for reliability, they are prone to entrainment and leaks due to the high number of end ports, resulting in energy consumption even when suction is not used actively. Portable on-demand suction devices, powered by batteries or mains electricity or driven by compressed air or oxygen through a Venturi effect, offer an alternative that operates only when required and may reduce environmental impact and costs.

A prospective audit was conducted across three UK hospitals (Raigmore Hospital, Inverness; Aberdeen Royal Infirmary, Aberdeen and Sheffield Children's Hospital, Sheffield). The data collected focused on equipment usage, with no patient interaction and included the number of suction ports for each ward and the frequency and duration of suction use. Electrical consumption was inferred from logbooks and kW ratings of pumps. Carbon emissions were calculated using UK government greenhouse gas conversion factors [1].

Across all hospitals, high-acuity areas such as operating theatres, endoscopy and ICUs reported daily suction use. In contrast, low-acuity wards showed substantial underutilisation. For example, single episodes of suction use of < 30 s over the course of a month were reported in medical wards with ≤ 26 wall ports. Similar patterns were observed across the audited hospitals, indicating systemic overprovision of central vacuum system infrastructure (Table 1). Most suction procedures last < 3 min, with many lasting < 30 s. Usage for 3 min represents just 0.2% of a 24-h period, meaning the system remains clinically unused for 99.8% of the day while pumps maintain a continuous negative pressure.

Compressor logbooks at Raigmore Hospital revealed intermittent operation of 3 kW motors on a ‘duty cycle’ – sequentially running whenever the vacuum pressure falls to a lower limit. Even with minimal clinical demand, negative pressure must be maintained across the entire pipe network. Average weekly energy use was logged at 490.5 kWh, giving an annual total of 25,506 kWh which emits 4515 kg of carbon dioxide equivalent (CO2e) and equates to 1300 tonnes of CO2e emissions across the UK.

Annual financial costs for running the Raigmore Hospital central vacuum system pumps amount to £7396 ($9941, €8551) for electricity, not including maintenance. Based on 123,000 acute beds across the UK [2], the potential annual green dividend of reviewing central vacuum system provision is > £2.0 million ($2.69 million, €2.31 million). For new build hospitals, a current stage 4 tender for a 450-bed hospital lists the central vacuum installation at £1.47 million ($1.97 million, €1.70 million) [personal communication].

The findings reveal a fundamental inefficiency in hospital-wide central vacuum system deployment. Whilst deemed essential in high-acuity areas, central vacuum system infrastructure is vastly underutilised in general wards. The design of the system, with the need to maintain vacuum over and above any allowable loss, multiple cycling pumps, extensive pipework and unavoidable leaks, means that energy consumption is driven not by clinical need but by the requirement to maintain a negative vacuum pressure.

Portable on-demand suction devices offer a practical alternative with the environmental advantage of reduced energy consumption. They operate only when needed in general wards and for emergencies as mains and battery powered units (though require maintenance and a battery replacement programme). Venturi units have the advantage of very low maintenance requirements and at a cost-point similar to the existing vacuum control devices. The units are powered by compressed medical air or oxygen, and some allow simultaneous vacuum and oxygen delivery. However, because they are powered by compressed gas, the embedded carbon and financial cost must be considered when analysing leanest service options. For high-volume surgical use in operating theatres and endoscopy units, on-demand systems such as Stryker Neptune (Stryker UK, Newbury, UK) [3] or Drainstar (M.A.S. Medical s.r.l., Racconigi, Italy) [4] allow direct discharge to grey water systems, which improves staff wellbeing by reducing moving and handling requirements, and decreases clinical waste tonnage and transport of surgical fluid for high temperature incineration [5]. Though the recurring cost of filter procurement for this equipment is high, it should be compared with the capital and maintenance costs of hospital-wide central vacuum systems.

This work shows that centralised vacuum systems are environmentally and financially unsustainable in low-acuity hospital wards and could be replaced by on-demand units, with efficient alternatives also available for high-acuity areas. Despite suction being used for < 1% of the day in many areas, central vacuum system pumps must always maintain vacuum, generating substantial carbon emissions and financial costs. Transitioning to a hybrid model, or designing future hospitals without central vacuum systems, offers an opportunity to reduce carbon emissions, lower costs and maintain high standards of clinical care.

多中心审核全院中央真空系统的效率。
真空吸盘是维持气道通畅的重要临床工具;管理呼吸系统疾病;并促进引流治疗和专门的伤口护理。大多数医院提供与中央真空系统相连的壁挂式吸口。这些系统包括多个泵;备份单元;还有大量的管道工程。虽然设计是为了提高可靠性,但由于末端端口数量多,它们容易夹带和泄漏,即使在不主动使用吸力时也会导致能源消耗。便携式按需抽吸装置,由电池或电源供电,或通过文丘里效应由压缩空气或氧气驱动,提供了一种仅在需要时运行的替代方案,可以减少对环境的影响和成本。对三家英国医院(因弗内斯的Raigmore医院、阿伯丁的阿伯丁皇家医院和谢菲尔德的谢菲尔德儿童医院)进行了前瞻性审计。收集的数据主要集中在设备使用情况,没有患者互动,包括每个病房的吸痰口数量、吸痰使用的频率和持续时间。电力消耗是从日志和泵的千瓦额定值推断出来的。碳排放量是用英国政府温室气体转换系数[1]计算的。在所有医院中,高视力区域,如手术室、内窥镜检查和icu报告了每日吸痰使用情况。相比之下,低视力病房显示出大量的利用不足。例如,在≤26个壁口的病房中,报告了一个月内单次使用30s的吸痰。在被审计的医院中也观察到类似的模式,表明中央真空系统基础设施的系统性供应过剩(表1)。大多数抽吸过程持续3分钟,许多持续30秒。3分钟的使用时间仅占24小时时间的0.2%,这意味着在泵保持连续负压的情况下,系统在99.8%的时间内处于临床未使用状态。Raigmore医院的压缩机日志显示,在“占空比”下,3kw电机间歇性运行,只要真空压力降至下限,就会依次运行。即使临床需求最小,整个管网也必须保持负压。平均每周的能源使用量为490.5千瓦时,每年的总使用量为25,506千瓦时,排放了4515公斤二氧化碳当量(CO2e),相当于整个英国排放了1300吨二氧化碳当量。每年运行Raigmore医院中央真空系统泵的电费高达7396英镑(9941美元,8551欧元),这还不包括维护费用。基于英国123,000个急症床位,审查中央真空系统提供的潜在年度绿色红利为200万英镑(269万美元,231万欧元)。对于新建医院来说,目前第四阶段的450个床位的招标中,中央真空装置的价格为147万英镑(197万美元,170万欧元)[个人通讯]。调查结果揭示了医院范围内中央真空系统部署的根本效率低下。虽然在高敏度地区被认为是必不可少的,但中央真空系统基础设施在普通病房的利用率远远不足。系统的设计需要保持真空超过任何允许的损失,多个循环泵,广泛的管道和不可避免的泄漏,这意味着能源消耗不是由临床需要驱动,而是由保持负真空压力的要求驱动。便携式按需抽吸装置提供了一种实用的替代方案,具有降低能耗的环境优势。它们仅在普通病房和紧急情况需要时作为电源和电池供电设备运行(尽管需要维护和更换电池方案)。文丘里装置的优点是维护要求非常低,成本与现有的真空控制装置相似。这些装置由压缩的医用空气或氧气提供动力,有些可以同时进行真空和氧气输送。然而,由于它们是由压缩气体驱动的,因此在分析最节能的服务选择时,必须考虑隐含的碳和财务成本。对于手术室和内镜检查单位的大量手术使用,按需系统如Stryker Neptune (Stryker UK, Newbury, UK)[3]或Drainstar (M.A.S. Medical s.r.l, Racconigi, Italy)[4]允许直接排放到灰水系统,通过减少移动和处理要求来改善员工的健康,并减少用于高温焚烧的医疗废物吨位和手术液的运输[5]。虽然这种设备的过滤器采购的经常性成本很高,但它应该与医院范围内的中央真空系统的资金和维护成本进行比较。 这项工作表明,集中式真空系统在环境和经济上都是不可持续的,在低锐度的医院病房,可以被按需设备所取代,在高锐度的地区也有有效的替代品。尽管在许多地区,每天有1%的时间使用吸力,但中央真空系统泵必须始终保持真空,从而产生大量的碳排放和财务成本。过渡到混合模式,或者设计没有中央真空系统的未来医院,提供了减少碳排放、降低成本和保持高标准临床护理的机会。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Anaesthesia
Anaesthesia 医学-麻醉学
CiteScore
21.20
自引率
9.30%
发文量
300
审稿时长
6 months
期刊介绍: The official journal of the Association of Anaesthetists is Anaesthesia. It is a comprehensive international publication that covers a wide range of topics. The journal focuses on general and regional anaesthesia, as well as intensive care and pain therapy. It includes original articles that have undergone peer review, covering all aspects of these fields, including research on equipment.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书