Fractionated Atrial Electrograms Identify Sites of Autonomic Responses During Pulmonary Vein Isolation: Insights for Cardioneuroablation

IF 1.9 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Juan F. Agudelo-Uribe, Rafael Correa-Velásquez, Juan D. Ramírez-Barrera, Margarita Londoño-Arango, Teresa Barrio-López, Jesús Almendral
{"title":"Fractionated Atrial Electrograms Identify Sites of Autonomic Responses During Pulmonary Vein Isolation: Insights for Cardioneuroablation","authors":"Juan F. Agudelo-Uribe,&nbsp;Rafael Correa-Velásquez,&nbsp;Juan D. Ramírez-Barrera,&nbsp;Margarita Londoño-Arango,&nbsp;Teresa Barrio-López,&nbsp;Jesús Almendral","doi":"10.1002/joa3.70423","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Cardioneuroablation (CNA) targets atrial ganglionated plexuses (GP) to treat vagally mediated bradyarrhythmias. Automated electrogram fractionation mapping has been proposed as a surrogate tool for GP localization, but optimal parameter settings remain undefined. We aimed to systematically evaluate automated fractionation mapping parameters and their association with autonomic responses during ablation.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Data from 26 patients were analyzed, including 12 patients undergoing atrial fibrillation (AF) ablation in sinus rhythm and 14 patients undergoing CNA. Using high-density electroanatomical mapping, three predefined values were tested for each fractionation parameter—signal width, refractoriness, and amplitude threshold—yielding 27 automated configurations. A purely anatomical localization strategy was also evaluated. In the AF cohort, all configurations were systematically assessed, generating 336 fractionation maps and 870 ablation sites. Autonomic response during radiofrequency delivery was defined as an absolute RR interval change &gt; 10%. Selected high-performing configurations were subsequently evaluated in the CNA cohort (614 ablation sites).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>In the AF cohort, 25 of 27 configurations showed a significant association with autonomic response. After Bonferroni correction, 15 configurations remained significant. The configuration with signal width 15 ms, refractoriness 25 ms, and amplitude threshold 0.05 mV demonstrated the highest sensitivity, although several configurations showed comparable performance. When evaluated in the CNA cohort, this configuration yielded a sensitivity of 74.1% and specificity of 73.9%.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>This pilot study provides a reproducible framework for evaluating automated fractionation mapping parameters associated with autonomic responses during CNA. The findings support biologically plausible parameter–performance relationships and identify parameter ranges potentially suitable for procedural use.</p>\n </section>\n </div>","PeriodicalId":15174,"journal":{"name":"Journal of Arrhythmia","volume":"42 4","pages":""},"PeriodicalIF":1.9000,"publicationDate":"2026-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/joa3.70423","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Arrhythmia","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/joa3.70423","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Cardioneuroablation (CNA) targets atrial ganglionated plexuses (GP) to treat vagally mediated bradyarrhythmias. Automated electrogram fractionation mapping has been proposed as a surrogate tool for GP localization, but optimal parameter settings remain undefined. We aimed to systematically evaluate automated fractionation mapping parameters and their association with autonomic responses during ablation.

Methods

Data from 26 patients were analyzed, including 12 patients undergoing atrial fibrillation (AF) ablation in sinus rhythm and 14 patients undergoing CNA. Using high-density electroanatomical mapping, three predefined values were tested for each fractionation parameter—signal width, refractoriness, and amplitude threshold—yielding 27 automated configurations. A purely anatomical localization strategy was also evaluated. In the AF cohort, all configurations were systematically assessed, generating 336 fractionation maps and 870 ablation sites. Autonomic response during radiofrequency delivery was defined as an absolute RR interval change > 10%. Selected high-performing configurations were subsequently evaluated in the CNA cohort (614 ablation sites).

Results

In the AF cohort, 25 of 27 configurations showed a significant association with autonomic response. After Bonferroni correction, 15 configurations remained significant. The configuration with signal width 15 ms, refractoriness 25 ms, and amplitude threshold 0.05 mV demonstrated the highest sensitivity, although several configurations showed comparable performance. When evaluated in the CNA cohort, this configuration yielded a sensitivity of 74.1% and specificity of 73.9%.

Conclusions

This pilot study provides a reproducible framework for evaluating automated fractionation mapping parameters associated with autonomic responses during CNA. The findings support biologically plausible parameter–performance relationships and identify parameter ranges potentially suitable for procedural use.

Abstract Image

分割心房电图识别肺静脉隔离期间自主神经反应的位置:对心血管消融的见解
背景:心神经消融(CNA)以心房神经节丛(GP)为靶点治疗迷走神经介导的慢速心律失常。自动电图分割映射已被提议作为GP定位的替代工具,但最佳参数设置仍未确定。我们的目的是系统地评估消融过程中自动分割线定位参数及其与自主神经反应的关系。方法分析26例患者的资料,其中12例行窦性心律房颤(AF)消融,14例行CNA。利用高密度电解剖图,对每个分馏参数(信号宽度、耐火度和振幅阈值)进行了三个预定义值测试,从而产生27种自动化配置。纯解剖定位策略也进行了评估。在房颤队列中,系统评估了所有配置,生成了336个分块图和870个消融点。射频输送过程中的自主反应被定义为绝对RR间隔变化>; 10%。随后在CNA队列(614个消融点)中评估选定的高性能配置。结果在房颤队列中,27种构型中有25种与自主神经反应显著相关。经Bonferroni校正后,仍有15种构型显著。信号宽度为15 ms,耐火度为25 ms,幅度阈值为0.05 mV的配置显示出最高的灵敏度,尽管其他配置的性能相当。当在CNA队列中进行评估时,该配置的敏感性为74.1%,特异性为73.9%。该初步研究提供了一个可重复的框架,用于评估与CNA期间自主神经反应相关的自动分馏图参数。研究结果支持生物学上合理的参数-性能关系,并确定了可能适合程序使用的参数范围。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Arrhythmia
Journal of Arrhythmia CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
2.90
自引率
10.00%
发文量
127
审稿时长
45 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书