Highlighting the Importance of Authentic Reference Chemicals in the Unambiguous Identification of Unknowns during Routine Sample Analysis─An OPCW Proficiency Test Case
Carolyn J. Koester, Teneile M. Alfaro, David S. Cho, Todd H. Corzett, Mark L. Dreyer, Saphon Hok, Derrick C. Kaseman, Adele F. Panasci-Nott, Alexander K. Vu, Armando Alcaraz, Carlos A. Valdez
{"title":"Highlighting the Importance of Authentic Reference Chemicals in the Unambiguous Identification of Unknowns during Routine Sample Analysis─An OPCW Proficiency Test Case","authors":"Carolyn J. Koester, Teneile M. Alfaro, David S. Cho, Todd H. Corzett, Mark L. Dreyer, Saphon Hok, Derrick C. Kaseman, Adele F. Panasci-Nott, Alexander K. Vu, Armando Alcaraz, Carlos A. Valdez","doi":"10.1021/jasms.6c00191","DOIUrl":null,"url":null,"abstract":"<p>The unequivocal identification\r\nof unknown chemicals during routine\r\nsample analysis is a constant occurrence in the field of analytical\r\nchemistry. The process can be painstakingly tedious, particularly\r\nin situations where a tentatively identified unknown may possess isomeric\r\ncounterparts yielding identical accurate mass values and very similar\r\nmass spectra. In this work, we present the experimental process involved\r\nin the correct identification of 1,4-oxathiane 4,4-dioxide and differentiation\r\nfrom its constitutional isomer, 2-hydroxyethyl vinyl sulfone, when\r\npresent in a silica gel matrix featured in the 54th Environmental\r\nOrganisation for the Prohibition of Chemical Weapons (OPCW) proficiency\r\ntest. After discovering that the library-generated mass spectrum for\r\n2-hydroxyethyl vinyl sulfone in a preliminary gas chromatography–mass\r\nspectrometry (GC-MS) analysis did not match the one from an authentic\r\nreference chemical, we embarked on an unknown structure determination\r\ncampaign involving GC-MS, liquid chromatography-tandem mass spectrometry\r\n(LC-MS-MS), and nuclear magnetic resonance (NMR) spectroscopy. All\r\nof the information obtained, coupled to the synthesis of an authentic\r\nreference chemical, was used to determine the identity of the unknown\r\nas 1,4-oxathiane 4,4-dioxide. The process described in this work highlights\r\nthe important role played by authentic reference chemicals in the\r\nidentification of unknown chemicals during routine sample analysis.</p>","PeriodicalId":672,"journal":{"name":"Journal of the American Society for Mass Spectrometry","volume":"37 8","pages":"1718–1723"},"PeriodicalIF":2.9000,"publicationDate":"2026-08-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the American Society for Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://pubs.acs.org/doi/10.1021/jasms.6c00191","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/13 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
The unequivocal identification
of unknown chemicals during routine
sample analysis is a constant occurrence in the field of analytical
chemistry. The process can be painstakingly tedious, particularly
in situations where a tentatively identified unknown may possess isomeric
counterparts yielding identical accurate mass values and very similar
mass spectra. In this work, we present the experimental process involved
in the correct identification of 1,4-oxathiane 4,4-dioxide and differentiation
from its constitutional isomer, 2-hydroxyethyl vinyl sulfone, when
present in a silica gel matrix featured in the 54th Environmental
Organisation for the Prohibition of Chemical Weapons (OPCW) proficiency
test. After discovering that the library-generated mass spectrum for
2-hydroxyethyl vinyl sulfone in a preliminary gas chromatography–mass
spectrometry (GC-MS) analysis did not match the one from an authentic
reference chemical, we embarked on an unknown structure determination
campaign involving GC-MS, liquid chromatography-tandem mass spectrometry
(LC-MS-MS), and nuclear magnetic resonance (NMR) spectroscopy. All
of the information obtained, coupled to the synthesis of an authentic
reference chemical, was used to determine the identity of the unknown
as 1,4-oxathiane 4,4-dioxide. The process described in this work highlights
the important role played by authentic reference chemicals in the
identification of unknown chemicals during routine sample analysis.
期刊介绍:
The Journal of the American Society for Mass Spectrometry presents research papers covering all aspects of mass spectrometry, incorporating coverage of fields of scientific inquiry in which mass spectrometry can play a role.
Comprehensive in scope, the journal publishes papers on both fundamentals and applications of mass spectrometry. Fundamental subjects include instrumentation principles, design, and demonstration, structures and chemical properties of gas-phase ions, studies of thermodynamic properties, ion spectroscopy, chemical kinetics, mechanisms of ionization, theories of ion fragmentation, cluster ions, and potential energy surfaces. In addition to full papers, the journal offers Communications, Application Notes, and Accounts and Perspectives