Evaluation of deep learning-based reconstruction models on non-TOF BGO PET/CT: impact of acquisition times and BSREM penalization factors on lesion detectability and SNR.

IF 3.5 2区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Anna Stenvall, David Minarik, Elin Trägårdh, Sofia Kvernby
{"title":"Evaluation of deep learning-based reconstruction models on non-TOF BGO PET/CT: impact of acquisition times and BSREM penalization factors on lesion detectability and SNR.","authors":"Anna Stenvall, David Minarik, Elin Trägårdh, Sofia Kvernby","doi":"10.1186/s40658-026-00919-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>New long field-of-view (FOV) PET scanners using bismuth germanate (BGO) detectors without time-of-flight (TOF) capability are now available. These systems incorporate deep learning-based TOF (DLb-TOF) models to compensate for the absence of TOF. There is a lack of studies systematically investigating the optimal balance between signal to noise ratio and lesion detectability across a broader range of acquisition times and β-values for these DLb-TOF models. This study aims to evaluate the trade-off between acquisition time, signal-to-noise ratio (SNR) and lesion detectability to guide optimization of clinical protocol.</p><p><strong>Materials and methods: </strong>Twenty patients referred for a clinical [<sup>18</sup>F]fluorodeoxyglucose (FDG) PET scan were included. Each patient received 3.5 MBq/kg of [<sup>18</sup>F]FDG and underwent a whole-body PET acquisition (120 s/bed) on a digital BGO PET/CT (32 cm FOV) 60 min post-injection. Data were reconstructed into images (384 × 384 matrix) representing different acquisition times (120 s, 90 s, 60 s, 45 s, 30 s and 15 s) using BSREM with β-values ranging from 50 to 1100. Three DLb-TOF models (Low, Medium, High) were applied. Volumes of interest were placed in the liver and two avid lesions per patient. SNR were calculated as SUVmean<sub>liver</sub>/SD<sub>liver</sub> and detectability were calculated as SUVpeak<sub>tumor</sub>/SUVpeak<sub>liver</sub>.</p><p><strong>Results: </strong>SNR increased with longer acquisition times and higher β-values. DLb-TOF models improved SNR across all settings, with the Low DLb-TOF model producing the largest increase. Lesion detectability depended on the acquisition time and β-value. At longer acquisition times (120 s, 90 s), β100 provided the highest detectability, while shorter times (60-15 s) required higher β-value (β300) for optimal detectability. Among DLb-TOF models, the High model gave the best detectability overall, though the Low model performed better at lower β-values.</p><p><strong>Conclusion: </strong>SNR increased with higher β-values, longer acquisition times, and DLb-TOF application. Lesion detectability, defined as the ratio of SUV<sub>peak</sub> in the lesion to SUV<sub>peak</sub> in the liver, depended on the β-value, acquisition time, and the DLb-TOF model used. The Low DLb-TOF model had the best SNR but at the expense of detectability. The optimal parameters for the evaluated BGO PET/CT system, balancing SNR and lesion detectability within a clinical reasonable acquisition time, were 60-90 s with β-values of 500-300, in combination with the Medium DLb-TOF model, when 3.5 MBq/kg [<sup>18</sup>F]FDG was administered.</p>","PeriodicalId":11559,"journal":{"name":"EJNMMI Physics","volume":" ","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-07-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13369060/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40658-026-00919-8","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: New long field-of-view (FOV) PET scanners using bismuth germanate (BGO) detectors without time-of-flight (TOF) capability are now available. These systems incorporate deep learning-based TOF (DLb-TOF) models to compensate for the absence of TOF. There is a lack of studies systematically investigating the optimal balance between signal to noise ratio and lesion detectability across a broader range of acquisition times and β-values for these DLb-TOF models. This study aims to evaluate the trade-off between acquisition time, signal-to-noise ratio (SNR) and lesion detectability to guide optimization of clinical protocol.

Materials and methods: Twenty patients referred for a clinical [18F]fluorodeoxyglucose (FDG) PET scan were included. Each patient received 3.5 MBq/kg of [18F]FDG and underwent a whole-body PET acquisition (120 s/bed) on a digital BGO PET/CT (32 cm FOV) 60 min post-injection. Data were reconstructed into images (384 × 384 matrix) representing different acquisition times (120 s, 90 s, 60 s, 45 s, 30 s and 15 s) using BSREM with β-values ranging from 50 to 1100. Three DLb-TOF models (Low, Medium, High) were applied. Volumes of interest were placed in the liver and two avid lesions per patient. SNR were calculated as SUVmeanliver/SDliver and detectability were calculated as SUVpeaktumor/SUVpeakliver.

Results: SNR increased with longer acquisition times and higher β-values. DLb-TOF models improved SNR across all settings, with the Low DLb-TOF model producing the largest increase. Lesion detectability depended on the acquisition time and β-value. At longer acquisition times (120 s, 90 s), β100 provided the highest detectability, while shorter times (60-15 s) required higher β-value (β300) for optimal detectability. Among DLb-TOF models, the High model gave the best detectability overall, though the Low model performed better at lower β-values.

Conclusion: SNR increased with higher β-values, longer acquisition times, and DLb-TOF application. Lesion detectability, defined as the ratio of SUVpeak in the lesion to SUVpeak in the liver, depended on the β-value, acquisition time, and the DLb-TOF model used. The Low DLb-TOF model had the best SNR but at the expense of detectability. The optimal parameters for the evaluated BGO PET/CT system, balancing SNR and lesion detectability within a clinical reasonable acquisition time, were 60-90 s with β-values of 500-300, in combination with the Medium DLb-TOF model, when 3.5 MBq/kg [18F]FDG was administered.

基于深度学习的非tof BGO PET/CT重建模型评价:采集时间和BSREM惩罚因子对病灶可检测性和信噪比的影响
背景:使用没有飞行时间(TOF)能力的锗酸铋(BGO)探测器的新型长视场(FOV) PET扫描仪现在可用。这些系统结合了基于深度学习的TOF (DLb-TOF)模型来弥补TOF的缺失。对于这些DLb-TOF模型,在更广泛的采集时间和β值范围内,信号噪声比和病变可检测性之间的最佳平衡缺乏系统的研究。本研究旨在评估采集时间、信噪比(SNR)和病变可检测性之间的权衡,以指导临床方案的优化。材料和方法:纳入20例进行临床[18F]氟脱氧葡萄糖(FDG) PET扫描的患者。每位患者接受3.5 MBq/kg [18F]FDG注射,并在注射后60分钟在数字BGO PET/CT (32 cm FOV)上进行全身PET采集(120秒/张)。采用β值为50 ~ 1100的BSREM模型,将数据重构成不同采集时间(120、90、60、45、30、15 s)的图像(384 × 384矩阵)。采用低、中、高3种DLb-TOF模型。感兴趣的体积被放置在肝脏和每个病人的两个严重病变。信噪比计算为SUVmeanliver/SDliver,检出率计算为SUVpeaktumor/SUVpeakliver。结果:信噪比随采集时间的延长和β值的增大而增大。DLb-TOF模型在所有设置下都提高了信噪比,其中Low DLb-TOF模型的提高幅度最大。病灶的可检出性取决于采集时间和β值。在较长的采集时间(120 s, 90 s), β100提供最高的可探测性,而较短的采集时间(60-15 s)需要较高的β值(β300)才能获得最佳的可探测性。在DLb-TOF模型中,虽然低模型在较低的β值下表现较好,但高模型的总体可探测性最好。结论:β值越高,采集时间越长,应用DLb-TOF时信噪比越高。病变可检测性,定义为病变中SUVpeak与肝脏中SUVpeak的比值,取决于β值、采集时间和所使用的DLb-TOF模型。低DLb-TOF模型具有最好的信噪比,但以可探测性为代价。当给药3.5 MBq/kg [18F]FDG时,所评估的BGO PET/CT系统的最佳参数为60-90 s, β值为500-300,在临床合理采集时间内平衡信噪比和病变可检出性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
EJNMMI Physics
EJNMMI Physics Physics and Astronomy-Radiation
CiteScore
6.70
自引率
10.00%
发文量
78
审稿时长
13 weeks
期刊介绍: EJNMMI Physics is an international platform for scientists, users and adopters of nuclear medicine with a particular interest in physics matters. As a companion journal to the European Journal of Nuclear Medicine and Molecular Imaging, this journal has a multi-disciplinary approach and welcomes original materials and studies with a focus on applied physics and mathematics as well as imaging systems engineering and prototyping in nuclear medicine. This includes physics-driven approaches or algorithms supported by physics that foster early clinical adoption of nuclear medicine imaging and therapy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书