Interobserver Variability of Intraoperative Soft Tissue Laxity Assessment in Robotic-Assisted Total Knee Arthroplasty Using the ROSA System.

IF 1.4 4区 医学 Q3 ORTHOPEDICS
Hiroki Kishi, Daisuke Nishiyama, Daisuke Fukui, Manabu Yamanaka, Hiroshi Yamada
{"title":"Interobserver Variability of Intraoperative Soft Tissue Laxity Assessment in Robotic-Assisted Total Knee Arthroplasty Using the ROSA System.","authors":"Hiroki Kishi, Daisuke Nishiyama, Daisuke Fukui, Manabu Yamanaka, Hiroshi Yamada","doi":"10.1055/a-2908-1899","DOIUrl":null,"url":null,"abstract":"<p><strong>Abstract: </strong>This study evaluated the interobserver reliability of intraoperative soft tissue laxity assessment during robotic-arm-assisted total knee arthroplasty (TKA) using the Robotic Surgical Assistant (ROSA) system. Although robotic systems have improved the precision of bone alignment, the reproducibility of soft tissue evaluations-critical for achieving joint balance-remains uncertain. We retrospectively analyzed 37 knees from 30 patients who underwent ROSA-assisted TKA. Soft tissue laxity was manually assessed under varus and valgus stress at extension and 90 degrees of knee flexion. Two parameters were recorded: joint gap widening and hip-knee angle (HKA). Interobserver agreement was evaluated using intraclass correlation coefficients (ICC), Pearson's correlation coefficients, and Bland-Altman analysis. Angular measurements, particularly HKA at extension, demonstrated excellent interobserver reliability (ICC = 0.94). In contrast, gap-based measurements showed greater variability, with lateral laxity under varus stress at extension demonstrating poor reliability (ICC = 0.31). Agreement between observers was generally stronger at extension than at 90 degrees of flexion. Contralateral assessments showed better consistency. While robotic systems improve alignment accuracy in TKA, intraoperative soft tissue laxity evaluation remains prone to interobserver variability. Gap-based measurements are less reliable than angular assessments, underscoring the need for standardized evaluation methods to improve consistency in soft tissue evaluation during robotic TKA.</p>","PeriodicalId":48798,"journal":{"name":"Journal of Knee Surgery","volume":" ","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2026-07-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Knee Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2908-1899","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Abstract: This study evaluated the interobserver reliability of intraoperative soft tissue laxity assessment during robotic-arm-assisted total knee arthroplasty (TKA) using the Robotic Surgical Assistant (ROSA) system. Although robotic systems have improved the precision of bone alignment, the reproducibility of soft tissue evaluations-critical for achieving joint balance-remains uncertain. We retrospectively analyzed 37 knees from 30 patients who underwent ROSA-assisted TKA. Soft tissue laxity was manually assessed under varus and valgus stress at extension and 90 degrees of knee flexion. Two parameters were recorded: joint gap widening and hip-knee angle (HKA). Interobserver agreement was evaluated using intraclass correlation coefficients (ICC), Pearson's correlation coefficients, and Bland-Altman analysis. Angular measurements, particularly HKA at extension, demonstrated excellent interobserver reliability (ICC = 0.94). In contrast, gap-based measurements showed greater variability, with lateral laxity under varus stress at extension demonstrating poor reliability (ICC = 0.31). Agreement between observers was generally stronger at extension than at 90 degrees of flexion. Contralateral assessments showed better consistency. While robotic systems improve alignment accuracy in TKA, intraoperative soft tissue laxity evaluation remains prone to interobserver variability. Gap-based measurements are less reliable than angular assessments, underscoring the need for standardized evaluation methods to improve consistency in soft tissue evaluation during robotic TKA.

机器人辅助全膝关节置换术中使用ROSA系统评估术中软组织松弛度的观察者间差异。
目的:本研究利用机器人手术助理(ROSA)系统评估机械臂辅助全膝关节置换术(TKA)中术中软组织松弛度评估的观察者间可靠性。尽管机器人系统已经提高了骨对齐的精度,但软组织评估的可重复性(对实现关节平衡至关重要)仍然不确定。方法:我们回顾性分析了30例接受rosa辅助TKA的患者的37个膝关节。在膝关节伸展和屈曲90°时,手动评估内翻和外翻应力下的软组织松弛程度。记录两个参数:关节间隙扩大和髋膝角(HKA)。使用类内相关系数(ICC)、Pearson相关系数和Bland-Altman图评估观察者间的一致性。结果:角度测量,特别是扩展时的HKA,显示出极好的观察者间信度(ICC = 0.94)。相比之下,基于间隙的测量显示出更大的可变性,在伸展时内翻应力下的侧向松弛显示出较差的可靠性(ICC = 0.31)。观察者之间的一致意见通常在伸展时比在90°屈曲时更强。对侧评估显示较好的一致性。结论:虽然机器人系统提高了TKA的对准精度,但术中软组织松弛度评估仍然容易出现观察者之间的差异。基于间隙的测量不如角度评估可靠,强调了标准化评估方法的必要性,以提高机器人TKA期间软组织评估的一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.50
自引率
5.90%
发文量
139
期刊介绍: The Journal of Knee Surgery covers a range of issues relating to the orthopaedic techniques of arthroscopy, arthroplasty, and reconstructive surgery of the knee joint. In addition to original peer-review articles, this periodical provides details on emerging surgical techniques, as well as reviews and special focus sections. Topics of interest include cruciate ligament repair and reconstruction, bone grafting, cartilage regeneration, and magnetic resonance imaging.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书