Development and validation of a pulmonary embolism quantification method using single photon emission computed tomography ventilation and perfusion scintigraphy.
IF 2.7 3区 医学Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Harini Fernando, Christopher Sibley-Allen, Charlotte Fowler, Livieratos Lefteris
{"title":"Development and validation of a pulmonary embolism quantification method using single photon emission computed tomography ventilation and perfusion scintigraphy.","authors":"Harini Fernando, Christopher Sibley-Allen, Charlotte Fowler, Livieratos Lefteris","doi":"10.1186/s13550-026-01379-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pulmonary embolism (PE) remains an important cause of cardiovascular death and morbidity worldwide. Conventionally, PE imaging is reported in a binary manner, with reports stating the presence or absence of a thrombus. No generally accepted graduated quantification systems are available to guide management or predict the development of acute or long-term complications. V/Q SPECT is a well-established method for diagnosing PE in acute and chronic settings. The use of volumetric perfusion and ventilation functional images could potentially calculate the volumetric PE burden. This project aimed to develop and optimise a volumetric V/Q SPECT PE quantification technique, utilising V/Q SPECT phantoms, 100 consecutive clinical V/Q SPECT studies, and advanced V/Q SPECT image analysis with semiautomated %PE quantification methods.</p><p><strong>Results: </strong>Lung phantoms with PE defects of known volume demonstrated that the choice of acquisition, reconstruction parameters, and the volumetric cutoff threshold influences %PE results. The V/Q quotient image utilised a semiautomated technique with phantom-derived optimisation parameters, which was found to be the most accurate quantification method.</p><p><strong>Conclusion: </strong>This study, combining phantom and clinical study analysis, developed a successful method for semiautomated V/Q SPECT %PE quantification, which has the potential utility of guiding patient management and outcome prediction.</p>","PeriodicalId":11611,"journal":{"name":"EJNMMI Research","volume":" ","pages":""},"PeriodicalIF":2.7000,"publicationDate":"2026-07-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13627622/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s13550-026-01379-y","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Pulmonary embolism (PE) remains an important cause of cardiovascular death and morbidity worldwide. Conventionally, PE imaging is reported in a binary manner, with reports stating the presence or absence of a thrombus. No generally accepted graduated quantification systems are available to guide management or predict the development of acute or long-term complications. V/Q SPECT is a well-established method for diagnosing PE in acute and chronic settings. The use of volumetric perfusion and ventilation functional images could potentially calculate the volumetric PE burden. This project aimed to develop and optimise a volumetric V/Q SPECT PE quantification technique, utilising V/Q SPECT phantoms, 100 consecutive clinical V/Q SPECT studies, and advanced V/Q SPECT image analysis with semiautomated %PE quantification methods.
Results: Lung phantoms with PE defects of known volume demonstrated that the choice of acquisition, reconstruction parameters, and the volumetric cutoff threshold influences %PE results. The V/Q quotient image utilised a semiautomated technique with phantom-derived optimisation parameters, which was found to be the most accurate quantification method.
Conclusion: This study, combining phantom and clinical study analysis, developed a successful method for semiautomated V/Q SPECT %PE quantification, which has the potential utility of guiding patient management and outcome prediction.
EJNMMI ResearchRADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING&nb-
CiteScore
5.90
自引率
3.10%
发文量
72
审稿时长
13 weeks
期刊介绍:
EJNMMI Research publishes new basic, translational and clinical research in the field of nuclear medicine and molecular imaging. Regular features include original research articles, rapid communication of preliminary data on innovative research, interesting case reports, editorials, and letters to the editor. Educational articles on basic sciences, fundamental aspects and controversy related to pre-clinical and clinical research or ethical aspects of research are also welcome. Timely reviews provide updates on current applications, issues in imaging research and translational aspects of nuclear medicine and molecular imaging technologies.
The main emphasis is placed on the development of targeted imaging with radiopharmaceuticals within the broader context of molecular probes to enhance understanding and characterisation of the complex biological processes underlying disease and to develop, test and guide new treatment modalities, including radionuclide therapy.