Judith C. A. Cluitmans, Piet Meijer, Sanna R. Rijpma, An K. Stroobants
{"title":"Accuracy of Factor VIII Assays for Measuring Damoctocog Alfa Pegol: A Comparative Analysis","authors":"Judith C. A. Cluitmans, Piet Meijer, Sanna R. Rijpma, An K. Stroobants","doi":"10.1111/hae.70350","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Accurate monitoring of treatment with modified factor VIII products remains challenging, due to known variability in assay performance across different laboratory methods and reagents. For the extended half-life product Damoctocog alfa pegol discrepancies have been reported.</p>\n </section>\n \n <section>\n \n <h3> Aims</h3>\n \n <p>This multicentre field study was designed to evaluate the analytical accuracy of Damoctocog alfa pegol using various methods for one-stage clotting and chromogenic assays.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The study was conducted under conditions closely mimicking routine laboratory settings. Five plasma samples were spiked with varying concentrations of Damoctocog alfa pegol (0.05 to 1.3 IU/ml) and distributed to 118 participating laboratories. FVIII activity of the samples was analysed using 12 one-stage and 7 chromogenic methods. Accuracy results were evaluated per specific reagent with a maximal acceptable mean bias of 20% from the target. Interlaboratory variation was established for reagents with ≥3 participants.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The reliability of factor VIII measurements differ between the tested reagents. Usage of SynthasIL and Pathromtin (one stage) and Biophen (chromogenic) is in most cases reliable, although inter-lab variation contributes to inaccuracy. The reagents STA PTT A and Triniclot (one stage), as well as Siemens (chromogenic) (strongly) underestimate FVIII activity and are therefore unsuitable for accurate measurement of Damoctocog alfa pegol.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>We have evaluated accuracy results for 19 Factor VIII assays in order to provide evidence-based guidance for clinical laboratories in selecting appropriate assay methodologies for reliable quantification of the activity of Damoctocog alfa pegol. Assay performance should be locally validated as large interlaboratory variation is observed.</p>\n </section>\n </div>","PeriodicalId":12819,"journal":{"name":"Haemophilia","volume":"32 5","pages":"1282-1287"},"PeriodicalIF":3.0000,"publicationDate":"2026-09-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/hae.70350","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Haemophilia","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/hae.70350","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/7/5 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
Accurate monitoring of treatment with modified factor VIII products remains challenging, due to known variability in assay performance across different laboratory methods and reagents. For the extended half-life product Damoctocog alfa pegol discrepancies have been reported.
Aims
This multicentre field study was designed to evaluate the analytical accuracy of Damoctocog alfa pegol using various methods for one-stage clotting and chromogenic assays.
Methods
The study was conducted under conditions closely mimicking routine laboratory settings. Five plasma samples were spiked with varying concentrations of Damoctocog alfa pegol (0.05 to 1.3 IU/ml) and distributed to 118 participating laboratories. FVIII activity of the samples was analysed using 12 one-stage and 7 chromogenic methods. Accuracy results were evaluated per specific reagent with a maximal acceptable mean bias of 20% from the target. Interlaboratory variation was established for reagents with ≥3 participants.
Results
The reliability of factor VIII measurements differ between the tested reagents. Usage of SynthasIL and Pathromtin (one stage) and Biophen (chromogenic) is in most cases reliable, although inter-lab variation contributes to inaccuracy. The reagents STA PTT A and Triniclot (one stage), as well as Siemens (chromogenic) (strongly) underestimate FVIII activity and are therefore unsuitable for accurate measurement of Damoctocog alfa pegol.
Conclusion
We have evaluated accuracy results for 19 Factor VIII assays in order to provide evidence-based guidance for clinical laboratories in selecting appropriate assay methodologies for reliable quantification of the activity of Damoctocog alfa pegol. Assay performance should be locally validated as large interlaboratory variation is observed.
期刊介绍:
Haemophilia is an international journal dedicated to the exchange of information regarding the comprehensive care of haemophilia. The Journal contains review articles, original scientific papers and case reports related to haemophilia care, with frequent supplements. Subjects covered include:
clotting factor deficiencies, both inherited and acquired: haemophilia A, B, von Willebrand''s disease, deficiencies of factor V, VII, X and XI
replacement therapy for clotting factor deficiencies
component therapy in the developing world
transfusion transmitted disease
haemophilia care and paediatrics, orthopaedics, gynaecology and obstetrics
nursing
laboratory diagnosis
carrier detection
psycho-social concerns
economic issues
audit
inherited platelet disorders.