Matthan Tharakan, Kishore Balasubramanian, Rohan Aggarwal, Kiana Prather, Helen Shi, Ellen L Air, Brian Dalm, Arjun Khanna, Ryan Kochanski, Paul S Larson, Nader Pouratian, Andrew K Conner
{"title":"Frameless Intraoperative CT-Guided Deep Brain Stimulation: A Multi-Institutional Analysis of Stereotactic Accuracy and Operative Metrics.","authors":"Matthan Tharakan, Kishore Balasubramanian, Rohan Aggarwal, Kiana Prather, Helen Shi, Ellen L Air, Brian Dalm, Arjun Khanna, Ryan Kochanski, Paul S Larson, Nader Pouratian, Andrew K Conner","doi":"10.1159/000553285","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>MRI-guided stereotaxy provides high accuracy but is limited by the need for an interventional/intraoperative MRI environment and associated workflow complexity. ClearPoint Software v3.0 extends an established stereotactic workflow to a conventional operating room using intraoperative CT (iCT) imaging. We aimed to evaluate stereotactic accuracy and operative metrics of the ClearPoint frameless iCT image-guided system for deep brain stimulation (DBS) lead implantation.</p><p><strong>Methods: </strong>We performed a multicenter retrospective review of consecutive DBS lead placement procedures supported by ClearPoint Neuro Navigation System Software Version 3.0 across 7 sites during a limited market release (February 1, 2025-May 1, 2025). Primary outcomes were radial targeting error (distance between planned trajectory endpoint and final device tip position in X-Y) and operative duration (skin incision to closure).</p><p><strong>Results: </strong>Twenty-three procedures were analyzed. Median age was 67 years (range 32-89). Most cases were bilateral (87%, n = 20). Targets included subthalamic nucleus (39%, n = 9), ventral intermediate nucleus (39%, n = 9), and globus pallidus internus (22%, n = 5). Targeting accuracy was high, with median radial error of 0.20 mm on the left (range 0-1.60) and 0.20 mm on the right (range 0-1.0). Median operative duration was 127 min (range 69-225). Median intraoperative scans were 6, and procedures required a median of 2 adjustments per side.</p><p><strong>Conclusion: </strong>In this first multi-institution report of ClearPoint Software v3.0 for DBS lead placement, the frameless iCT workflow demonstrated consistent submillimeter accuracy with favorable operative metrics in a standard operating room setting.</p>","PeriodicalId":22078,"journal":{"name":"Stereotactic and Functional Neurosurgery","volume":" ","pages":"1-12"},"PeriodicalIF":2.1000,"publicationDate":"2026-06-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Stereotactic and Functional Neurosurgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000553285","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"NEUROIMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: MRI-guided stereotaxy provides high accuracy but is limited by the need for an interventional/intraoperative MRI environment and associated workflow complexity. ClearPoint Software v3.0 extends an established stereotactic workflow to a conventional operating room using intraoperative CT (iCT) imaging. We aimed to evaluate stereotactic accuracy and operative metrics of the ClearPoint frameless iCT image-guided system for deep brain stimulation (DBS) lead implantation.
Methods: We performed a multicenter retrospective review of consecutive DBS lead placement procedures supported by ClearPoint Neuro Navigation System Software Version 3.0 across 7 sites during a limited market release (February 1, 2025-May 1, 2025). Primary outcomes were radial targeting error (distance between planned trajectory endpoint and final device tip position in X-Y) and operative duration (skin incision to closure).
Results: Twenty-three procedures were analyzed. Median age was 67 years (range 32-89). Most cases were bilateral (87%, n = 20). Targets included subthalamic nucleus (39%, n = 9), ventral intermediate nucleus (39%, n = 9), and globus pallidus internus (22%, n = 5). Targeting accuracy was high, with median radial error of 0.20 mm on the left (range 0-1.60) and 0.20 mm on the right (range 0-1.0). Median operative duration was 127 min (range 69-225). Median intraoperative scans were 6, and procedures required a median of 2 adjustments per side.
Conclusion: In this first multi-institution report of ClearPoint Software v3.0 for DBS lead placement, the frameless iCT workflow demonstrated consistent submillimeter accuracy with favorable operative metrics in a standard operating room setting.
期刊介绍:
''Stereotactic and Functional Neurosurgery'' provides a single source for the reader to keep abreast of developments in the most rapidly advancing subspecialty within neurosurgery. Technological advances in computer-assisted surgery, robotics, imaging and neurophysiology are being applied to clinical problems with ever-increasing rapidity in stereotaxis more than any other field, providing opportunities for new approaches to surgical and radiotherapeutic management of diseases of the brain, spinal cord, and spine. Issues feature advances in the use of deep-brain stimulation, imaging-guided techniques in stereotactic biopsy and craniotomy, stereotactic radiosurgery, and stereotactically implanted and guided radiotherapeutics and biologicals in the treatment of functional and movement disorders, brain tumors, and other diseases of the brain. Background information from basic science laboratories related to such clinical advances provides the reader with an overall perspective of this field. Proceedings and abstracts from many of the key international meetings furnish an overview of this specialty available nowhere else. ''Stereotactic and Functional Neurosurgery'' meets the information needs of both investigators and clinicians in this rapidly advancing field.