Transitioning from Laboratory-Developed Tests to a Single Commercial Reagent Kit in a National Newborn Screening Program: Impact on Analytical Performance and Harmonization.

IF 3.5 Q1 GENETICS & HEREDITY
Rachel S Carling, Zoe J Barclay, Sophie C Ward, Marie Appleton, Robert Barski, Harry Benn, Kelly Chambers, Paul Coakley, Helena Kemp, Nicola Crabbe, Sarah Dowden, Toby Greenfield, Sarah L Hogg, Saima Hussein, Rhiannon Marr, Oliver Parkes, Darren Powell, Tejswurree Ramgoolam, Joshua Ssali, Nazia Taj, Katherine Wright, Teresa H Y Wu, James R Bonham
{"title":"Transitioning from Laboratory-Developed Tests to a Single Commercial Reagent Kit in a National Newborn Screening Program: Impact on Analytical Performance and Harmonization.","authors":"Rachel S Carling, Zoe J Barclay, Sophie C Ward, Marie Appleton, Robert Barski, Harry Benn, Kelly Chambers, Paul Coakley, Helena Kemp, Nicola Crabbe, Sarah Dowden, Toby Greenfield, Sarah L Hogg, Saima Hussein, Rhiannon Marr, Oliver Parkes, Darren Powell, Tejswurree Ramgoolam, Joshua Ssali, Nazia Taj, Katherine Wright, Teresa H Y Wu, James R Bonham","doi":"10.3390/ijns12020041","DOIUrl":null,"url":null,"abstract":"<p><p>Newborn screening in England is a national program with laboratories adhering to common screening algorithms. Until recently, screening for inherited metabolic disorders was provided by ten laboratories using laboratory-developed tests (LDTs) and three using commercial assays: harmonization of results proved challenging. Introduction of hereditary tyrosinemia type 1 screening meant LDTs required modification to include the measurement of succinylacetone, and subsequent re-validation. This provided an opportunity to implement a single commercial reagent kit in all laboratories. It was anticipated that this would improve analytical performance and harmonization. This study aimed to determine whether these goals were achieved. Verification across the 13 laboratories revealed that the commercial kit reduced inter-laboratory variation for all analytes demonstrating improved harmonization. However, this was achieved by applying instrument-specific correction factors to all analytes, the magnitude of which were significant, indicating a lack of standardization. Performance of succinylacetone was limited by instrument-dependent background interference from the methionine stable isotope label, underscoring the need to establish evidence-based screening cut-off values (COV) rather than adopting published thresholds. This study emphasizes the need for traceable reference materials to improve laboratory quality and the value of screening outcome data.</p>","PeriodicalId":14159,"journal":{"name":"International Journal of Neonatal Screening","volume":"12 2","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-06-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13300528/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Neonatal Screening","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/ijns12020041","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"GENETICS & HEREDITY","Score":null,"Total":0}
引用次数: 0

Abstract

Newborn screening in England is a national program with laboratories adhering to common screening algorithms. Until recently, screening for inherited metabolic disorders was provided by ten laboratories using laboratory-developed tests (LDTs) and three using commercial assays: harmonization of results proved challenging. Introduction of hereditary tyrosinemia type 1 screening meant LDTs required modification to include the measurement of succinylacetone, and subsequent re-validation. This provided an opportunity to implement a single commercial reagent kit in all laboratories. It was anticipated that this would improve analytical performance and harmonization. This study aimed to determine whether these goals were achieved. Verification across the 13 laboratories revealed that the commercial kit reduced inter-laboratory variation for all analytes demonstrating improved harmonization. However, this was achieved by applying instrument-specific correction factors to all analytes, the magnitude of which were significant, indicating a lack of standardization. Performance of succinylacetone was limited by instrument-dependent background interference from the methionine stable isotope label, underscoring the need to establish evidence-based screening cut-off values (COV) rather than adopting published thresholds. This study emphasizes the need for traceable reference materials to improve laboratory quality and the value of screening outcome data.

从实验室开发的测试过渡到单一商业试剂盒在国家新生儿筛查计划:对分析性能和协调的影响。
新生儿筛查在英国是一个国家项目,实验室坚持共同的筛查算法。直到最近,遗传代谢紊乱的筛查是由10个实验室使用实验室开发的测试(LDTs)和3个使用商业分析提供的:结果的统一证明具有挑战性。遗传性酪氨酸血症1型筛查的引入意味着LDTs需要修改以包括琥珀酰丙酮的测量,并随后重新验证。这提供了在所有实验室实施单一商业试剂盒的机会。预期这将改善分析性能和协调。本研究旨在确定这些目标是否实现。13个实验室的验证表明,商业试剂盒减少了所有分析物的实验室间差异,证明了一致性的改善。然而,这是通过对所有分析物应用仪器特定的校正因子来实现的,其幅度显着,表明缺乏标准化。琥珀酰丙酮的性能受到蛋氨酸稳定同位素标签的仪器相关背景干扰的限制,强调需要建立基于证据的筛选临界值(COV),而不是采用已公布的阈值。本研究强调需要可追溯的参考材料,以提高实验室质量和筛选结果数据的价值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Neonatal Screening
International Journal of Neonatal Screening Medicine-Pediatrics, Perinatology and Child Health
CiteScore
6.70
自引率
20.00%
发文量
56
审稿时长
11 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书