PediLoc® blade plate yields faster consolidation rates than LCP pediatric condylar plate in distal femoral osteotomies: a retrospective comparison and description of surgical technique.

IF 4.5 2区 医学 Q1 ORTHOPEDICS
Samuel Tschudi, Cyrill Delay, Britta Krautwurst, Domenic Grisch, Thomas Dreher
{"title":"PediLoc<sup>®</sup> blade plate yields faster consolidation rates than LCP pediatric condylar plate in distal femoral osteotomies: a retrospective comparison and description of surgical technique.","authors":"Samuel Tschudi, Cyrill Delay, Britta Krautwurst, Domenic Grisch, Thomas Dreher","doi":"10.1186/s10195-026-00946-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The PediLoc<sup>®</sup> Locking Cannulated Blade Plate System was originally designed for proximal femoral osteotomies but has also been used for multidimensional correction of distal femoral deformities. The present study delineates the surgical technique and undertakes a retrospective analysis of the consolidation rates of distal femoral osteotomies performed using the aforementioned system, comparing them with those performed using the DePuy Synthes<sup>®</sup> 90° LCP Pediatric Condylar Plate. We hypothesized that the semi-rigid design of the blade plate results in faster osteotomy healing rates compared with other locking screw plate systems.</p><p><strong>Materials and methods: </strong>This retrospective single-centre study included patients who underwent distal femoral osteotomy with the PediLoc<sup>®</sup> Blade Plate (Blade Plate) and the DePuy Synthes<sup>®</sup> 90° LCP Pediatric Condylar Plate (90° LCP-PCP) at the University Children's Hospital Zurich (2020-2024). Consolidation rates were assessed using \"Regenerating Bone Defects Using New Biomedical Engineering\" (REBORNE), \"Radiographic Union Score for Tibial Fractures\" (RUST) and modified RUST (mRUST) scores at 6 weeks and 3 months postoperatively and compared using using Wilcoxon signed-rank test. For the comparison between both intervention groups, the unpaired t-test was used for normally distributed data and the Mann-Whitney U test for non-normally distributed and ordinal data. For all tests, values of p < 0.05 were considered statistically significant. Geometric planes of osteotomy were analysed, and a stepwise correction technique was described.</p><p><strong>Results: </strong>A total of 111 osteotomies with complete radiographic follow-up were analysed (77 Blade Plate [69.4%], 34 90° LCP-PCP [30.6%]). At 6 weeks, early consolidation (RUST score ≥ 10) was observed in 39% of Blade Plate cases versus 6% in the 90° LCP group. By 3 months, consolidation reached 91% in the Blade Plate group and 70% in the 90° LCP-PCP group. One Blade Plate case was excluded due to an implant-associated intraoperative fracture.</p><p><strong>Conclusions: </strong>The PediLoc<sup>®</sup> Blade Plate facilitates stable, multidirectional correction of distal femoral deformities and achieves earlier radiographic consolidation of osteotomies compared with the 90° LCP-PCP without increased complication rates. Since faster consolidation was observed only in the RUST scores, the clinical relevance of this finding remains to be determined through prospective randomized controlled trials.</p><p><strong>Level of evidence: </strong>Level III, retrospective comparative study.</p><p><strong>Trial registration: </strong>Clinicaltrials.gov, 2025-0000, registration date 14 April 2025, retrospectively registered, https://register.</p><p><strong>Clinicaltrials: </strong>gov/prs/beta/records.</p>","PeriodicalId":48603,"journal":{"name":"Journal of Orthopaedics and Traumatology","volume":" ","pages":""},"PeriodicalIF":4.5000,"publicationDate":"2026-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13550261/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedics and Traumatology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s10195-026-00946-2","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The PediLoc® Locking Cannulated Blade Plate System was originally designed for proximal femoral osteotomies but has also been used for multidimensional correction of distal femoral deformities. The present study delineates the surgical technique and undertakes a retrospective analysis of the consolidation rates of distal femoral osteotomies performed using the aforementioned system, comparing them with those performed using the DePuy Synthes® 90° LCP Pediatric Condylar Plate. We hypothesized that the semi-rigid design of the blade plate results in faster osteotomy healing rates compared with other locking screw plate systems.

Materials and methods: This retrospective single-centre study included patients who underwent distal femoral osteotomy with the PediLoc® Blade Plate (Blade Plate) and the DePuy Synthes® 90° LCP Pediatric Condylar Plate (90° LCP-PCP) at the University Children's Hospital Zurich (2020-2024). Consolidation rates were assessed using "Regenerating Bone Defects Using New Biomedical Engineering" (REBORNE), "Radiographic Union Score for Tibial Fractures" (RUST) and modified RUST (mRUST) scores at 6 weeks and 3 months postoperatively and compared using using Wilcoxon signed-rank test. For the comparison between both intervention groups, the unpaired t-test was used for normally distributed data and the Mann-Whitney U test for non-normally distributed and ordinal data. For all tests, values of p < 0.05 were considered statistically significant. Geometric planes of osteotomy were analysed, and a stepwise correction technique was described.

Results: A total of 111 osteotomies with complete radiographic follow-up were analysed (77 Blade Plate [69.4%], 34 90° LCP-PCP [30.6%]). At 6 weeks, early consolidation (RUST score ≥ 10) was observed in 39% of Blade Plate cases versus 6% in the 90° LCP group. By 3 months, consolidation reached 91% in the Blade Plate group and 70% in the 90° LCP-PCP group. One Blade Plate case was excluded due to an implant-associated intraoperative fracture.

Conclusions: The PediLoc® Blade Plate facilitates stable, multidirectional correction of distal femoral deformities and achieves earlier radiographic consolidation of osteotomies compared with the 90° LCP-PCP without increased complication rates. Since faster consolidation was observed only in the RUST scores, the clinical relevance of this finding remains to be determined through prospective randomized controlled trials.

Level of evidence: Level III, retrospective comparative study.

Trial registration: Clinicaltrials.gov, 2025-0000, registration date 14 April 2025, retrospectively registered, https://register.

Clinicaltrials: gov/prs/beta/records.

PediLoc®钢板比LCP儿童髁钢板在股骨远端截骨术中的巩固率更快:手术技术的回顾性比较和描述。
背景:PediLoc®锁定空心钢板系统最初设计用于股骨近端截骨,但也用于股骨远端畸形的多维矫正。本研究描述了手术技术,并对使用上述系统进行股骨远端截骨术的巩固率进行了回顾性分析,并将其与使用DePuy Synthes®90°LCP儿童髁钢板进行了比较。我们假设,与其他锁定螺钉钢板系统相比,半刚性钢板设计的截骨愈合速度更快。材料和方法:本回顾性单中心研究纳入了在苏黎世大学儿童医院(2020-2024)使用PediLoc®刀片钢板(Blade Plate)和DePuy Synthes®90°LCP儿科髁钢板(90°LCP- pcp)行股骨远端截骨术的患者。在术后6周和3个月,使用“使用新生物医学工程再生骨缺损”(REBORNE)、“胫骨骨折放射学联合评分”(RUST)和改良RUST (mRUST)评分评估巩固率,并使用Wilcoxon sign -rank检验进行比较。两干预组间比较,正态分布数据采用非配对t检验,非正态分布和有序数据采用Mann-Whitney U检验。结果:共分析了111例截骨手术并进行了完整的影像学随访(77例Blade Plate[69.4%], 34例90°LCP-PCP[30.6%])。6周时,39%的Blade Plate病例出现早期实变(RUST评分≥10),而90°LCP组为6%。3个月时,Blade Plate组的巩固率达到91%,90°LCP-PCP组的巩固率达到70%。一例Blade Plate病例因术中植入物相关骨折而被排除。结论:与90°LCP-PCP相比,PediLoc®钢板可实现股骨远端畸形的稳定、多向矫正,并能更早地实现截骨术的x线巩固,且并发症发生率不增加。由于仅在RUST评分中观察到更快的巩固,因此该发现的临床相关性仍需通过前瞻性随机对照试验来确定。证据等级:III级,回顾性比较研究。试验注册:Clinicaltrials.gov, 2025-0000,注册日期为2025年4月14日,回顾性注册,https://register.Clinicaltrials: gov/prs/beta/records。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Orthopaedics and Traumatology
Journal of Orthopaedics and Traumatology Medicine-Orthopedics and Sports Medicine
CiteScore
4.30
自引率
0.00%
发文量
56
审稿时长
13 weeks
期刊介绍: The Journal of Orthopaedics and Traumatology, the official open access peer-reviewed journal of the Italian Society of Orthopaedics and Traumatology, publishes original papers reporting basic or clinical research in the field of orthopaedic and traumatologic surgery, as well as systematic reviews, brief communications, case reports and letters to the Editor. Narrative instructional reviews and commentaries to original articles may be commissioned by Editors from eminent colleagues. The Journal of Orthopaedics and Traumatology aims to be an international forum for the communication and exchange of ideas concerning the various aspects of orthopaedics and musculoskeletal trauma.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书